What factors drive polyethylene wear in total knee arthroplasty? RESULTS OF A LARGE RETRIEVAL SERIES

被引:18
作者
Currier, J. H. [1 ]
Currier, B. H. [1 ]
Abdel, M. P. [2 ]
Berry, D. J. [2 ]
Titus, A. J. [2 ]
Van Citters, D. W. [1 ]
机构
[1] Dartmouth Coll, Thayer Sch Educ, Hanover, NH 03755 USA
[2] Dartmouth Coll, Hanover, NH 03755 USA
关键词
CROSS-LINKED-POLYETHYLENE; CONVENTIONAL POLYETHYLENE; TIBIAL INSERTS; FIXED-BEARING; OSTEOLYSIS; MECHANISMS; FAILURE; PATIENT; SURFACE; UHMWPE;
D O I
10.1302/0301-620X.103B11.BJJ-2020-2334.R1
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Aims Wear of the polyethylene (PE) tibial insert of total knee arthroplasty (TKA) increases the risk of revision surgery with a significant cost burden on the healthcare system. This study quantifies wear performance of tibial inserts in a large and diverse series of retrieved TKAs to evaluate the effect of factors related to the patient, knee design, and bearing material on tibial insert wear performance. Methods An institutional review board-approved retrieval archive was surveyed for modular PE tibial inserts over a range of in vivo duration (mean 58 months (0 to 290)). Five knee designs, totalling 1,585 devices, were studied. Insert wear was estimated from measured thickness change using a previously published method. Linear regression statistical analyses were used to test association of 12 patient and implant design variables with calculated wear rate. Results Five patient-specific variables and seven implant-specific variables were evaluated for significant association with lower insert wear rate. Six were significant when controlling for other factors: greater patient age, female sex, shorter duration in vivo, polished tray, highly cross-linked PE (HXLPE), and constrained knee design. Conclusion This study confirmed that knee wear rate increased with duration in vivo. Older patients and females had significantly lower wear rates. Polished modular tibial tray surfaces, HXLPE, and constrained TKA designs were device design factors associated with significantly reduced wear rate.
引用
收藏
页码:1695 / 1701
页数:7
相关论文
共 30 条
[1]   The 2012 Frank Stinchfield Award: Decreasing Patient Activity With Aging: Implications for Crosslinked Polyethylene Wear [J].
Battenberg, Andrew K. ;
Hopkins, Jeffrey S. ;
Kupiec, Andrew D. ;
Schmalzried, Thomas P. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2013, 471 (02) :386-392
[2]  
Benjamin J, 2001, CLIN ORTHOP RELAT R, P131
[3]   Knee Wear Measured in Retrievals: A Polished Tray Reduces Insert Wear [J].
Berry, Daniel J. ;
Currier, John H. ;
Mayor, Michael B. ;
Collier, John P. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2012, 470 (07) :1860-1868
[4]   Wear of moderately cross-linked polyethylene in fixed-bearing total knee replacements [J].
Brockett, Claire L. ;
Jennings, Louise M. ;
Hardaker, Catherine ;
Fisher, John .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2012, 226 (H7) :529-535
[5]   Osteolysis after total knee arthroplasty: Influence of tibial baseplate surface finish and sterilization of polyethylene insert - Findings at five to ten years postoperatively [J].
Collier, MB ;
Engh, CA ;
McAuley, JP ;
Ginn, SD ;
Engh, GA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2005, 87A (12) :2702-2708
[6]   Clinical assessment of tibial polyethylene thickness [J].
Collier, MB ;
Jewett, BA ;
Engh, CA .
JOURNAL OF ARTHROPLASTY, 2003, 18 (07) :860-866
[7]   Polyethylene damage on the nonarticular surface of modular total knee prostheses [J].
Cuckler, JM ;
Lemons, J ;
Tamarapalli, JR ;
Beck, P .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (410) :248-253
[8]   Comparison of Wear and Oxidation in Retrieved Conventional and Highly Cross-Linked UHMWPE Tibial Inserts [J].
Currier, Barbara H. ;
Currier, John H. ;
Franklin, Katherine J. ;
Mayor, Michael B. ;
Reinitz, Steven D. ;
Van Citters, Douglas W. .
JOURNAL OF ARTHROPLASTY, 2015, 30 (12) :2349-2353
[9]   Can Microcomputed Tomography Measure Retrieved Polyethylene Wear? Comparing Fixed-bearing and Rotating-platform Knees [J].
Engh, Charles A., Jr. ;
Zimmerman, Rebecca L. ;
Hopper, Robert H., Jr. ;
Engh, Gerard A. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2013, 471 (01) :86-93
[10]   In vivo deterioration of tibial baseplate locking mechanisms in contemporary modular total knee components [J].
Engh, GA ;
Lounici, S ;
Rao, AR ;
Collier, MB .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2001, 83A (11) :1660-1665