Assessing the variability of 23Na MRI in skeletal muscle tissue: Reproducibility and repeatability of tissue sodium concentration measurements in the lower leg at 3 T

被引:28
作者
Gerhalter, Teresa [1 ,2 ,3 ]
Gast, Lena V. [1 ]
Marty, Benjamin [2 ,3 ]
Uder, Michael [1 ]
Carlier, Pierre G. [2 ,3 ]
Nagel, Armin M. [1 ,4 ,5 ]
机构
[1] FAU, Univ Hosp Erlangen, Inst Radiol, Erlangen, Germany
[2] Inst Myol, NMR Lab, Paris, France
[3] CEA, NMR Lab, DRF, IBFJ,MIRCen, Paris, France
[4] DKFZ, Div Med Phys Radiol, Heidelberg, Germany
[5] FAU, Inst Med Phys, Erlangen, Germany
关键词
quantification; repeatability; reproducibility; skeletal muscle tissue; sodium MRI; total tissue sodium content; HEALTHY-SUBJECTS; DEPOSITION; AGREEMENT; EXERCISE;
D O I
10.1002/nbm.4279
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The goal of this study was to evaluate the reproducibility and repeatability of tissue sodium concentration (TSC) measurements using Na-23 MRI in skeletal muscle tissue. Na-23 MRI was performed at 3 T on the right lower leg of eight healthy volunteers (aged 28 +/- 4 years). The examinations were repeated at the same site after similar to 22 weeks to assess the variability over a medium-term period. Additionally, they were scanned at a second site shortly before or shortly after the first visit (within 3 weeks) to evaluate the inter-site reproducibility. Moreover, we analysed the effect of B-0 correction on the variability. Coefficients of variations (CVs) from mean TSC values as well as Bland-Altman plots were used to assess intra-site repeatability and inter-site reproducibility. In phantom measurements, the B-0 correction improved the quantitative accuracy. We observed differences of up to 4.9 mmol/L between the first and second visit and a difference of up to 3.7 mmol/L between the two different sites. The CV for the medium-term repeatability was 15% and the reproducibility CV was 9%. The Bland-Altman plots indicated high agreement between the visits in all muscle regions. The systematic bias of -0.68 mmol/L between site X and Y (P = 0.03) was slightly reduced to -0.64 mmol/L after B-0 correction (P = 0.04). This work shows that TSC measurements in healthy skeletal muscle tissue can be performed with good repeatability and reproducibility, which is of importance for future longitudinal or multicentre studies.
引用
收藏
页数:9
相关论文
共 34 条
[1]   Hyperkalemic Periodic Paralysis and Permanent Weakness: 3-T MR Imaging Depicts Intracellular 23Na Overload-Initial Results [J].
Amarteifio, Erick ;
Nagel, Armin M. ;
Weber, Marc-Andre ;
Jurkat-Rott, Karin ;
Lehmann-Horn, Frank .
RADIOLOGY, 2012, 264 (01) :154-163
[2]  
Bansal N, 2000, J MAGN RESON IMAGING, V11, P532
[3]   Energy metabolism and interstitial fluid displacement in human gastrocnemius during short ischemic cycles [J].
Binzoni, T ;
Quaresima, V ;
Barattelli, G ;
Hiltbrand, E ;
Gürke, L ;
Terrier, F ;
Cerretelli, P ;
Ferrari, M .
JOURNAL OF APPLIED PHYSIOLOGY, 1998, 85 (04) :1244-1251
[4]  
Bland JM, 1999, STAT METHODS MED RES, V8, P135, DOI 10.1177/096228029900800204
[5]   3D 23Na MRI of human skeletal muscle at 7 Tesla: initial experience [J].
Chang, Gregory ;
Wang, Ligong ;
Schweitzer, Mark E. ;
Regatte, Ravinder R. .
EUROPEAN RADIOLOGY, 2010, 20 (08) :2039-2046
[6]   Human skeletal muscle: Sodium MR imaging and quantification-potential applications in exercise and disease [J].
Constantinides, CD ;
Gillen, JS ;
Boada, FE ;
Pomper, MG ;
Bottomley, PA .
RADIOLOGY, 2000, 216 (02) :559-568
[7]   Magnetic resonance-determined sodium removal from tissue stores in hemodialysis patients [J].
Dahlmann, Anke ;
Doerfelt, Kathrin ;
Eicher, Florian ;
Linz, Peter ;
Kopp, Christoph ;
Moessinger, Irina ;
Horn, Stephan ;
Bueschges-Seraphin, Beatrix ;
Wabel, Peter ;
Hammon, Matthias ;
Cavallaro, Alexander ;
Eckardt, Kai-Uwe ;
Kotanko, Peter ;
Levin, Nathan W. ;
Johannes, Bernd ;
Uder, Michael ;
Luft, Friedrich C. ;
Mueller, Dominik N. ;
Titze, Jens M. .
KIDNEY INTERNATIONAL, 2015, 87 (02) :434-441
[8]   Reliability and agreement of sodium (23Na) MRI in calf muscle and skin of healthy subjects from the US [J].
Dyke, Jonathan P. ;
Meyring-Wosten, Anna ;
Zhao, Yize ;
Linz, Peter ;
Thijssen, Stephan ;
Kotanko, Peter .
CLINICAL IMAGING, 2018, 52 :100-105
[9]   Localized B0 shimming based on 23Na MRI at 7 T [J].
Gast, Lena, V ;
Henning, Anke ;
Hensel, Bernhard ;
Uder, Michael ;
Nagel, Armin M. .
MAGNETIC RESONANCE IN MEDICINE, 2020, 83 (04) :1339-1347
[10]   23Na MRI depicts early changes in ion homeostasis in skeletal muscle tissue of patients with duchenne muscular dystrophy [J].
Gerhalter, Teresa ;
Gast, Lena, V ;
Marty, Benjamin ;
Martin, Jan ;
Trollmann, Regina ;
Schuessler, Stephanie ;
Roemer, Frank ;
Laun, Frederik B. ;
Uder, Michael ;
Schroeder, Rolf ;
Carlier, Pierre G. ;
Nagel, Armin M. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2019, 50 (04) :1103-1113