SOLANG: A user-friendly code to calculate the geometry factor using Monte Carlo simulations. Application to alpha-particle spectrometry

被引:12
作者
Cornejo Diaz, N. A. [2 ]
Martin Sanchez, A. [1 ]
de la Torre Perez, J. [1 ]
机构
[1] Univ Extremadura, Dept Fis, E-06071 Badajoz, Spain
[2] Ctr Protecc & Higiene Radiac, Havana 6195, Cuba
关键词
Geometrical factor; Monte Carlo simulation; Alpha-particle spectrometry; SOURCE-DETECTOR SYSTEM; SOLID-ANGLE; CIRCULAR DETECTOR; SERIES EXPANSION; LINEAR SOURCE; DISK SOURCE;
D O I
10.1016/j.apradiso.2011.01.038
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
Monte Carlo simulation was applied to calculate the effective solid angle (or geometry factor) presented by a plane radioactive source at a detector entrance window. A fast and user-friendly computer program SOLANG was written to perform the calculations for disk or rectangular sources and circular non-coaxial detector disks. Results can be achieved with great precision, depending on the number of simulated trajectories. Some checks and applications to the calculation of efficiencies of semiconductor detectors and gas ionization chambers used to measure alpha particles are presented. Their results were very reliable. The code is available free of charge on request to the authors. (C) 2011 Elsevier Ltd. All rights reserved.
引用
收藏
页码:822 / 824
页数:3
相关论文
共 16 条
[2]   Geometrical efficiency for a parallel disk source and detector [J].
Carrillo, HRV .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 371 (03) :535-537
[3]   SOLID ANGLE SUBTENDED BY 2 RECTANGLES [J].
COOK, J .
NUCLEAR INSTRUMENTS & METHODS, 1980, 178 (2-3) :561-564
[4]   Assessment of the suitability of different random number generators for Monte Carlo simulations in gamma-ray spectrometry [J].
Diaz, N. Cornejo ;
Gil, A. Vergara ;
Jurado Vargas, M. .
APPLIED RADIATION AND ISOTOPES, 2010, 68 (03) :469-473
[5]   A comparison of different analytical methods of determining the solid angle of a circular coaxial source-detector system [J].
Galiano, E ;
Rodrigues, M .
APPLIED RADIATION AND ISOTOPES, 2006, 64 (04) :497-501
[6]   Response to "Comments on "A comparison of different analytical methods of determining the solid angle of a circular coaxial source-detector system"" [J].
Galiano, Eduardo .
APPLIED RADIATION AND ISOTOPES, 2007, 65 (09) :1067-1069
[7]   Defined solid-angle counter with variable geometry [J].
Garcia-Torano, E. ;
Ramiro, T. Duran ;
Burgos, C. ;
Ahedo, M. Begona .
APPLIED RADIATION AND ISOTOPES, 2008, 66 (6-7) :881-885
[8]  
Knoll G.F., 1989, Radiation Detection and Measurement, V2nd
[9]  
MATSUMOTO M, 2002, MERSENNE TWISTER RAN
[10]   SOLID ANGLE SUBTENDED AT A DISK SOURCE BY A NON-PARALLEL DISK DETECTOR [J].
OBLOZINS.P ;
RIBANSKY, I .
NUCLEAR INSTRUMENTS & METHODS, 1971, 94 (01) :187-&