Evaluation and commissioning of a surface based system for respiratory sensing in 4D CT

被引:20
作者
Spadea, Maria Francesca [1 ,2 ]
Baroni, Guido [2 ]
Gierga, David P. [3 ,4 ]
Turcotte, Julie C. [3 ,4 ]
Chen, George T. Y. [3 ,4 ]
Sharp, Gregory C. [3 ,4 ]
机构
[1] Magna Graecia Univ Catanzaro, Dept Expt & Clin Med, I-88100 Catanzaro, Italy
[2] Politecn Milano Univ, Dept Bioengn, Milan, Italy
[3] Harvard Univ, Sch Med, Dept Radiat Oncol, Boston, MA 02114 USA
[4] Massachusetts Gen Hosp, Dept Radiat Oncol, Boston, MA 02114 USA
关键词
4D CT; respiratory motion; surface imaging; multi sensor analysis; 4-DIMENSIONAL COMPUTED-TOMOGRAPHY; RADIATION-THERAPY; RADIOTHERAPY;
D O I
10.1120/jacmp.v12i1.3288
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study is to assess the temporal and reconstruction accuracy of a surface imaging system, the GateCT under ideal conditions, and compare the device with a commonly used respiratory surrogate: the Varian RPM. A clinical CT scanner, run in cine mode, was used with two optical devices, GateCT and RPM, to detect respiratory motion. A radiation detector, GM-10, triggers the X-ray on/off to GateCT system, while the RPM is directly synchronized with the CT scanner through an electronic connection. Two phantoms were imaged: the first phantom translated on a rigid plate along the anterior-posterior (AP) direction, and was used to assess the temporal synchronization of each optical system with the CT scanner. The second phantom, consisting of five spheres translating 3 cm peak-to-peak in the superior-inferior direction, was used to assess the quality of rebinned images created by GateCT and RPM. Calibration assessment showed a nearly perfect synchronization with the scanner for both the RPM and GateCT systems, thus demonstrating the good performance of the radiation detector. Results for the volume rebinning test showed discrepancies in volumes for the 3D reconstruction (compared to ground truth) of up to 36% for GateCT and up to 40% for RPM. No statistical difference was proven between the two systems in volume sorting. Errors are mainly due to phase detection inaccuracies and to the large motion of the phantom. This feasibility study assessed the consistency of two optical systems in synchronizing the respiratory signal with the image acquisition. A new patient protocol based on both RPM and GateCT will be soon started.
引用
收藏
页码:162 / 169
页数:8
相关论文
共 10 条
[1]   A phantom evaluation of a stereo-vision surface imaging system for radiotherapy patient setup [J].
Bert, C ;
Metheany, KG ;
Doppke, K ;
Chen, GTY .
MEDICAL PHYSICS, 2005, 32 (09) :2753-2762
[2]   Radiotherapy of mobile tumors [J].
Jiang, Steve B. .
SEMINARS IN RADIATION ONCOLOGY, 2006, 16 (04) :239-248
[3]   The management of respiratory motion in radiation oncology report of AAPM Task Group 76 [J].
Keall, Paul J. ;
Mageras, Gig S. ;
Balter, James M. ;
Emery, Richard S. ;
Forster, Kenneth M. ;
Jiang, Steve B. ;
Kapatoes, Jeffrey M. ;
Low, Daniel A. ;
Murphy, Martin J. ;
Murray, Brad R. ;
Ramsey, Chester R. ;
Van Herk, Marcel B. ;
Vedam, S. Sastry ;
Wong, John W. ;
Yorke, Ellen .
MEDICAL PHYSICS, 2006, 33 (10) :3874-3900
[4]  
Krengli M, 2009, RADIAT ONCOL, V4, DOI [10.1186/1748-717X-4-9, 10.1186/1748-717X-5-10]
[5]   Advances in 4D medical imaging and 4D radiation therapy [J].
Li, Guang ;
Citrin, Deborah ;
Camphausen, Kevin ;
Mueller, Boris ;
Burman, Chandra ;
Mychalczak, Borys ;
Miller, Robert W. ;
Song, Yulin .
TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2008, 7 (01) :67-81
[6]   Four-dimensional computed tomography: Image formation and clinical protocol [J].
Rietzel, E ;
Pan, TS ;
Chen, GTY .
MEDICAL PHYSICS, 2005, 32 (04) :874-889
[7]   Image guidance: Treatment target localization systems [J].
Sharpe, Michael B. ;
Craig, Tim ;
Moseley, Douglas J. .
IMRT, IGRT, SBRT: ADVANCES IN THE TREATMENT PLANNING AND DELIVERY OF RADIOTHERAPY, 2007, 40 :72-93
[8]   Tumor trailing strategy for intensity-modulated radiation therapy of moving targets [J].
Trofimov, Alexei ;
Vrancic, Christian ;
Chan, Timothy C. Y. ;
Sharp, Gregory C. ;
Bortfeld, Thomas .
MEDICAL PHYSICS, 2008, 35 (05) :1718-1733
[9]   Acquiring a four-dimensional computed tomography dataset using an external respiratory signal [J].
Vedam, SS ;
Keall, PJ ;
Kini, VR ;
Mostafavi, H ;
Shukla, HP ;
Mohan, R .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (01) :45-62
[10]   A novel respiratory detection method based on automated analysis of ultrasound diaphragm video [J].
Xu, QY ;
Hamilton, RJ .
MEDICAL PHYSICS, 2006, 33 (04) :916-921