2 mm Radius Loop Antenna and Linear Active Balun for Near Field Measurement of Magnetic Field in MRI-Conditional Testing of Medical Devices

被引:15
作者
Attaran, Ali [1 ]
Handler, William B. [1 ]
Chronik, Blaine A. [1 ]
机构
[1] Western Univ, Dept Phys & Astron, London, ON N6A 3K7, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
Magnetic resonance imaging; Probes; Baluns; Antennas; Radio frequency; Antenna measurements; Biomedical imaging; Active balun; magnetic field probe; magnetic resonance imaging (MRI) medical device testing; PROBES;
D O I
10.1109/TEMC.2019.2896519
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper presents the design and construction of a magnetic field probe with an active balun for conditional testing of medical devices within a magnetic resonance imaging (MRI). The magnetic field probe was designed using a small loop antenna with 2 mm radius to have high spatial resolution and accuracy. It was tuned and matched at a center frequency of 128 MHz, which corresponds to 3 T MRI systems or equivalent radio frequency (RF) exposure systems for MRI-conditional testing of medical devices. An active balun with a differential transistor topology and a passive low-profile transformer were employed to boost the detected magnetic field signal level lead to higher sensitivity. It also has high input impedance that improves the decoupling of the probe to nearby medical devices. The designed magnetic field probe and active balun have been fabricated on a double-sided printed circuit board, FR4 thickness of 1.57 mm and a copper thickness of 35 mu m, with overall footprint of 22 mm x 11 mm. A verification test setup was developed to generate a known field and calibrate the probe based upon an analytic calculation of field, FDTD simulation and a 10-mm radius passive tuned/matched loop antenna.
引用
收藏
页码:186 / 193
页数:8
相关论文
共 16 条
[1]  
[Anonymous], ELECTROMAGNETIC FIEL
[2]  
[Anonymous], 2017, DTS10974 AAMIISO
[3]   Biosynthesis of palladium nanoparticles using Diospyros kaki leaf extract and determination of antibacterial efficacy [J].
Attar, Azade ;
Yapaoz, Melda Altikatoglu .
PREPARATIVE BIOCHEMISTRY & BIOTECHNOLOGY, 2018, 48 (07) :629-634
[4]   Reliable RF B/E-Field Probes for Time-Domain Monitoring of EM Exposure During Medical Device Testing [J].
Attaran, Ali ;
Handler, William Bradfield ;
Wawrzyn, Krzysztof ;
Menon, Ravi S. ;
Chronik, Blaine A. .
IEEE TRANSACTIONS ON ANTENNAS AND PROPAGATION, 2017, 65 (09) :4815-4823
[5]  
Attarar All., 2017, 2017 IEEE 30th Canadian Conference on Electrical and Computer Engineering (CCECE), P1
[6]  
Huber K. M., 2008, P INT SOC MAG RESON, V16, P1
[7]   STANDARD PROBES FOR ELECTROMAGNETIC-FIELD MEASUREMENTS [J].
KANDA, M .
IEEE TRANSACTIONS ON ANTENNAS AND PROPAGATION, 1993, 41 (10) :1349-1364
[8]   An Accurate Circuit Model for Input Impedance and Radiation Pattern of Two-Port Loop Antennas as E- and H-Probe [J].
Kashanianfard, Mani ;
Sarabandi, Kamal .
IEEE TRANSACTIONS ON ANTENNAS AND PROPAGATION, 2017, 65 (01) :114-120
[9]   A novel automated phasor measurement system for validated and traceable EMC/EMI near-field analysis [J].
Kuehn S. ;
Pfeifer S. ;
Kochali B. ;
Grobbelaar E. ;
Sepan P. ;
Kuster N. .
IEEE Electromagnetic Compatibility Magazine, 2016, 5 (02) :41-46
[10]  
Mispelter J., 2006, NMR Probeheads for Biophysical and Biomedical Experiments: Theoretical Principles & Practical Guidelines, DOI DOI 10.1142/P438