Insertion Speed Affects the Initial Stability of Dental Implants

被引:10
作者
Hsu, Yuan-Yang [1 ]
Tsai, Ming-Tzu [2 ]
Huang, Heng-Li [1 ,3 ]
Fuh, Lih-Jyh [1 ,4 ]
Hsu, Jui-Ting [1 ,3 ,5 ]
机构
[1] China Med Univ, Coll Dent, Sch Dent, 91 Hsueh Shih Rd, Taichung 404, Taiwan
[2] Hungkuang Univ, Dept Biomed Engn, Taichung 433, Taiwan
[3] Asia Univ, Dept Bioinformat & Med Engn, Taichung 413, Taiwan
[4] China Med Univ & Hosp, Dept Dent, Taichung 404, Taiwan
[5] China Med Univ, Coll Biomed Engn, Dept Biomed Engn, Taichung 404, Taiwan
关键词
Dental implant; Initial stability; Inserted speed; Artificial bone; ITV; ISQ; PTV; IN-VITRO; TIGHTENING SPEED; BONE; OSSEOINTEGRATION; THICKNESS;
D O I
10.1007/s40846-022-00742-3
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Purpose This study investigated the effects of insertion speed of dental implants on their stability. Methods Dental implants were inserted at speeds of 4, 10, 20, and 30 rpm, respectively, into artificial bones of two different bone qualities; namely, good bone (GB) and poor bone (PB). Therefore, the four insertion speeds and two bone qualities totaled eight groups, with five specimens per group. During the insertion process, the maximum insertion torque value (ITV) was measured, along with two stability parameters: periotest value (PTV) and initial stability quotient (ISQ). Results Dental implants in the GB group exhibited better stability than those in the PB group (p < 0.001). Significant differences in ITVs were observed among the four insertion speeds in both the GB and PB groups (p = 0.004 and p = 0.034, respectively). In general, the higher the insertion speed, the lower the ITV; in particular, the ITVs for implants inserted at 4 rpm were higher than those measured at 20 and 30 rpm (p = 0.008). However, regardless of the bone quality, the PTVs did not differ significantly among the four insertion speeds (p = 0.066). In both the GB and PB groups, the ISQs differed significantly differences among the four insertion speeds (p = 0.016 and p = 0.004, respectively). Conclusion The stability parameters measured for dental implants in the GB group were all higher than those measured in the PB group. In general, a higher insertion speed resulted in a lower ITV value. The PTV did not differ significantly, while the ISQ differed significantly among the four insertion speeds.
引用
收藏
页码:516 / 525
页数:10
相关论文
共 33 条
[1]  
A F1839-08 American Society for Testing and Materials, 2008, F183908 ASTM
[2]  
A F543-17 American Society for Testing and Materials, 2008, F54317 ASTM
[3]   Comparative study of the accuracy of CBCT implant site measurements using different software programs [J].
Al-Ekrish, Asma'a A. .
SAUDI DENTAL JOURNAL, 2021, 33 (06) :355-361
[4]  
Al-Sabbagh Mohanad, 2019, Dent Clin North Am, V63, P461, DOI 10.1016/j.cden.2019.02.001
[5]   Methods to Improve Osseointegration of Dental Implants in Low Quality (Type-IV) Bone: An Overview [J].
Alghamdi, Hamdan S. .
JOURNAL OF FUNCTIONAL BIOMATERIALS, 2018, 9 (01)
[6]  
Atsumi M, 2007, INT J ORAL MAX IMPL, V22, P743
[7]   Effect of the coefficient of friction and tightening speed on the preload induced at the dental implant complex with the finite element method [J].
Bulaqi, Haddad Arabi ;
Mashhadi, Mahmoud Mousavi ;
Geramipanah, Farideh ;
Safari, Hamed ;
Paknejad, Mojgan .
JOURNAL OF PROSTHETIC DENTISTRY, 2015, 113 (05) :405-411
[8]  
Chakrapani Swarna, 2015, Contemp Clin Dent, V6, P12, DOI 10.4103/0976-237X.149284
[9]   Temperature Field in Bone During Robotic Dental Implant Drilling: Theoretical Models and In Vitro Experiments [J].
Cui, Qi ;
Wang, Russell ;
Faller, Lisa-Marie ;
Kan, Tianshu ;
Jiang, Xianfeng ;
Zhu, Fudong ;
Zhu, Weidong ;
Liu, Yunfeng .
JOURNAL OF MEDICAL AND BIOLOGICAL ENGINEERING, 2022, 42 (02) :253-262
[10]   The insertion torque-depth curve integral as a measure of implant primary stability: An in vitro study on polyurethane foam blocks [J].
Di Stefano, Danilo Alessio ;
Arosio, Paolo ;
Gastaldi, Giorgio ;
Gherlone, Enrico .
JOURNAL OF PROSTHETIC DENTISTRY, 2018, 120 (05) :706-714