Dual-Energy Computed Tomography Proton-Dose Calculation with Scripting and Modified Hounsfield Units

被引:12
作者
Kassaee, Anthony [1 ]
Cheng, Chingyun [2 ]
Yin, Lingshu [1 ]
Zou, Wei [1 ]
Li, Taoran [1 ]
Lin, Alexander [1 ]
Swisher-McClure, Samuel [1 ]
Lukens, John N. [1 ]
Lustig, Robert A. [1 ]
O'Reilly, Shannon [1 ]
Dong, Lei [1 ]
Hytonen, Roni [3 ]
Teo, Boon-Keng Kevin [1 ]
机构
[1] Univ Penn, Dept Radiat Oncol, TRC 2 West,3400 Civ Ctr Blvd, Philadelphia, PA 19104 USA
[2] Rutgers Canc Inst New Jersey, Dept Radiat Oncol, New Brunswick, NJ USA
[3] Varian Med Syst, Palo Alto, CA USA
关键词
dual-energy CT; proton therapy; stopping-power ratios; CT; IMPLEMENTATION; THERAPY;
D O I
10.14338/IJPT-20-00075.1
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To describe an implementation of dual-energy computed tomography (DECT) for calculation of proton stopping-power ratios (SPRs) in a commercial treatment-planning system. The process for validation and the workflow for safe deployment of DECT is described, using single-energy computed tomography (SECT) as a safety check for DECT dose calculation. Materials and Methods: The DECT images were acquired at 80 kVp and 140 kVp and were processed with computed tomography scanner software to derive the electron density and effective atomic number images. Reference SPRs of tissue-equivalent plugs from Gammex (Middleton, Wisconsin) and CIRS (Computerized Imaging Reference Systems, Norfolk, Virginia) electron density phantoms were used for validation and comparison of SECT versus DECT calculated through the Eclipse treatment planning system (Varian Medical Systems, Palo Alto, California) application programming interface scripting tool. An in-house software was also used to create DECT SPR computed tomography images for comparison with the script output. In the workflow, using the Eclipse system application programming interface script, clinical plans were optimized with the SECT image set and then forward-calculated with the DECT SPR for the final dose distribution. In a second workflow, the plans were optimized using DECT SPR with reduced range-uncertainty margins. Results: For the Gammex phantom, the root mean square error in SPR was 1.08% for DECT versus 2.29% for SECT for 10 tissue-surrogates, excluding the lung. For the CIRS Phantom, the corresponding results were 0.74% and 2.27%. When evaluating the head and neck plan, DECT optimization with 2% range-uncertainty margins achieved a small reduction in organ-at-risk doses compared with that of SECT plans with 3.5% range-uncertainty margins. For the liver case, DECT was used to identify and correct the lipiodol SPR in the SECT plan. Conclusion: It is feasible to use DECT for proton-dose calculation in a commercial treatment planning system in a safe manner. The range margins can be reduced to 2% in some sites, including the head and neck.
引用
收藏
页码:62 / 72
页数:11
相关论文
共 15 条
[1]  
[Anonymous], 2014, J APPL CLIN MED PHYS
[2]   Experimental validation of two dual-energy CT methods for proton therapy using heterogeneous tissue samples [J].
Bair, Esther ;
Lalonde, Arthur ;
Zhang, Rongxiao ;
Jee, Kyung-Wook ;
Yang, Kai ;
Sharp, Gregory ;
Liu, Bob ;
Royle, Gary ;
Bouchard, Hugo ;
Lu, Hsiao-Ming .
MEDICAL PHYSICS, 2018, 45 (01) :48-59
[3]   The potential of dual-energy CT to reduce proton beam range uncertainties [J].
Bar, Esther ;
Lalonde, Arthur ;
Royle, Gary ;
Lu, Hsiao-Ming ;
Bouchard, Hugo .
MEDICAL PHYSICS, 2017, 44 (06) :2332-2344
[4]   Dual-Energy Computed Tomography: Physical Principles, Approaches to Scanning, Usage, and Implementation: Part 2 [J].
Forghani, Reza ;
De Man, Bruno ;
Gupta, Rajiv .
NEUROIMAGING CLINICS OF NORTH AMERICA, 2017, 27 (03) :385-+
[5]   COMPUTERIZED TRANSVERSE AXIAL SCANNING (TOMOGRAPHY) .1. DESCRIPTION OF SYSTEM [J].
HOUNDFIELD, GN .
BRITISH JOURNAL OF RADIOLOGY, 1973, 46 (552) :1016-1022
[6]   Comparison of proton therapy treatment planning for head tumors with a pencil beam algorithm on dual and single energy CT images [J].
Hudobivnik, Nace ;
Schwarz, Florian ;
Johnson, Thorsten ;
Agolli, Linda ;
Dedes, George ;
Tessonnier, Thomas ;
Verhaegen, Frank ;
Thieke, Christian ;
Belka, Claus ;
Sommer, Wieland H. ;
Parodi, Katia ;
Landry, Guillaume .
MEDICAL PHYSICS, 2016, 43 (01) :495-504
[7]   Experimental verification of ion stopping power prediction from dual energy CT data in tissue surrogates [J].
Huenemohr, Nora ;
Krauss, Bernhard ;
Tremmel, Christoph ;
Ackermann, Benjamin ;
Jaekel, Oliver ;
Greilich, Steffen .
PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (01) :83-96
[8]   Influence of intravenous contrast agent on dose calculation in proton therapy using dual energy CT [J].
Lalonde, Arthur ;
Xie, Yunhe ;
Burgdorf, Brendan ;
O'Reilly, Shannon ;
Ingram, William Scott ;
Yin, Lingshu ;
Zou, Wei ;
Dong, Lei ;
Bouchard, Hugo ;
Teo, Boon-Ken Kevin .
PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (12)
[10]   The calibration of CT Hounsfield units for radiotherapy treatment planning [J].
Schneider, U ;
Pedroni, E ;
Lomax, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (01) :111-124