Quantitative in vivo microdialysis in pharmacokinetic studies:: Some reminders

被引:27
作者
Cano-Cebrián, MJ [1 ]
Zornoza, T [1 ]
Polache, A [1 ]
Granero, L [1 ]
机构
[1] Univ Valencia, Dept Pharm & Pharmaceut Technol, E-46100 Burjassot, Spain
关键词
D O I
10.2174/1389200053586109
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
This paper reviews the empirical methods of quantitative microdialysis that have been used to interpret the results obtained from pharmacokinetic studies. The concept of extraction efficiency or recovery and the properties of recovery in vivo (variation with flow rate, time dependency and influence of the mode of administration) are considered. The most frequently used methods for determining recovery in vivo are described and evaluated in the light of recent theoretical studies. Specifically, we review the variation of flow rate method, the very slow flow method, the no net flux method and the delivery and retrodialysis methods. Special emphasis is placed on the description of each method, demonstrating its applicability to pharmacokinetic studies conducted under steady-state or transient conditions, and also its limitations. Finally, the more relevant studies that have compared the suitability of these methods are reviewed.
引用
收藏
页码:83 / 90
页数:8
相关论文
共 54 条
  • [1] MICRODIALYSIS - THEORY AND APPLICATION
    BENVENISTE, H
    HUTTEMEIER, PC
    [J]. PROGRESS IN NEUROBIOLOGY, 1990, 35 (03) : 195 - 215
  • [2] DETERMINATION OF BRAIN INTERSTITIAL CONCENTRATIONS BY MICRODIALYSIS
    BENVENISTE, H
    HANSEN, AJ
    OTTOSEN, NS
    [J]. JOURNAL OF NEUROCHEMISTRY, 1989, 52 (06) : 1741 - 1750
  • [3] Recovery as a function of the osmolality of the perfusion medium in microdialysis experiments (Reprinted from Analytica Chimica Acta, vol 375)
    Borg, N
    Ståhle, L
    [J]. ANALYTICA CHIMICA ACTA, 1999, 379 (03) : 319 - 325
  • [4] Methodological aspects of the use of a calibrator in in vivo microdialysis -: Further development of the retrodialysis method
    Bouw, MR
    Hammarlund-Udenaes, M
    [J]. PHARMACEUTICAL RESEARCH, 1998, 15 (11) : 1673 - 1679
  • [5] STEADY-STATE THEORY FOR QUANTITATIVE MICRODIALYSIS OF SOLUTES AND WATER INVIVO AND INVITRO
    BUNGAY, PM
    MORRISON, PF
    DEDRICK, RL
    [J]. LIFE SCIENCES, 1990, 46 (02) : 105 - 119
  • [6] Probe calibration in transient microdialysis in vivo
    Bungay, PM
    Dedrick, RL
    Fox, E
    Balis, FM
    [J]. PHARMACEUTICAL RESEARCH, 2001, 18 (03) : 361 - 366
  • [7] BUNGAY PM, 1991, CURR SEP, V10, P106
  • [8] Quantitative dual-probe microdialysis:: mathematical model and analysis
    Chen, KC
    Höistad, M
    Kehr, J
    Fuxe, K
    Nicholson, C
    [J]. JOURNAL OF NEUROCHEMISTRY, 2002, 81 (01) : 94 - 107
  • [9] A review of microdialysis sampling for pharmacokinetic applications
    Davies, MI
    [J]. ANALYTICA CHIMICA ACTA, 1999, 379 (03) : 227 - 249
  • [10] Methodological issues in microdialysis sampling for pharmacokinetic studies
    de Lange, ECM
    de Boer, AG
    Breimer, DD