"Smart" Continuous Glucose Monitoring Sensors: On-Line Signal Processing Issues

被引:53
作者
Sparacino, Giovanni [1 ]
Facchinetti, Andrea [1 ]
Cobelli, Claudio [1 ]
机构
[1] Univ Padua, Dept Informat Engn, I-35131 Padua, Italy
关键词
diabetes; prediction; filtering; calibration; model; time-series; DIABETIC-PATIENTS; TIME-SERIES; GLYCEMIC CONTROL; PLASMA-GLUCOSE; INSULIN; HYPOGLYCEMIA; OSCILLATIONS; VARIABILITY; CALIBRATION; EXCURSIONS;
D O I
10.3390/s100706751
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The availability of continuous glucose monitoring (CGM) sensors allows development of new strategies for the treatment of diabetes. In particular, from an on-line perspective, CGM sensors can become. smart. by providing them with algorithms able to generate alerts when glucose concentration is predicted to exceed the normal range thresholds. To do so, at least four important aspects have to be considered and dealt with on-line. First, the CGM data must be accurately calibrated. Then, CGM data need to be filtered in order to enhance their signal-to-noise ratio (SNR). Thirdly, predictions of future glucose concentration should be generated with suitable modeling methodologies. Finally, generation of alerts should be done by minimizing the risk of detecting false and missing true events. For these four challenges, several techniques, with various degrees of sophistication, have been proposed in the literature and are critically reviewed in this paper.
引用
收藏
页码:6751 / 6772
页数:22
相关论文
共 61 条
[1]  
[Anonymous], 2005, Dover Books on Engineering
[2]  
Bequette B Wayne, 2010, J Diabetes Sci Technol, V4, P404
[3]  
Bode Bruce, 2004, Diabetes Technol Ther, V6, P105, DOI 10.1089/152091504773731285
[4]   Is blood glucose predictable from previous values? A solicitation for data [J].
Bremer, T ;
Gough, DA .
DIABETES, 1999, 48 (03) :445-451
[5]  
Buckingham Bruce, 2005, Diabetes Technol Ther, V7, P792, DOI 10.1089/dia.2005.7.792
[6]  
Buckingham Bruce, 2008, J Diabetes Sci Technol, V2, P300
[7]  
Buckingham BA, 2006, DIABETES TECHNOL THE, V8, P318
[8]   Integral-based filtering of continuous glucose sensor measurements for glycaemic control in critical care [J].
Chase, J. Geoffrey ;
Hann, Christopher E. ;
Jackson, Monique ;
Lin, Jessica ;
Lotz, Thomas ;
Wong, Xing-Wei ;
Shaw, Geoffrey M. .
COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2006, 82 (03) :238-247
[9]   Prevention of hypoglycemia using risk assessment with a continuous glucose monitoring system [J].
Choleau, C ;
Dokladal, P ;
Klein, JC ;
Ward, WK ;
Wilson, GS ;
Reach, G .
DIABETES, 2002, 51 (11) :3263-3273
[10]   Calibration of a subcutaneous amperometric glucose sensor implanted for 7 days in diabetic patients - Part 2. Superiority of the one-point calibration method [J].
Choleau, C ;
Klein, JC ;
Reach, G ;
Aussedat, B ;
Demaria-Pesce, V ;
Wilson, GS ;
Gifford, R ;
Ward, WK .
BIOSENSORS & BIOELECTRONICS, 2002, 17 (08) :647-654