Bispectral index, entropy, and quantitative electroencephalogram during single-agent xenon anesthesia

被引:40
作者
Laitio, Ruut M. [1 ,2 ]
Kaskinoro, Kimmo [1 ,2 ]
Sarkela, Mika O. K. [1 ,2 ]
Kaisti, Kaike K. [1 ,2 ]
Salmi, Elina [1 ,2 ]
Maksimow, Anu [1 ,2 ]
Langsjo, Jaakko W. [1 ,2 ]
Aantaa, Riku [1 ,2 ]
Kangas, Katja [1 ,2 ]
Jaaskelainen, Satu [1 ,2 ]
Scheinin, Harry [1 ,2 ]
机构
[1] Turku Univ Hosp, Dept Anesthesiol & Intens Care, FIN-20521 Turku, Finland
[2] Univ Turku, Turku PET Ctr, Turku, Finland
关键词
D O I
10.1097/01.anes.0000296106.52472.a6
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background: The aim was to evaluate the performance of anesthesia depth monitors, Bispectral Index (BIS) and Entropy, during single-agent xenon anesthesia in 17 healthy subjects. Methods: After mask induction with xenon and intubation, anesthesia was continued with xenon only. BIS, State Entropy and Response Entropy, and electroencephalogram were monitored throughout induction, steady-state anesthesia, and emergence. The performance of BIS, State Entropy, and Response Entropy were evaluated with prediction probability, sensitivity, and specificity analyses. The power spectrum of the raw electroencephalogram signal was calculated. Results: The mean (SD) xenon concentration during anesthesia was 66.4% (2.4%). BIS, State Entropy, and Response Entropy demonstrated low prediction probability values at loss of response (0.455, 0.656, and 0.619) but 1 min after that the values were high (0.804, 0.941, and 0.929). Thereafter, equally good performance was demonstrated for all indices. At emergence, the prediction probability values to distinguish between steady-state anesthesia and return of response for BIS, State Entropy, and Response Entropy were 0.988, 0.892, and 0.992. No statistical differences between the performances of the monitors were observed. Quantitative electroencephalogram analyses showed generalized increase in total power (P < 0.001), delta (P < 0.001) and theta activity (P < 0.001), and increased alpha activity (P = 0.003) in the frontal brain regions. Conclusions: Electroencephalogram-derived depth of sedation indices BIS and Entropy showed a delay to detect loss of response during induction of xenon anesthesia. Both monitors performed well in distinguishing between conscious and unconscious states during steady-state anesthesia. Xenon-induced changes in electroencephalogram closely resemble those induced by propofol.
引用
收藏
页码:63 / 70
页数:8
相关论文
共 42 条
[1]   A SIMPLE STUDY OF AWARENESS AND DREAMING DURING ANAESTHESIA [J].
BRICE, DD ;
HETHERINGTON, RR ;
UTTING, JE .
BRITISH JOURNAL OF ANAESTHESIA, 1970, 42 (06) :535-+
[2]   Randomized controlled trial of the haemodynamic and recovery effects of xenon or propofol anaesthesia [J].
Coburn, M ;
Kunitz, O ;
Baumert, JH ;
Hecker, K ;
Haaf, S ;
Zühlsdorff, A ;
Beeker, T ;
Rossaint, R .
BRITISH JOURNAL OF ANAESTHESIA, 2005, 94 (02) :198-202
[3]   Xenon reduces glutamate-, AMPA-, and kainate-induced membrane currents in cortical neurones [J].
Dinse, A ;
Föhr, KJ ;
Georgieff, M ;
Beyer, C ;
Bulling, A ;
Weigt, HU .
BRITISH JOURNAL OF ANAESTHESIA, 2005, 94 (04) :479-485
[4]   Monitoring depth of anesthesia - With emphasis on the application of the bispectral index and the middle latency auditory evoked response to the prevention of recall [J].
Drummond, JC .
ANESTHESIOLOGY, 2000, 93 (03) :876-882
[5]   Remifentanil versus alfentanil - Comparative pharmacokinetics and pharmacodynamics in healthy adult male volunteers [J].
Egan, TD ;
Minto, CF ;
Hermann, DJ ;
Barr, J ;
Muir, KT ;
Shafer, SL .
ANESTHESIOLOGY, 1996, 84 (04) :821-833
[6]   Reduction in the incidence of awareness using BIS monitoring [J].
Ekman, A ;
Lindholm, ML ;
Lennmarken, C ;
Sandin, R .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2004, 48 (01) :20-26
[7]   Ketamine and postoperative pain -: a quantitative systematic review of randomised trials [J].
Elia, N ;
Tramèr, MR .
PAIN, 2005, 113 (1-2) :61-70
[8]  
Ellerkmann RK, 2004, ANESTHESIOLOGY, V101, P1275
[9]   How does xenon produce anaesthesia? [J].
Franks, NP ;
Dickinson, R ;
de Sousa, SLM ;
Hall, AC ;
Lieb, WR .
NATURE, 1998, 396 (6709) :324-324
[10]   Cardiovascular effects of xenon and nitrous oxide in patients during fentanyl-midazolam anaesthesia [J].
Goto, T ;
Hanne, P ;
Ishiguro, Y ;
Ichinose, F ;
Niimi, Y ;
Morita, S .
ANAESTHESIA, 2004, 59 (12) :1178-1183