Image quality evaluation of YVO4:Eu phosphor screens for use in X-ray medical imaging detectors

被引:17
作者
Kandarakis, I
Cavouras, D
Kanellopoulos, E
Nomicos, CD
Panayiotakis, GS
机构
[1] Inst Educ Technol, Dept Med Instrumentat Technol, Athens 12210, Greece
[2] Inst Educ Technol, Dept Elect, Athens 12210, Greece
[3] Univ Patras, Sch Med, Dept Med Phys, Patras 26500, Greece
关键词
D O I
10.1016/S1350-4487(98)00058-4
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
YVO4:Eu is a phosphor material emitting red light (621 nm) and providing excellent spectral compatibility with red sensitive optical detectors. However, it has never been used in medical X-ray imaging. In this study the imaging performance of six laboratory prepared YVO4:Eu test screens with coating weights from 30 to 140 mg/cm(2) was examined by evaluating the modulation transfer function (MTF), the X-ray luminescence efficiency (XLE), and the detective quantum efficiency (DQE). DQE was evaluated by employing a method based on XLE and emission spectra measurements. The 30 mg/cm(2) screen had a significantly higher MTF than thicker screens while its zero frequency DQE peak value was 0.27 at 30 kVp. Results obtained indicate that YVO4:Eu may be of value for use in X-ray imaging detectors. (C) 1998 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:481 / 486
页数:6
相关论文
共 21 条
[11]  
KANDARAKIS I, 1997, NUCL INSTR METH PH A, V399, P345
[12]   X-RAY EFFICIENCY OF POWDER PHOSPHORS [J].
LUDWIG, GW .
JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1971, 118 (07) :1152-&
[13]  
MOTZ J W, 1978, Medical Physics (Woodbury), V5, P8, DOI 10.1118/1.594405
[14]   A study of X-ray luminescence and spectral compatibility of europium-activated yttrium-vanadate (YVO4:Eu) screens for medical imaging applications [J].
Panayiotakis, G ;
Cavouras, D ;
Kandarakis, I ;
Nomicos, C .
APPLIED PHYSICS A-MATERIALS SCIENCE & PROCESSING, 1996, 62 (05) :483-486
[15]  
SHAW R, 1984, P SOC PHOTO-OPT INST, V454, P128, DOI 10.1117/12.939323
[16]  
Shaw R., 1963, J PHOTOGR SCI, V11, P199
[17]  
Storm E., 1967, LA3753 U CAL LOS AL
[18]   MOLYBDENUM TARGET X-RAY-SPECTRA - A SEMIEMPIRICAL MODEL [J].
TUCKER, DM ;
BARNES, GT ;
WU, XZ .
MEDICAL PHYSICS, 1991, 18 (03) :402-407
[19]   SEMIEMPIRICAL MODEL FOR GENERATING TUNGSTEN TARGET X-RAY-SPECTRA [J].
TUCKER, DM ;
BARNES, GT ;
CHAKRABORTY, DP .
MEDICAL PHYSICS, 1991, 18 (02) :211-218
[20]   X-ray detectors for digital radiography [J].
Yaffe, MJ ;
Rowlands, JA .
PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (01) :1-39