Image quality evaluation of YVO4:Eu phosphor screens for use in X-ray medical imaging detectors

被引:17
作者
Kandarakis, I
Cavouras, D
Kanellopoulos, E
Nomicos, CD
Panayiotakis, GS
机构
[1] Inst Educ Technol, Dept Med Instrumentat Technol, Athens 12210, Greece
[2] Inst Educ Technol, Dept Elect, Athens 12210, Greece
[3] Univ Patras, Sch Med, Dept Med Phys, Patras 26500, Greece
关键词
D O I
10.1016/S1350-4487(98)00058-4
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
YVO4:Eu is a phosphor material emitting red light (621 nm) and providing excellent spectral compatibility with red sensitive optical detectors. However, it has never been used in medical X-ray imaging. In this study the imaging performance of six laboratory prepared YVO4:Eu test screens with coating weights from 30 to 140 mg/cm(2) was examined by evaluating the modulation transfer function (MTF), the X-ray luminescence efficiency (XLE), and the detective quantum efficiency (DQE). DQE was evaluated by employing a method based on XLE and emission spectra measurements. The 30 mg/cm(2) screen had a significantly higher MTF than thicker screens while its zero frequency DQE peak value was 0.27 at 30 kVp. Results obtained indicate that YVO4:Eu may be of value for use in X-ray imaging detectors. (C) 1998 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:481 / 486
页数:6
相关论文
共 21 条
[1]  
[Anonymous], PHYS MED IMAGING REC
[2]  
Barnes G. T., 1979, PHYSICS MED IMAGING, P138
[3]   THE IMAGE QUALITY CHARACTERISTICS OF A NOVEL ULTRA-HIGH-RESOLUTION FILM-SCREEN SYSTEM [J].
BEUTEL, J ;
MICKEWICH, DJ ;
ISSLER, SL ;
SHAW, R .
PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (09) :1195-1206
[4]   ANALYSIS OF THE DETECTIVE QUANTUM EFFICIENCY OF A RADIOGRAPHIC SCREEN FILM COMBINATION [J].
BUNCH, PC ;
HUFF, KE ;
VANMETTER, R .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1987, 4 (05) :902-909
[5]   An evaluation of the Y2O3:Eu3+ scintillator for application in medical x-ray detectors and image receptors [J].
Cavouras, D ;
Kandarakis, I ;
Panayiotakis, GS ;
Evangelou, EK ;
Nomicos, CD .
MEDICAL PHYSICS, 1996, 23 (12) :1965-1975
[6]  
GIAKOUMAKIS GE, 1991, APPL PHYS A-MATER, V52, P7
[7]  
Gurwich AM, 1995, RADIAT MEAS, V24, P325
[8]  
HENDEE WR, 1970, MED RAD PHYSICS, P145
[9]   X-ray induced luminescence and spatial resolution of La2O2S:Tb phosphor screens [J].
Kandarakis, I ;
Cavouras, D ;
Panayiotakis, G ;
Agelis, T ;
Nomicos, C ;
Giakoumakis, G .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (02) :297-307
[10]   Evaluating x-ray detectors for radiographic applications: A comparison of ZnSCdS:Ag with Gd2O2S:Tb and Y2O2S:Tb screens [J].
Kandarakis, I ;
Cavouras, D ;
Panayiotakis, GS ;
Nomicos, CD .
PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (07) :1351-1373