A calibration method for whole-body counters, using Monte Carlo simulation

被引:14
|
作者
Ishikawa, T
Matsumoto, M
Uchiyama, M
机构
[1] Natl. Inst. of Radiological Sciences, Inage-ku, Chiba, 263
关键词
D O I
10.1093/oxfordjournals.rpd.a031585
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
A Monte Carlo simulation code was developed to estimate the counting efficiencies in whole-body counting for various body sizes. The code consists of mathematical models and parameters which are categorised into three groups: a geometrical model for phantoms and detectors, a photon transport model, and a detection system model. Photon histories were simulated with these models. The counting efficiencies for five Cs-137 block phantoms of different sizes were calculated by the code and compared with those measured with a whole-body counter at NIRS (Japan). The phantoms corresponded to a newborn, a 5 month old, a 6 year old, an 11 year old and an adult. The differences between the measured and calculated values were within 6%. For the adult phantom, the difference was 0.5%. The results suggest that the Monte Carlo simulation code can be used to estimate the counting efficiencies for various body sizes.
引用
收藏
页码:283 / 288
页数:6
相关论文
共 50 条
  • [1] Monte Carlo calibration of whole-body counters with NaI(Tl) detectors in stretcher geometry
    Breustedt, B.
    Eschner, W.
    RADIATION PROTECTION DOSIMETRY, 2010, 139 (04) : 510 - 518
  • [2] Virtual calibration of whole-body counters to consider the size dependency of counting efficiency using Monte Carlo simulations
    Park, MinSeok
    Kim, Han Sung
    Yoo, Jaeryong
    Kim, Chan Hyeong
    Jang, Won Il
    Park, Sunhoo
    NUCLEAR ENGINEERING AND TECHNOLOGY, 2021, 53 (12) : 4122 - 4129
  • [3] Application of voxel phantoms and Monte Carlo method to whole-body counter calibration
    Kinase, S.
    Takagi, S.
    Noguchi, H.
    Saito, K.
    RADIATION PROTECTION DOSIMETRY, 2007, 125 (1-4) : 189 - 193
  • [4] MONTE CARLO SIMULATION OF A WHOLE-BODY COUNTER USING IGOR PHANTOMS
    Bochud, Francois O.
    Laedermann, Jean-Pascal
    Baechler, Sebastien
    Bailat, Claude J.
    Boschung, Markus
    Aroua, Abbas
    Mayer, Sabine
    RADIATION PROTECTION DOSIMETRY, 2014, 162 (03) : 280 - 288
  • [5] INTERCOMPARISON OF WHOLE-BODY COUNTERS USING A MULTINUCLIDE CALIBRATION PHANTOM
    FENWICK, JD
    MCKENZIE, AL
    BODDY, K
    PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (02): : 191 - 198
  • [6] Calibration of a whole body counter using Monte Carlo methods
    Breustedt, Bastian
    Schieck, Hans Paetz Gen
    Schicha, Harald
    Eschner, Wolfgang
    Zeitschrift fur Medizinische Physik, 2004, 14 (02): : 85 - 94
  • [7] WHOLE-BODY COUNTERS
    REMENCHI.AP
    MILLER, CE
    JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1966, 197 (01): : 47 - &
  • [8] COMPARISON OF TWO ANTHROPOMORPHIC PHANTOMS AS A CALIBRATION TOOL FOR WHOLE-BODY COUNTER USING MONTE CARLO SIMULATIONS
    Manohari, M.
    Mathiyarasu, R.
    Rajagopal, V.
    Venkatraman, B.
    RADIATION PROTECTION DOSIMETRY, 2015, 164 (03) : 298 - 303
  • [9] Monte Carlo calculations for efficiency calibration of a whole-body monitor using BOMAB phantoms of different sizes
    Bhati, S.
    Patni, H. K.
    Ghare, V. P.
    Singh, I. S.
    Nadar, M. Y.
    RADIATION PROTECTION DOSIMETRY, 2012, 148 (04) : 414 - 419
  • [10] EVALUATION OF COUNTING EFFICIENCIES OF A WHOLE-BODY COUNTER USING MONTE CARLO SIMULATION WITH VOXEL PHANTOMS
    Takahashi, Masa
    Kinase, Sakae
    Kramer, Richard
    RADIATION PROTECTION DOSIMETRY, 2011, 144 (1-4) : 407 - 410