Comparison between two different approaches in the uncertainty estimation for the chromatographic analysis of aflatoxin M1: from single laboratory validation data and from proficiency test results

被引:1
作者
Nocentini, M. [1 ]
Focardi, C. [1 ]
Ninci, S. [1 ]
Biancalani, G. [1 ]
机构
[1] Ist Zooprofilatt Sperimentale Reg Lazio & Toscana, Sez Firenze, I-50010 Florence, Italy
关键词
aflatoxin M-1; immunoaffinity clean-up; validation; proficiency testing; IMMUNOAFFINITY COLUMNS; LIQUID-CHROMATOGRAPHY; ANALYTICAL-CHEMISTRY; CALIBRATION; CLEANUP;
D O I
10.3920/WMJ2008.1092
中图分类号
TS2 [食品工业];
学科分类号
0832 ;
摘要
The analytical method to determine aflatoxin M-1 in milk by HPLC with immunoaffinity column clean-up and fluorescence detection has been validated by using certified reference materials and spiked samples. These validation data have been analysed to obtain the uncertainty estimation following the bottom-up approach as reported by Eurachem. The results obtained with this method have been compared with the expanded uncertainty calculated from data obtained during the years 2002-2007 by participation in Fapas proficiency testing.
引用
收藏
页码:381 / 390
页数:10
相关论文
共 18 条
  • [1] [Anonymous], 2002, Off. J. Eur. Comm, VL221, P8
  • [2] [Anonymous], 2006, Off. J. Eur. Communities, VL70, P12
  • [3] Barwick VJ, 1999, ANALYST, V124, P981, DOI 10.1039/a901845j
  • [4] BARWICK VJ, 1998, LGCVAM1998088
  • [5] Estimation of measurement uncertainty for the determination of aflatoxin M1 in milk using immunoaffinity clean-up procedures
    Calaresu, G
    Piras, P
    Carta, G
    Galarini, R
    Chessa, G
    [J]. ACCREDITATION AND QUALITY ASSURANCE, 2006, 11 (1-2) : 10 - 16
  • [6] Assessment of quality performance parameters for straight line calibration curves related to the spread of the abscissa values around their mean
    De Beer, Jacques O.
    De Beer, Thomas R.
    Goeyens, Leo
    [J]. ANALYTICA CHIMICA ACTA, 2007, 584 (01) : 57 - 65
  • [7] Interlaboratory study of a liquid chromatography method for erythromycin: determination of uncertainty
    Dehouck, P
    Heyden, YV
    Smeyers-Verbeke, J
    Massart, DL
    Marini, RD
    Chiap, P
    Hubert, P
    Crommen, J
    Van de Wauw, W
    De Beer, J
    Cox, R
    Mathieu, G
    Reepmeyer, JC
    Voigt, B
    Estevenon, O
    Nicolas, A
    Van Schepdael, A
    Adams, E
    Hoogmartens, J
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2003, 1010 (01) : 63 - 74
  • [8] Determination of uncertainty in analytical measurements from collaborative study results on the analysis of a phenoxymethylpenicillin sample
    Dehouck, P
    Vander Heydon, Y
    Smeyers-Verbeke, J
    Massart, DL
    Crommen, J
    Hubert, P
    Marini, RD
    Smeets, OSNM
    Decristoforo, G
    Van de Wauw, W
    De Beer, J
    Quaglia, MG
    Stella, C
    Veuthey, JL
    Estevenon, O
    Van Schepdael, A
    Roets, E
    Hoogmartens, J
    [J]. ANALYTICA CHIMICA ACTA, 2003, 481 (02) : 261 - 272
  • [9] ELLISON S.L. R., 2000, QUANTIFYING UNCERTAI
  • [10] *EUR CIT, 1998, FITN PURP AN METH LA