Reference signal less Fourier analysis based motion artifact removal algorithm for wearable photoplethysmography devices to estimate heart rate during physical exercises

被引:24
作者
Pankaj [1 ]
Kumar, Ashish [2 ]
Komaragiri, Rama [1 ]
Kumar, Manjeet [3 ]
机构
[1] Bennett Univ, Dept Elect & Commun Engn, Greater Noida 201310, Uttar Pradesh, India
[2] Vellore Inst Technol, Sch Elect Engn, Chennai 600127, Tamil Nadu, India
[3] Delhi Technol Univ DTU Rohini, Dept Elect & Commun Engn, Delhi 110042, India
关键词
Photoplethysmographic(PPG) signal; Fast Fourier transform(FFT); Fourier decomposition method (FDM); Motion artifacts(MA); Heart rate (HR) estimation; HR tracking; PPG SIGNALS; REDUCTION; FRAMEWORK;
D O I
10.1016/j.compbiomed.2021.105081
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
Context: Accurate and reliable heart rate (HR) estimation using photoplethysmographic (PPG)-enabled wearable devices in real-time during daily life activities is challenging. Problem: A PPG signal recorded using a wearable PPG device is corrupted by motion artifacts. Therefore, the main challenge of monitoring HR in real time is the accurate reconstruction of a clean PPG signal by suppressing motion artifacts. Proposed approach: The proposed algorithm employs the Fourier theory-based Fourier decomposition method (FDM) to suppress motion artifacts and a fast Fourier transform (FFT)-based method to estimate the HR. In this paper, a computationally efficient algorithm that does not require a reference accelerometer signal to suppress motion artifacts to estimate HR in real time during physical activities is proposed. Methodology: The noisy PPG signal is decomposed into a desired set of orthogonal Fourier intrinsic band functions (FIBFs). A clean PPG signal is obtained by discarding the FIBFs corrupted with noise and superpositioning the clean FIBFs. Clean FIBFs were further used to estimate the HR. Results: The proposed method is evaluated by computing the mean absolute error (MAE) and percentage absolute error (PAE) on two publicly available datasets, IEEE SPC (training and test) and BAMI (BAMI-I and BAMI-II). The MAE and PAE values computed with the proposed method using the IEEE SPC dataset were (1.87, 1.71). The MAE and PAE values computed using the proposed method on the BAMI-I and BAMI-II datasets were (1.33, 1.13) and (1.45, 1.17), respectively. The computed MAE and PAE values were more accurate than those of state-of-the-art techniques presented in the literature. Conclusion: Owing to the improved accuracy and speed, the proposed HR estimation algorithm can be implemented in wearable health monitoring devices for continuous and reliable HR estimation in real time. The proposed algorithm can be applied to denoise PPG signals with different sampling rates.
引用
收藏
页数:14
相关论文
共 39 条
  • [1] Ajdaraga E, 2017, 2017 25TH TELECOMMUNICATION FORUM (TELFOR), P685
  • [2] Photoplethysmography and its application in clinical physiological measurement
    Allen, John
    [J]. PHYSIOLOGICAL MEASUREMENT, 2007, 28 (03) : R1 - R39
  • [3] CASINOR: Combination of adaptive filters using single noise reference signal for heart rate estimation from PPG signals
    Arunkumar, K. R.
    Bhaskar, M.
    [J]. SIGNAL IMAGE AND VIDEO PROCESSING, 2020, 14 (08) : 1507 - 1515
  • [4] Heart rate estimation from wrist-type photoplethysmography signals during physical exercise
    Arunkumar, K. R.
    Bhaskar, M.
    [J]. BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2020, 57
  • [5] Heart rate estimation from photoplethysmography signal for wearable health monitoring devices
    Arunkumar, K. R.
    Bhaskar, M.
    [J]. BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2019, 50 : 1 - 9
  • [6] Bashar Syed Khairul, 2018, 2018 IEEE EMBS International Conference on Biomedical & Health Informatics (BHI), P79, DOI 10.1109/BHI.2018.8333374
  • [7] Reduced complexity algorithm for heart rate monitoring from PPG signals using automatic activity intensity classifier
    Biagetti, Giorgio
    Crippa, Paolo
    Falaschetti, Laura
    Orcioni, Simone
    Turchetti, Claudio
    [J]. BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2019, 52 : 293 - 301
  • [8] Pulse oximetry: Understanding its basic principles facilitates appreciation of its limitations
    Chan, Edward D.
    Chan, Michael M.
    Chan, Mallory M.
    [J]. RESPIRATORY MEDICINE, 2013, 107 (06) : 789 - 799
  • [9] Real-Time Robust Heart Rate Estimation From Wrist-Type PPG Signals Using Multiple Reference Adaptive Noise Cancellation
    Chowdhury, Sayeed Shafayet
    Hyder, Rakib
    Bin Hafiz, Md. Samzid
    Haque, Mohammad Ariful
    [J]. IEEE JOURNAL OF BIOMEDICAL AND HEALTH INFORMATICS, 2018, 22 (02) : 450 - 459
  • [10] Deep Learning for Heart Rate Estimation From Reflectance Photoplethysmography With Acceleration Power Spectrum and Acceleration Intensity
    Chung, Heewon
    Ko, Hoon
    Lee, Hooseok
    Lee, Jinseok
    [J]. IEEE ACCESS, 2020, 8 : 63390 - 63402