Doppler echocardiographic assessment of valvular regurgitation severity by measurement of the vena contracta: An in vitro validation study

被引:26
作者
Mascherbauer, J
Rosenhek, R
Bittner, B
Binder, J
Simon, P
Maurer, G
Schima, H
Baumgartner, H
机构
[1] Univ Vienna, Vienna Gen Hosp, Dept Cardiol, Dept Med Stat, A-1090 Vienna, Austria
[2] Univ Vienna, Vienna Gen Hosp, Ludwig Boltzmann Inst Cardiosurg Res, A-1090 Vienna, Austria
[3] Univ Vienna, Vienna Gen Hosp, Dept Cardiothorac Surg, A-1090 Vienna, Austria
基金
奥地利科学基金会;
关键词
D O I
10.1016/j.echo.2005.03.021
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. Noninvasive quantitation of valvular regurgitation remains a difficult problem. Measurements of the vena contracta (VC) by color Doppler echocardiography have been proposed but limited data are available on the actual accuracy of this method. Methods: To evaluate how closely the color Doppler VC reflects the true fluid dynamic VC and the anatomic regurgitant orifice and whether this measurement is affected by flow changes, various models of valvular regurgitation were studied in an in vitro flow circuit. The VC diameter was measured with color Doppler using two different ultrasound systems (Agilent Sonos 5500; Agilent Technologies Inc, Palo Alto, Calif and Vingmed CPM 800; GE Healthcare, Chalfront St Giles, UK). Optical planimetry of the anatomic regurgitant orifice was performed, the true VC diameter was determined by laser particle flow visualization. Results: Because of flow contraction, the true VC diameter was consistently smaller than the anatomic regurgitant orifice diameter. Anatomic orifice and true VC only marginally changed with flow rate. The diameter of the color Doppler VC, however, not only overestimated the anatomic orifice diameter by 45% to 60% and the true VC diameter by 130% to 160%, but was also highly affected by the flow rate and the ultrasound system. Despite these limitations a color Doppler VC diameter of 0.77 cm or more (Agilent) and 0.89 cm or more (Vingmed) detected severe regurgitation with a sensitivity of 93% and 84% and a specificity of 96% and 79%, respectively. Conclusions: Color Doppler estimates of the VC markedly overestimate regurgitant orifice and true VC. In contrast to the true VC, Doppler measurements are significantly affected by flow rate and by the ultrasound system used. Nevertheless, they allow semiquantitative assessment of valvular regurgitation separating severe from nonsevere regurgitation with acceptable accuracy.
引用
收藏
页码:999 / 1006
页数:8
相关论文
共 32 条
[1]   A NEW METHOD FOR QUANTITATION OF MITRAL REGURGITATION BASED ON COLOR FLOW DOPPLER IMAGING OF FLOW CONVERGENCE PROXIMAL TO REGURGITANT ORIFICE [J].
BARGIGGIA, GS ;
TRONCONI, L ;
SAHN, DJ ;
RECUSANI, F ;
RAISARO, A ;
DESERVI, S ;
VALDESCRUZ, LM ;
MONTEMARTINI, C .
CIRCULATION, 1991, 84 (04) :1481-1489
[2]  
Baumgartner H, 1991, J Am Soc Echocardiogr, V4, P57
[3]  
BAUMGARTNER H, 1990, AM J CARDIOL, V67, P314
[4]   COMPARING THE AREAS UNDER 2 OR MORE CORRELATED RECEIVER OPERATING CHARACTERISTIC CURVES - A NONPARAMETRIC APPROACH [J].
DELONG, ER ;
DELONG, DM ;
CLARKEPEARSON, DI .
BIOMETRICS, 1988, 44 (03) :837-845
[5]  
Dujardin KS, 1997, CIRCULATION, V96, P3409
[6]   EFFECTIVE MITRAL REGURGITANT ORIFICE AREA - CLINICAL USE AND PITFALLS OF THE PROXIMAL ISOVELOCITY SURFACE-AREA METHOD [J].
ENRIQUEZSARANO, M ;
MILLER, FA ;
HAYES, SN ;
BAILEY, KR ;
TAJIK, AJ ;
SEWARD, JB .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1995, 25 (03) :703-709
[7]   COLOR FLOW IMAGING COMPARED WITH QUANTITATIVE DOPPLER ASSESSMENT OF SEVERITY OF MITRAL REGURGITATION - INFLUENCE OF ECCENTRICITY OF JET AND MECHANISM OF REGURGITATION [J].
ENRIQUEZSARANO, M ;
TAJIK, AJ ;
BAILEY, KR ;
SEWARD, JB .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 21 (05) :1211-1219
[8]   QUANTITATIVE DOPPLER ASSESSMENT OF VALVULAR REGURGITATION [J].
ENRIQUEZSARANO, M ;
BAILEY, KR ;
SEWARD, JB ;
TAJIK, AJ ;
KROHN, MJ ;
MAYS, JM .
CIRCULATION, 1993, 87 (03) :841-848
[9]   CHANGES IN EFFECTIVE REGURGITANT ORIFICE THROUGHOUT SYSTOLE IN PATIENTS WITH MITRAL-VALVE PROLAPSE - A CLINICAL-STUDY USING THE PROXIMAL ISOVELOCITY SURFACE-AREA METHOD [J].
ENRIQUEZSARANO, M ;
SINAK, LJ ;
TAJIK, AJ ;
BAILEY, KR ;
SEWARD, JB .
CIRCULATION, 1995, 92 (10) :2951-2958
[10]  
GROSSMAN W, 1985, CARDIAC CATHETERIZAT, P289