Validation of SIMIND Monte Carlo Simulation Software for Modelling a Siemens Symbia T SPECT Scintillation Camera

被引:5
作者
Ejeh, John E. [1 ,2 ]
van Staden, Johan A. [1 ]
du Raan, Hanlie [1 ]
机构
[1] Univ Free State, Dept Med Phys, Bloemfontein, South Africa
[2] Univ Coll Hosp Ibadan, Dept Nucl Med, Ibadan, Nigeria
来源
WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2018, VOL 1 | 2019年 / 68卷 / 01期
关键词
Computer simulation; Code validation; SIMIND; Septal penetration fraction; Monte Carlo; SEGMENTATION; SCATTER;
D O I
10.1007/978-981-10-9035-6_106
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The use of Monte Carlo (MC) simulations in nuclear medicine is on the rise due to its usefulness in modelling many physical phenomena which are difficult or impossible to measure. To use MC for gamma camera image simulations, it is important to validate the simulated data with the measured data to ensure that what is simulated is a true representation of the physical system being modelled. Therefore, in this study, the SIMIND MC code for modelling a Siemens Symbia T dual head SPECT/CT gamma camera fitted with high energy all-purpose collimators (HEAP) was validated. The validation was done by acquiring and simulating certain NEMA stipulated performance measurements for I-131. These measured and simulated performance data were processed and the results compared. The FWHM and FWTM for the gamma camera for measured and simulated system spatial resolution were 14.4, 29.7 mm versus 13.9, 25.9 mm. System sensitivity performance measurements and simulation produced results of 49 cps/MBq and 48 cps/MBq respectively. The corresponding measured and simulated values for the septal penetration were 3.8 and 4.2%. From this study it is evident that the SIMIND MC Code can be used with confidence to mimic I-131 studies acquired with a Siemens Symbia T SPECT/CT gamma camera fitted with HEAP collimators.
引用
收藏
页码:573 / 576
页数:4
相关论文
共 18 条
[1]  
[Anonymous], NAT ASS REP CLIM CHA
[2]  
Azarm Ahmadreza, 2015, World J Nucl Med, V14, P160, DOI 10.4103/1450-1147.163242
[3]   Accuracy of Monte Carlo simulations compared to in-vivo MDCT dosimetry [J].
Bostani, Maryam ;
Mueller, Jonathon W. ;
McMillan, Kyle ;
Cody, Dianna D. ;
Cagnon, Chris H. ;
DeMarco, John J. ;
McNitt-Gray, Michael F. .
MEDICAL PHYSICS, 2015, 42 (02) :1080-1086
[4]   Strategies for CT tissue segmentation for Monte Carlo calculations in nuclear medicine dosimetry [J].
Braad, P. E. N. ;
Andersen, T. ;
Hansen, S. B. ;
Hoilund-Carlsen, P. F. .
MEDICAL PHYSICS, 2016, 43 (12) :6507-6516
[5]  
Dewaraja YK, 2000, J NUCL MED, V41, P123
[6]  
Ferreira T., 2012, IMAGE J USER GUID 1
[7]   Dual-window scatter correction and energy window setting in cerebral blood flow SPECT:: a Monte Carlo study [J].
Gustafsson, A ;
Ärlig, Å ;
Jacobsson, L ;
Ljungberg, M ;
Wikkelsö, C .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (11) :3431-3440
[8]   Uncertainty propagation for SPECT/CT-based renal dosimetry in 177Lu peptide receptor radionuclide therapy [J].
Gustafsson, Johan ;
Brolin, Gustav ;
Cox, Maurice ;
Ljungberg, Michael ;
Johansson, Lena ;
Gleisner, Katarina Sjogreen .
PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (21) :8329-8346
[9]   Validation of the GATE Monte Carlo simulation platform for modelling a CsI(Tl) scintillation camera dedicated to small-animal imaging [J].
Lazaro, D ;
Buvat, I ;
Loudos, G ;
Strul, D ;
Santin, G ;
Giokaris, N ;
Donnarieix, D ;
Maigne, L ;
Spanoudaki, V ;
Styliaris, S ;
Staelens, S ;
Breton, V .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (02) :271-285
[10]   A MONTE-CARLO PROGRAM FOR THE SIMULATION OF SCINTILLATION CAMERA CHARACTERISTICS [J].
LJUNGBERG, M ;
STRAND, SE .
COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 1989, 29 (04) :257-272