Feasibility of Deep Learning-Based PET/MR Attenuation Correction in the Pelvis Using Only Diagnostic MR Images

被引:42
作者
Bradshaw, Tyler J. [1 ]
Zhao, Gengyan [2 ]
Jang, Hyungseok [3 ]
Liu, Fang [1 ]
McMillan, Alan B. [1 ]
机构
[1] Univ Wisconsin, Sch Med & Publ Hlth, Dept Radiol, Madison, WI 53706 USA
[2] Univ Wisconsin, Sch Med & Publ Hlth, Dept Med Phys, Madison, WI 53706 USA
[3] Univ Calif San Diego, Dept Radiol, San Diego, CA 92103 USA
基金
美国国家卫生研究院;
关键词
PET/MR; attenuation correction; deep learning; QUANTITATIVE-EVALUATION; CT IMAGES; SEGMENTATION; GENERATION; TIME;
D O I
10.18383/j.tom.2018.00016
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This study evaluated the feasibility of using only diagnostically relevant magnetic resonance (MR) images together with deep learning for positron emission tomography (PET)/MR attenuation correction (deepMRAC) in the pelvis. Such an approach could eliminate dedicated MRAC sequences that have limited diagnostic utility but can substantially lengthen acquisition times for multibed position scans. We used axial T2 and T1 LAVA Flex magnetic resonance imaging images that were acquired for diagnostic purposes as inputs to a 3D deep convolutional neural network. The network was trained to produce a discretized (air, water, fat, and bone) substitute computed tomography (CT) (CTsub). Discretized (CTref-discrete) and continuously valued (CTref) reference CT images were created to serve as ground truth for network training and attenuation correction, respectively. Training was performed with data from 12 subjects. CTsub, CTref, and the system MRAC were used for PET/MR attenuation correction, and quantitative PET values of the resulting images were compared in 6 test subjects. Overall, the network produced CTsub with Dice coefficients of 0.79 +/- 0.03 for cortical bone, 0.98 +/- 0.01 for soft tissue (fat: 0.94 +/- 0.0; water: 0.88 +/- 0.02), and 0.49 +/- 0.17 for bowel gas when compared with CTref-discrete. The root mean square error of the whole PET image was 4.9% by using deepMRAC and 11.6% by using the system MRAC. In evaluating 16 soft tissue lesions, the distribution of errors for maximum standardized uptake value was significantly narrower using deepMRAC (-1.0% +/- 1.3%) than using system MRAC method (0.0% +/- 6.4%) according to the Brown-Forsy the test (P < .05). These results indicate that improved PET/MR attenuation correction can be achieved in the pelvis using only diagnostically relevant MR images.
引用
收藏
页码:138 / 147
页数:10
相关论文
共 33 条
  • [1] [Anonymous], EUR J NUCL MED MOL I
  • [2] [Anonymous], ENS MULT MOD ARCH RO
  • [3] [Anonymous], MED PHYS
  • [4] Whole-body bone segmentation from MRI for PET/MRI attenuation correction using shape- based averaging
    Arabi, Hossein
    Zaidi, Habib
    [J]. MEDICAL PHYSICS, 2016, 43 (11) : 5848 - 5861
  • [5] Quantitative Evaluation of Segmentation- and Atlas-Based Attenuation Correction for PET/MR on Pediatric Patients
    Bezrukov, Ilja
    Schmidt, Holger
    Gatidis, Sergios
    Mantlik, Frederic
    Schaefer, Juergen F.
    Schwenzer, Nina
    Pichler, Bernd J.
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2015, 56 (07) : 1067 - 1074
  • [6] Simultaneous carotid PET/MR: feasibility and improvement of magnetic resonance-based attenuation correction
    Bini, Jason
    Eldib, Mootaz
    Robson, Philip M.
    Calcagno, Claudia
    Fayad, Zahi A.
    [J]. INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2016, 32 (01) : 61 - 71
  • [7] Delso G, 2011, J NUCL MED, V52, P1914, DOI 10.2967/jnumed.111.092726
  • [8] 3D deeply supervised network for automated segmentation of volumetric medical images
    Dou, Qi
    Yu, Lequan
    Chen, Hao
    Jin, Yueming
    Yang, Xin
    Qin, Jing
    Heng, Pheng-Ann
    [J]. MEDICAL IMAGE ANALYSIS, 2017, 41 : 40 - 54
  • [9] Automatic Substitute Computed Tomography Generation and Contouring for Magnetic Resonance Imaging (MRI)-Alone External Beam Radiation Therapy From Standard MRI Sequences
    Dowling, Jason A.
    Sun, Jidi
    Pichler, Peter
    Rivest-Henault, David
    Ghose, Soumya
    Richardson, Haylea
    Wratten, Chris
    Martin, Jarad
    Arm, Jameen
    Best, Leah
    Chandra, Shekhar S.
    Fripp, Jurgen
    Menk, Frederick W.
    Greer, Peter B.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2015, 93 (05): : 1144 - 1153
  • [10] Towards quantitative PET/MRI: a review of MR-based attenuation correction techniques
    Hofmann, Matthias
    Pichler, Bernd
    Schoelkopf, Bernhard
    Beyer, Thomas
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2009, 36 : 93 - 104