Real-time assessment of flow reversal in an eccentric arterial stenotic model

被引:16
作者
Ai, Lisong [1 ]
Zhang, Lequan [1 ]
Dai, Wangde [2 ]
Hu, Changhong [1 ]
Shung, K. Kirk [1 ]
Hsiai, Tzung K. [1 ]
机构
[1] Univ So Calif, Dept Biomed Engn & Cardiovasc Med, Los Angeles, CA 90089 USA
[2] Hosp Good Samaritan, Inst Heart, Los Angeles, CA 90017 USA
关键词
PW Doppler; Atherosclerotic plaque; Shear stress; CFD; MEMS sensors; OSCILLATORY SHEAR-STRESS; MANGANESE SUPEROXIDE-DISMUTASE; BLOOD-FLOW; INTRAVASCULAR ULTRASOUND; SPATIAL VARIATIONS; ENDOTHELIAL-CELLS; MONOCYTE ADHESION; PULSATILE FLOW; NADPH OXIDASE; IN-VIVO;
D O I
10.1016/j.jbiomech.2010.06.021
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Plaque rupture is the leading cause of acute coronary syndromes and stroke. Plaque formation, otherwise known as stenosis, preferentially occurs in the regions of arterial bifurcation or curvatures. To date, real-time assessment of stenosis-induced flow reversal remains a clinical challenge. By interfacing microelectromechanical system (MEMS) thermal sensors with the high frequency pulsed wave (PW) Doppler ultrasound, we proposed to assess flow reversal in the presence of an eccentric stenosis. We developed a 3-D stenotic model (inner diameter of 6 mm, an eccentric stenosis with a height of 2.75 mm, and width of 21 mm) simulating a superficial arterial vessel. We demonstrated that heat transfer from the sensing element (2 x 80 mu m(2)) to the flow field peaked as a function of flow rates at the throat of the stenosis along the center/midline of arterial model, and dropped downstream from the stenosis, where flow reversal was detected by the high frequency ultrasound device at 45 MHz. Computational fluid dynamics (CFD) codes are in agreement with the ultrasound-acquired flow profiles upstream, downstream, and at the throat of the stenosis. Hence, we characterized regions of eccentric stenosis in terms of changes in heat transfer along the midline of vessel and identified points of flow reversal with high spatial and temporal resolution. (C) 2010 Elsevier Ltd. All rights reserved.
引用
收藏
页码:2678 / 2683
页数:6
相关论文
共 45 条
[1]   Shear stress influences spatial variations in vascular Mn-SOD expression: implication for LDL nitration [J].
Ai, Lisong ;
Rouhanizadeh, Mahsa ;
Wu, Joseph C. ;
Takabe, Wakako ;
Yu, Hongyu ;
Alavi, Mohammad ;
Li, Rongsong ;
Chu, Yi ;
Miller, Jordan ;
Heistad, Donald D. ;
Hsiai, Tzung K. .
AMERICAN JOURNAL OF PHYSIOLOGY-CELL PHYSIOLOGY, 2008, 294 (06) :C1576-C1585
[2]   Real-Time Intravascular Shear Stress in the Rabbit Abdominal Aorta [J].
Ai, Lisong ;
Yu, Hongyu ;
Dai, Wangde ;
Hale, Sharon L. ;
Kloner, Robert A. ;
Hsiai, Tzung K. .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2009, 56 (06) :1755-1764
[3]   Optimization of intravascular shear stress assessment in vivo [J].
Ai, Lisong ;
Yu, Hongyu ;
Takabe, Wakako ;
Paraboschi, Anna ;
Yu, Fei ;
Kim, E. S. ;
Li, Rongsong ;
Hsiai, Tzung K. .
JOURNAL OF BIOMECHANICS, 2009, 42 (10) :1429-1437
[4]   Flows in stenotic vessels [J].
Berger, SA ;
Jou, LD .
ANNUAL REVIEW OF FLUID MECHANICS, 2000, 32 :347-382
[5]   A high-frequency pulsed-wave doppler ultrasound system for the detection and imaging of blood flow in the microcirculation [J].
Christopher, DA ;
Burns, PN ;
Starkoski, BG ;
Foster, FS .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1997, 23 (07) :997-1015
[6]  
DAVIES PF, 1986, P NATL ACAD SCI USA, V83, P2114, DOI 10.1073/pnas.83.7.2114
[7]   Experimental and numerical study of pulsatile flows through stenosis: Wall shear stress analysis [J].
Deplano, V ;
Siouffi, M .
JOURNAL OF BIOMECHANICS, 1999, 32 (10) :1081-1090
[8]  
ERMOLAEV YA, 1974, B EXP BIOL MED+, V77, P6
[9]  
Evans D.H., 2000, Doppler Ultrasound - physics, instrumentation and clinical applications
[10]   MECHANISMS LEADING TO MYOCARDIAL-INFARCTION - INSIGHTS FROM STUDIES OF VASCULAR BIOLOGY [J].
FUSTER, V .
CIRCULATION, 1994, 90 (04) :2126-2146