An assessment of a method of inflating cuffs with a nitrous oxide gas mixture to prevent an increase in intracuff pressure in five different tracheal tube designs

被引:13
作者
Karasawa, F [1 ]
Tokunaga, M [1 ]
Aramaki, Y [1 ]
Shizukuishi, M [1 ]
Satoh, T [1 ]
机构
[1] Natl Def Med Coll, Dept Anaesthesiol, Tokorozawa, Saitama 359, Japan
关键词
anaesthetics; gases; nitrous oxide; equipment; cuffs; tracheal; intubation; technique;
D O I
10.1046/j.1365-2044.2001.01792.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We assessed the efficacy of inflating cuffs with a nitrous oxide gas mixture to minimise changes in intracuff pressure during anaesthesia. Patients were randomly assigned to one of five groups of 15 subjects each, and the trachea was intubated with the Profile Soft-Seal Cuff, Hi-Contour, Reinforced, Sheridan or Trachelon tracheal tubes. Cuffs were inflated with 40% nitrous oxide and cuff pressure was measured during anaesthesia with 67% nitrous oxide. Concentration of nitrous oxide in the cuff was measured at the end of anaesthesia. Cuff pressure increased slightly but significantly in the Reinforced and Trachelon groups. Nitrous oxide concentration in the Reinforced, Sheridan, or Trachelon groups was slightly but significantly higher than that in the Profile or Hi-Contour groups. Cuff pressure never exceeded 22 mmHg and there were no air leaks. Therefore, inflating cuffs with 40% N2O preserves stable cuff pressure in all five tracheal tubes, despite differences in cuff and pilot balloon design.
引用
收藏
页码:155 / 159
页数:5
相关论文
共 18 条
  • [1] BRANDT L, 1983, ANAESTHESIST, V32, P459
  • [2] CAPAN LM, 1983, ANESTHESIOLOGY, V59, P202
  • [3] A NEW ENDOTRACHEAL-TUBE WITH A CUFF IMPERVIOUS TO NITROUS-OXIDE - CONSTANCY OF CUFF PRESSURE AND VOLUME
    FUJIWARA, M
    MIZOGUCHI, H
    KAWAMURA, J
    HAYASHI, I
    TAKAKUWA, R
    MORINAGA, N
    YANAGAWA, T
    CHIBA, T
    ODASHIRO, M
    [J]. ANESTHESIA AND ANALGESIA, 1995, 81 (05) : 1084 - 1086
  • [4] Hakim M., 1993, ACTA ANAESTH SCAND, V37, P250
  • [5] The effect on intracuff pressure of various nitrous oxide concentrations used for inflating an endotracheal tube cuff
    Karasawa, F
    Ohshima, T
    Takamatsu, I
    Ehata, T
    Fukuda, I
    Uchihashi, Y
    Satoh, T
    [J]. ANESTHESIA AND ANALGESIA, 2000, 91 (03) : 708 - 713
  • [6] KARASAWA F, IN PRESS ANESTHESIA
  • [7] KIM JM, 1980, ANESTH ANALG, V59, P291
  • [8] A NEW TRACHEOSTOMY TUBE .3. BRONCHOFIBEROPTIC EXAMINATION OF THE TRACHEA AFTER PROLONGED INTUBATION WITH THE NL TRACHEOSTOMY TUBE
    LOMHOLT, N
    BORGESKOV, S
    KIRKBY, B
    [J]. ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1981, 25 (05) : 407 - 411
  • [9] MANDOE H, 1992, ANESTH ANALG, V74, P897
  • [10] Choice of cuff inflation medium during nitrous oxide anaesthesia
    Mitchell, V
    Adams, T
    Calder, I
    [J]. ANAESTHESIA, 1999, 54 (01) : 32 - 36