In Vitro Examination of the HeartWare CircuLite Ventricular Assist Device in the Fontan Connection

被引:6
作者
Tree, Michael [1 ]
Trusty, Philip M. [2 ]
Slesnick, Timoth C. [3 ]
Yoganathan, Ajit [2 ]
Deshpande, Shriprasad [3 ]
Maher, Kevin [3 ]
机构
[1] Georgia Inst Technol, Dept Mech Engn, Atlanta, GA 30332 USA
[2] Georgia Inst Technol, Dept Biomed Engn, Atlanta, GA 30332 USA
[3] Emory Univ, Sch Med, Dept Pediat, Childrens Hlth Care Atlanta, Atlanta, GA USA
关键词
in vitro; HeartWare CircuLite; ventricular assist device; Fontan; FAILING FONTAN; TRANSPLANTATION; CIRCULATION; MANAGEMENT; FAILURE; SUPPORT; BRIDGE;
D O I
10.1097/MAT.0000000000000521
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The failing Fontan physiology may benefit from ventricular assist device (VAD) mechanical circulatory support, although a subpulmonary VAD placed at the Fontan connection has never successfully supported the Fontan circulation long term. The HeartWare CircuLite continuous flow VAD was examined for Fontan circulatory support in an in vitro mock circulation. The VAD was tested in three different scenarios: VAD in parallel, baffle restricted VAD in parallel, and VAD in series. Successful support was defined as simultaneous decrease in inferior vena cava (IVC) pressure of 5 mm Hg or more and an increase in cardiac output (CO) to 4.25 L/min or greater. The VAD in parallel scenario resulted in a CO decrease to 3.46 L/min and 2.22 mm Hg decrease in IVC pressure. The baffle restricted VAD in parallel scenario resulted in a CO increase to 3.9 L/min increase in CO and 20.5 mm Hg decrease in IVC pressure (at 90% restriction). The VAD in series scenario resulted in a CO of 1.75 L/min and 5.9 mm Hg decrease in IVC pressure. We successfully modeled chronic failing Fontan physiology using patient-specific hemodynamic and anatomic data. Although unsuccessful in supporting Fontan patients as defined here, the HeartWare CircuLite VAD demonstrates the possibility to reduce Fontan pressure and increase CO with a VAD in the Fontan connection. This study provides insight into pump performance and design issues when attempting to support Fontan circulation. Refinements in VAD design with specific parameters to help support this patient population is the subject of our future work.
引用
收藏
页码:482 / 489
页数:8
相关论文
共 19 条
  • [1] Heart Transplantation for the Failing Fontan
    Backer, Carl L.
    Russell, Hyde M.
    Pahl, Elfriede
    Monge, Michael C.
    Gambetta, Katheryn
    Kindel, Steven J.
    Gossett, Jeffrey G.
    Hardy, Courtney
    Costello, John M.
    Deal, Barbara J.
    [J]. ANNALS OF THORACIC SURGERY, 2013, 96 (04) : 1413 - 1419
  • [2] Performance evaluation of a pediatric viscous impeller pump for Fontan cavopulmonary assist
    Giridharan, Guruprasad A.
    Koenig, Steven C.
    Kennington, Jeffrey
    Sobieski, Michael A.
    Chen, Jun
    Frankel, Steven H.
    Rodefeld, Mark D.
    [J]. JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2013, 145 (01) : 249 - 257
  • [3] Goldberg DJ, 2011, EXPERT REV CARDIOVAS, V9, P785, DOI [10.1586/erc.11.75, 10.1586/ERC.11.75]
  • [4] Hall JE, 2011, TXB MED PHYSL
  • [5] Hemodynamic Phenotype of the Failing Fontan in an Adult Population
    Hebson, Camden L.
    McCabe, Nancy M.
    Elder, Robert W.
    Mahle, William T.
    McConnell, Michael
    Kogon, Brian E.
    Veledar, Emir
    Jokhadar, Maan
    Vincent, Robert N.
    Sahu, Anurag
    Book, Wendy M.
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 2013, 112 (12) : 1943 - 1947
  • [6] Successful Bridge Through Transplantation With Berlin Heart Ventricular Assist Device in a Child With Failing Fontan
    Hoganson, David M.
    Boston, Umar S.
    Gazit, Avihu Z.
    Canter, Charles E.
    Eghtesady, Pirooz
    [J]. ANNALS OF THORACIC SURGERY, 2015, 99 (02) : 707 - 709
  • [7] Protein-Losing Enteropathy and the Fontan Operation
    Johnson, Jonathan N.
    Driscoll, David J.
    O'Leary, Patrick W.
    [J]. NUTRITION IN CLINICAL PRACTICE, 2012, 27 (03) : 375 - 384
  • [8] Clinical benefits of partial circulatory support in New York Heart Association Class IIIB and Early Class IV patients
    Meyns, Bart P.
    Simon, Andre
    Klotz, Stefan
    Wittwer, Thorsten
    Schlensak, Christian
    Rega, Filip
    Burkhoff, Daniel
    [J]. EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2011, 39 (05) : 693 - 698
  • [9] Does CircuLite Synergy assist device as partial ventricular support have a place in modern management of advanced heart failure?
    Mohite, Prashant N.
    Sabashnikov, Anton
    Simon, Andre Ruediger
    Weymann, Alexander
    Patil, Nikhil Prakash
    Unsoeld, Bernhard
    Bireta, Christian
    Popov, Aron Frederik
    [J]. EXPERT REVIEW OF MEDICAL DEVICES, 2015, 12 (01) : 49 - 60
  • [10] Beyond a Broken Heart: Circulatory Dysfunction in the Failing Fontan
    Mori, Makoto
    Aguirre, Alfredo J.
    Elder, Robert W.
    Kashkouli, Ali
    Farris, Alton Brad
    Ford, Ryan M.
    Book, Wendy M.
    [J]. PEDIATRIC CARDIOLOGY, 2014, 35 (04) : 569 - 579