Acute Lead Dislodgements and In-Hospital Mortality in Patients Enrolled in the National Cardiovascular Data Registry Implantable Cardioverter Defibrillator Registry

被引:68
作者
Cheng, Alan [1 ]
Wang, Yongfei [2 ]
Curtis, Jeptha P. [2 ]
Varosy, Paul D. [3 ,4 ]
机构
[1] Johns Hopkins Univ, Sch Med, Sect Cardiac Electrophysiol, Div Cardiol, Baltimore, MD 21287 USA
[2] Yale Univ, Sch Med, Dept Internal Med, Sect Cardiovasc Med, New Haven, CT 06510 USA
[3] VA Eastern Colorado Hlth Care Syst, Aurora, CO USA
[4] Univ Colorado Denver, Sch Med, Aurora, CO USA
关键词
death; implantable cardioverter-defibrillator; lead dislodgements; National Cardiovascular Data Registry; HEART-FAILURE; COMPLICATIONS; PERFORATION; PACEMAKER;
D O I
10.1016/j.jacc.2010.06.037
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives We sought to describe the incidence of acute lead dislodgements and the consequences of these events in patients enrolled in the National Cardiovascular Data Registry (NCDR) Implantable Cardioverter-Defibrillator (ICD) Registry. Background Lead dislodgements are common adverse events in patients undergoing ICD implants. Little is known regarding who is at risk and the consequences of these events. Methods Patients enrolled between April 2006 and September 2008 were included. Acute lead dislodgement was defined as movement of the lead requiring another procedure for repositioning before discharge. Results Acute dislodgement occurred in 2,628 of 226,764 patients. Univariate variables associated with dislodgements included older age, female sex, and patients with atrial fibrillation, chronic lung disease, cerebrovascular disease, nonischemic cardiomyopathy, and lower ejection fractions (all p < 0.002). After multivariate adjustment, factors associated with an increased risk for dislodgement included New York Heart Association functional class IV heart failure, atrial fibrillation/flutter, having a cardiac resynchronization therapy-defibrillator device, and procedures performed by physicians trained under alternative pathways. A teaching/training hospital setting was not a factor (p = 0.64). Acute dislodgements had increased odds for other adverse events including cardiac arrest, cardiac tamponade, device infection, pneumothorax, and in-hospital death even after adjustment for potential confounders (all p < 0.0001). Conclusions Acute lead dislodgements occur more often in patients with more comorbidities and in patients undergoing implants by nonelectrophysiology-trained implanters. These events were strongly associated with increased odds for in-hospital death. (J Am Coll Cardiol 2010;56:1651-6) (C) 2010 by the American College of Cardiology Foundation
引用
收藏
页码:1651 / 1656
页数:6
相关论文
共 13 条
[1]   The relation between patients' outcomes and the volume of cardioverter-defibrillator implantation procedures performed by physicians treating medicare beneficiaries [J].
Al-Khatib, SM ;
Lucas, FL ;
Jollis, JG ;
Malenka, DJ ;
Wennberg, DE .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2005, 46 (08) :1536-1540
[2]   Complications of pacemakers and defibrillators in the elderly [J].
Bailey, SM ;
Wilkoff, BL .
AMERICAN JOURNAL OF GERIATRIC CARDIOLOGY, 2006, 15 (02) :102-107
[3]   Early left ventricular lead dislodgement related to hyperpnea respiration [J].
Chan, JYS ;
Fung, JWH ;
Yu, CM .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2006, 29 (04) :425-426
[4]   Clinical competency statement: Training pathways for implantation of cardioverter defibrillators and cardiac resynchronization devices [J].
Curtis, AB ;
Ellenbogen, KA ;
Hammill, SC ;
Hayes, DL ;
Reynolds, DW ;
Wilber, DJ ;
Cain, ME .
HEART RHYTHM, 2004, 1 (03) :371-375
[5]   Increased incidence of subacute lead perforation noted with one implantable cardioverter-defibrillator [J].
Danik, Stephan B. ;
Mansour, Moussa ;
Singh, Jagmeet ;
Reddy, Vivek Y. ;
Ellinor, Patrick T. ;
Milan, David ;
Heist, E. Kevin ;
d'Avila, Andre ;
Ruskin, Jeremy N. ;
Mela, Theofanie .
HEART RHYTHM, 2007, 4 (04) :439-442
[6]   The additional hospital costs generated in the management of complications of pacemaker and defibrillator implantations [J].
Ferguson, TB ;
Ferguson, CL ;
Crites, K ;
CrimminsReda, P .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1996, 111 (04) :742-752
[7]   Patient, procedural, and hardware factors associated with pacemaker lead failures in pediatrics and congenital heart disease [J].
Fortescue, EB ;
Berul, CI ;
Cecchin, F ;
Walsh, EP ;
Triedman, JK ;
Alexander, ME .
HEART RHYTHM, 2004, 1 (02) :150-159
[8]   NATIONAL ICD REGISTRY ANNUAL REPORT 2008 Review of the ICD Registry's Third Year, Expansion to include Lead Data and Pediatric ICD Procedures, and Role for Measuring Performance [J].
Hammill, Stephen C. ;
Kremers, Mark S. ;
Kadish, Alan H. ;
Stevenson, Lynne Warner ;
Heidenreich, Paul A. ;
Lindsay, Bruce D. ;
Mirro, Michael J. ;
Radford, Martha J. ;
McKay, Chuck ;
Wang, Yongfei ;
Lang, Christine M. ;
Pontzer, Kathryn ;
Rumsfed, John ;
Phurrough, Steve E. ;
Curtis, Jeptha P. ;
Brindis, Raph G. .
HEART RHYTHM, 2009, 6 (09) :1397-1401
[9]   Evaluation of Early Complications Related to De Novo Cardioverter Defibrillator Implantation Insights From the Ontario ICD Database [J].
Lee, Douglas S. ;
Krahn, Andrew D. ;
Healey, Jeffrey S. ;
Birnie, David ;
Crystal, Eugene ;
Dorian, Paul ;
Simpson, Christopher S. ;
Khaykin, Yaariv ;
Cameron, Douglas ;
Janmohamed, Amir ;
Yee, Raymond ;
Austin, Peter C. ;
Chen, Zhongliang ;
Hardy, Judy ;
Tu, Jack V. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2010, 55 (08) :774-782
[10]   Safety of transvenous cardiac resynchronization system implantation in patients with chronic heart failure -: Combined results of over 2,000 patients from a multicenter study program [J].
León, AR ;
Abraham, WT ;
Curtis, AB ;
Daubert, JP ;
Fisher, WG ;
Gurley, J ;
Hayes, DL ;
Lieberman, R ;
Petersen-Stejskal, S ;
Wheelan, K .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2005, 46 (12) :2348-2356