Biomechanical comparison between stainless steel, titanium and carbon-fiber reinforced polyetheretherketone volar locking plates for distal radius fractures

被引:47
作者
Mugnai, Raffaele [1 ]
Tarallo, Luigi [1 ]
Capra, Francesco
Catani, Fabio [1 ]
机构
[1] Modena Univ Hosp, Orthopaed & Traumatol Dept, Via Pietro Giardini 1355, I-41126 Modena, MO, Italy
关键词
Distal radius; Plate; Bending stiffness; Load to failure; Biomechanical; Material; ORTHOPEDIC-SURGERY; COMPOSITE-MATERIAL; INTERNAL-FIXATION; SHAFT FRACTURES; PLASTIC PLATES; SCREW DESIGN; BONE; IMPLANTS; COMPRESSION; STABILITY;
D O I
10.1016/j.otsr.2018.05.002
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction: As the popularity of volar locked plate fixation for distal radius fractures has increased, so have the number and variety of implants, including variations in plate design, the size and angle of the screws, the locking screw mechanism, and the material of the plates. Hypothesis: Carbon-fiber reinforced polyetheretherketone (CFR-PEEK) plate features similar biomechanical properties to metallic plates, representing, therefore, an optimal alternative for the treatment of distal radius fractures. Materials and Methods: Three different materials-composed plates were evaluated: stainless steel volar lateral column (Zimmer); titanium DVR (Hand Innovations); CFR-PEEK DiPHOS-RM (Lima Corporate). Six plates for each type were implanted in sawbones and an extra-articular rectangular osteotomy was created. Three plates for each material were tested for load to failure and bending stiffness in axial compression. Moreover, 3 constructs for each plate were evaluated after dynamically loading for 6000 cycles of fatigue. Results: The mean bending stiffness pre-fatigue was significantly higher for the stainless steel plate. The titanium plate yielded the higher load to failure both pre and post fatigue. After cyclic loading, the bending stiffness increased by a mean of 24% for the stainless steel plate; 33% for the titanium; and 17% for the CFR-PEEK plate. The mean load to failure post-fatigue increased by a mean of 10% for the stainless steel and 14% for CFR-PEEK plates, whereas it decreased ( -16%) for the titanium plate. Statistical analysis between groups reported significant values (p < 001) for all comparisons except for Hand Innovations vs. Zimmer bending stiffness post fatigue (p = .197). Discussion: The significant higher load to failure of the titanium plate, makes it indicated for patients with higher functional requirements or at higher risk of trauma in the post-operative period. The CFR-PEEK plate showed material-specific disadvantages, represented by little tolerance to plastic deformation, and lower load to failure. (C) 2018 Elsevier Masson SAS. All rights reserved.
引用
收藏
页码:877 / 882
页数:6
相关论文
共 36 条
[11]  
Hart A, 2015, J ORTHOP TRAUMA, V29, pE1, DOI 10.1097/BOT.0000000000000146
[12]   BONE-DENSITY AFTER RIGID PLATE FIXATION OF TIBIAL FRACTURES - A DUAL-ENERGY X-RAY ABSORPTIOMETRY STUDY [J].
JANES, GC ;
COLLOPY, DM ;
PRICE, R ;
SIKORSKI, JM .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1993, 75 (06) :914-917
[13]   Stability of volar locking plate systems for AO type C3 fractures of the distal radius: biomechanical study in a cadaveric model [J].
Kamei, Shuzo ;
Osada, Denju ;
Tamai, Kazuya ;
Kato, Nakayuki ;
Takai, Morimitsu ;
Kameda, Masahiro ;
Nohara, Yutaka .
JOURNAL OF ORTHOPAEDIC SCIENCE, 2010, 15 (03) :357-364
[14]   Influence of plate material and screw design on stiffness and ultimate load of locked plating in osteoporotic proximal humeral fractures [J].
Katthagen, Jan Christoph ;
Schwarze, Michael ;
Warnhoff, Mara ;
Voigt, Christine ;
Hurschler, Christof ;
Lill, Helmut .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2016, 47 (03) :617-624
[15]  
Kettunen Jukka, 2003, J Orthop Sci, V8, P772, DOI 10.1007/s00776-003-0718-7
[16]   Biomechanical Properties of 3-Dimensional Printed Volar Locking Distal Radius Plate: Comparison With Conventional Volar Locking Plate [J].
Kim, Sung-Jae ;
Jo, Young-Hoon ;
Choi, Wan-Sun ;
Lee, Chang-Hun ;
Lee, Bong-Gun ;
Kim, Joo-Hak ;
Lee, Kwang-Hyun .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2017, 42 (09) :747.e1-747.e6
[17]   The Use of Carbon-Fiber-Reinforced (CFR) PEEK Material in Orthopedic Implants: A Systematic Review [J].
Li, Chuan Silvia ;
Vannabouathong, Christopher ;
Sprague, Sheila ;
Bhandari, Mohit .
CLINICAL MEDICINE INSIGHTS-ARTHRITIS AND MUSCULOSKELETAL DISORDERS, 2015, 8 :33-45
[18]   Biomechanical Comparison of Volar Fixed-Angle Locking Plates for AO C3 Distal Radius Fractures: Titanium Versus Stainless Steel With Compression [J].
Marshall, Tyler ;
Momaya, Amit ;
Eberhardt, Alan ;
Chaudhari, Nilesh ;
Hunt, Thomas R., III .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2015, 40 (10) :2032-2038
[19]   A Biomechanical Evaluation of a Hybrid Dynamic Compression Plate and a CastLess Arthrodesis Plate for Pancarpal Arthrodesis in Dogs [J].
Meeson, Richard L. ;
Goodship, Allen E. ;
Arthurs, Gareth I. .
VETERINARY SURGERY, 2012, 41 (06) :738-744
[20]   Plate and screw design in fractures of the hand and wrist [J].
Mudgal, Chaitanya S. ;
Jupiter, Jesse B. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2006, (445) :68-80