Patient and healthcare professional satisfaction with a new, high accuracy blood glucose meter with color range indicator and wireless connectivity

被引:4
|
作者
Katz, Laurence B. [1 ]
Grady, Mike [2 ]
Stewart, Lorna [2 ]
Cameron, Hilary [3 ]
机构
[1] LifeScan Inc, Clin Affairs, Wayne, PA USA
[2] LifeScan Scotland Ltd, Clin Affairs Clin Operat, Inverness, Scotland
[3] LifeScan Scotland Ltd, Biostat, Inverness, Scotland
关键词
Accuracy; blood glucose; blood glucose meter; blood glucose strips; self-monitoring of blood glucose; wireless; PERFORMANCE; TRIAL;
D O I
10.1080/17434440.2016.1195260
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Accurate self-monitoring of blood glucose is a key component of effective self-management of glycemic control. The OneTouch VerioFlex (OTVF) blood glucose monitoring system (BGMS) was evaluated for accuracy in a clinical setting. Patients also used OTVF for a 1-wk trial period and reported their level of satisfaction with meter features. In a separate study, healthcare professionals used an on-line simulator of the BGMS and answered questions about its potential utility to their patients. OTVF was accurate over a wide glucose range and met lay user and system accuracy blood glucose standards described in ISO15197:2013 as well as the accuracy requirements to fulfill US FDA expectations for 510(k) clearance of BGMS. Patients and healthcare professionals felt the features of OTVF, which has the capability to connect wirelessly to mobile devices and interact wirelessly with diabetes management software, could provide significant benefits to them or their patients.
引用
收藏
页码:619 / 626
页数:8
相关论文
共 11 条
  • [11] Hematocrit and drug interference study of the new Arkray Glucocard S Onyx blood glucose meter meeting ISO 15197:2013 accuracy criteria - a contribution to patient safety
    Burrini, Claudio
    Morrocchi, Beatrice
    Giorgetti, Nicola
    Saeger, Beate
    EXPERT REVIEW OF MEDICAL DEVICES, 2025,