The calculation of dose enhancement close to platinum implants for skull radiography

被引:6
|
作者
Cheung, Joel Y. C.
Tang, Fuk-Hay [1 ]
机构
[1] Hong Kong Polytech Univ, Dept Hlth Technol & Informat, Hung Hum, Hong Kong, Peoples R China
[2] Princess Margaret Hosp, Dept Oncol, Hong Kong, Hong Kong, Peoples R China
来源
HEALTH PHYSICS | 2007年 / 93卷 / 04期
关键词
x rays; radiography; Monte Carlo; cancer;
D O I
10.1097/01.HP.0000264450.81683.51
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Materials with high atomic numbers experience the occurrence of the photoelectric effect when they are irradiated by low energy photons. A short range dose enhancement, due to the dominant photoelectric effect, close to platinum implants (Z = 78) in diagnostic radiography cannot be easily measured experimentally. The enhanced dose may increase the risk for adverse health effects from cancer or may damage vital brain structures close to the high atomic number implants. In the present work, Monte Carlo simulation using the LSCAT version of PRESTA EGS4 was employed to investigate the resulting dose enhancements. The results show that the highest estimated dose enhancement of 79% for brain tissues close to platinum implants was calculated for 65 kV x-ray energy and 180% for 120 kV x-ray energy.
引用
收藏
页码:267 / 272
页数:6
相关论文
共 50 条
  • [21] CALCULATION OF DOSE ENHANCEMENT FROM BACKSCATTERED ELECTRONS WITH ACCEPT AND EGS
    PALTA, JR
    DAFTARI, I
    AYYANGAR, K
    SUNTHARALINGAM, N
    MEDICAL PHYSICS, 1987, 14 (03) : 474 - 475
  • [22] An investigation of oxygen enhancement ratio modeling for proton dose calculation
    Denis Yoo
    Chang Ju Rah
    Erick Lee
    Ji Hwan Kim
    Jeong Hoon Park
    Byung Jun Min
    Journal of the Korean Physical Society, 2022, 81 : 1081 - 1090
  • [23] An investigation of oxygen enhancement ratio modeling for proton dose calculation
    Yoo, Denis
    Rah, Chang Ju
    Lee, Erick
    Kim, Ji Hwan
    Park, Jeong Hoon
    Min, Byung Jun
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2022, 81 (11) : 1081 - 1090
  • [24] Development of Monte Carlo Based Voxel Head Phantom for Dose Calculation in Periapical Radiography
    Oh, Y.
    Kang, C.
    An, S.
    Kim, J.
    Ji, Y.
    Lee, R.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [25] Development of a 3-D convolution/superposition algorithm for precise dose calculation in the skull
    Mack A.
    Weltz D.
    Scheib S.G.
    Wowra B.
    Böttcher H.
    Seifer V.
    Australasian Physics & Engineering Sciences in Medicine, 2006, 29 (1): : 1 - 12
  • [26] CALCULATION OF EFFECTIVE DOSE IN DIAGNOSTIC RADIOGRAPHY - COMPARISON OF THE CALCULATION METHODS PUBLISHED IN ICRP-26 AND ICRP-60
    KESKE, U
    HIERHOLZER, J
    MATSCHKE, S
    VENZ, S
    HOSTEN, N
    FELIX, R
    RADIOLOGY, 1995, 197 : 359 - 359
  • [27] COUPLED PHOTON-ELECTRON CALCULATION OF DOSE ENHANCEMENT IN THIN REGIONS
    DRESSER, MM
    GORDON, JD
    TRANSACTIONS OF THE AMERICAN NUCLEAR SOCIETY, 1976, 23 (JUN18): : 608 - 609
  • [28] MCPI©:: A sub-minute Monte Carlo dose calculation engine for prostate implants
    Chibani, O
    Williamson, JF
    MEDICAL PHYSICS, 2005, 32 (12) : 3688 - 3698
  • [29] A component method to delineate surgical spine implants for proton Monte Carlo dose calculation
    Chang, Chih-Wei
    Charyyev, Serdar
    Harms, Joseph
    Slopsema, Roelf
    Wolf, Jonathan
    Refai, Daniel
    Yoon, Tim
    McDonald, Mark W.
    Bradley, Jeffrey D.
    Leng, Shuai
    Zhou, Jun
    Yang, Xiaofeng
    Lin, Liyong
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2023, 24 (01):
  • [30] Dose calculation validation of AAA and Acuros XB algorithms in presence of titanium spinal implants
    Tait, J.
    Williams, T.
    Stacey, C. F.
    Hussein, M.
    Clark, C. H.
    Johnson, U.
    RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S1364 - S1365