Scalar Localization by Cone-Beam Computed Tomography of Cochlear Implant Carriers: A Comparative Study Between Straight and Periomodiolar Precurved Electrode Arrays

被引:98
作者
Boyer, Eric [1 ,2 ]
Karkas, Alexandre [1 ,2 ]
Attye, Arnaud [1 ,3 ]
Lefournier, Virginie [4 ]
Escude, Bernard [5 ]
Schmerber, Sebastien [1 ,2 ]
机构
[1] Univ Grenoble 1, Grenoble, France
[2] Univ Hosp Grenoble, Dept Otorhinolaryngol, F-38043 Grenoble 09, France
[3] CHU Grenoble, Dept Radiol, Univ Hosp, F-38043 Grenoble, France
[4] Grp Clin Mail, Grenoble, France
[5] Radiol Clin Pasteur, Toulouse, France
关键词
Cochlear implant; Cone-beam computed tomography; Dislocation; Scalar localization; TEMPORAL BONE; INSERTION TRAUMA; QUALITY-CONTROL; DESIGN; TOOL; POSITION; CONTOUR; MORPHOLOGY; STIFFNESS;
D O I
10.1097/MAO.0000000000000705
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Objective: To compare the incidence of dislocation of precurved versus straight flexible cochlear implant electrode arrays using cone-beam computed tomography (CBCT) image analyses. Study Design: Consecutive nonrandomized case-comparison study. Settings: Tertiary referral center. Patients: Analyses of patients' CBCT images after cochlear implant surgery. Intervention(s): Precurved and straight flexible electrode arrays from two different manufacturers were implanted. A round window insertion was performed in most cases. Two cases necessitated a cochleostomy. The patients' CBCT images were reconstructed in the coronal oblique, sagittal oblique, and axial oblique section. Main Outcome Measures: The insertion depth angle and the incidence of dislocation from the scala tympani to the scala vestibuli were determined. Results: The CBCT images and the incidence of dislocation were analyzed in 54 patients (61 electrode arrays). Thirty-one patients were implanted with a precurved perimodiolar electrode array and 30 patients with a straight flexible electrode array. A total of nine (15%) scalar dislocations were observed in both groups. Eight (26%) scalar dislocations were observed in the precurved array group and one (3%) in the straight array group. Dislocation occurred at an insertion depth angle between 170 and 190 degrees in the precurved array group and at approximately 370 degrees in the straight array group. Conclusion: With precurved arrays, dislocation usually occurs in the ascending part of the basal turn of the cochlea. With straight flexible electrode arrays, the incidence of dislocation was lower, and it seems that straight flexible arrays have a higher chance of a confined position within the scala tympani than perimodiolar precurved arrays.
引用
收藏
页码:422 / 429
页数:8
相关论文
共 41 条
  • [1] Arweiler-Harbeck D., 2012, ISRN OTOLARYNGOLOGY
  • [2] Quality control after cochlear implant surgery by means of rotational tomography
    Aschendorff, A
    Kubalek, T
    Turowski, B
    Zanella, T
    Hochmuth, T
    Schumacher, T
    Klenzner, T
    Laszig, R
    [J]. OTOLOGY & NEUROTOLOGY, 2005, 26 (01) : 34 - 37
  • [3] Quality control after insertion of the nucleus contour and contour advance electrode in adults
    Aschendorff, Antje
    Kromeier, Jan
    Klenzner, Thomas
    Laszig, Roland
    [J]. EAR AND HEARING, 2007, 28 (02) : 75S - 79S
  • [4] Flat-panel volume computed tomography for cochlear implant electrode array examination in isolated temporal bone specimens
    Bartling, Soenke H.
    Gupta, Rajiv
    Torkos, Attila
    Dullin, Christian
    Eckhardt, Grabbe
    Lenarz, Thomas
    Becker, Hartmut
    Stöver, Timo
    [J]. OTOLOGY & NEUROTOLOGY, 2006, 27 (04) : 491 - 498
  • [5] The Internal Dimensions of the Cochlear Scalae With Special Reference to Cochlear Electrode Insertion Trauma
    Biedron, Slavomir
    Prescher, Andreas
    Ilgner, Justus
    Westhofen, Martin
    [J]. OTOLOGY & NEUROTOLOGY, 2010, 31 (05) : 731 - 737
  • [6] Briggs Robert J S, 2006, Audiol Neurootol, V11 Suppl 1, P42, DOI 10.1159/000095613
  • [7] Cohen LT, 1996, AM J OTOL, V17, P859
  • [8] Coordes A, 2013, OTOL NEUROTOL, V34, P1027, DOI 10.1097/MAO.0b013e318280da2a
  • [9] High-resolution cone-beam computed tomography: a potential tool to improve atraumatic electrode design and position
    Cushing, Sharon L.
    Daly, Michael J.
    Treaba, Claudiu G.
    Chan, Harley
    Irish, Jonathan C.
    Blaser, Susan
    Gordon, Karen A.
    Papsin, Blake C.
    [J]. ACTA OTO-LARYNGOLOGICA, 2012, 132 (04) : 361 - 368
  • [10] Morphologic examination of the temporal bone by cone beam computed tomography: Comparison with multislice helical computed tomography
    Dahmani-Causse, M.
    Marx, M.
    Deguine, O.
    Fraysse, B.
    Lepage, B.
    Escude, B.
    [J]. EUROPEAN ANNALS OF OTORHINOLARYNGOLOGY-HEAD AND NECK DISEASES, 2011, 128 (05) : 230 - 235