Rotational artifacts in on-board cone beam computed tomography

被引:4
作者
Ali, E. S. M. [1 ]
Webb, R. [2 ]
Nyiri, B. J. [1 ,3 ]
机构
[1] Ottawa Hosp Canc Ctr, Dept Med Phys, Ottawa, ON, Canada
[2] Elekta, Montreal, PQ, Canada
[3] Univ Ottawa, Dept Radiol, Ottawa, ON, Canada
关键词
cone beam computed tomography; XVI; image artifacts; image lag; rotational errors; STEREOTACTIC BODY RADIOTHERAPY; SETUP ERRORS; QUALITY-ASSURANCE; CONSEQUENCES; SYSTEM;
D O I
10.1088/0031-9155/60/4/1461
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Rotational artifacts in image guidance systems lead to registration errors that affect non-isocentric treatments and dose to off-axis organs-at-risk. This study investigates a rotational artifact in the images acquired with the on-board cone beam computed tomography system XVI (Elekta, Stockholm, Sweden). The goals of the study are to identify the cause of the artifact, to characterize its dependence on other quantities, and to investigate possible solutions. A 30 cm diameter cylindrical phantom is used to acquire clockwise and counterclockwise scans at five speeds (120 to 360 deg min(-1)) on six Elekta linear accelerators from three generations (MLCi, MLCi2 and Agility). Additional scans are acquired with different pulse widths and focal spot sizes for the same mAs. Image quality is evaluated using a common phantom with an in-house three dimensional contrast transfer function attachment. A robust, operator-independent analysis is developed which quantifies rotational artifacts with 0.02 degrees accuracy and imaging system delays with 3 ms accuracy. Results show that the artifact is caused by mislabelling of the projections with a lagging angle due to various imaging system delays. For the most clinically used scan speed (360 deg min(-1)), the artifact is similar to 0.5 degrees, which corresponds to similar to 0.25 degrees error per scan direction with the standard Elekta procedure for angle calibration. This leads to a 0.5 mm registration error at 11 cm off-center. The artifact increases linearly with scan speed, indicating that the system delay is independent of scan speed. For the most commonly used pulse width of 40 ms, this delay is 34 +/- 1 ms, part of which is half the pulse width. Results are consistent among the three linac generations. A software solution that corrects the angles of individual projections is shown to eliminate the rotational error for all scan speeds and directions. Until such a solution is available from the manufacturer, three clinical solutions are presented, which reduce the rotational error without compromising image quality.
引用
收藏
页码:1461 / 1476
页数:16
相关论文
共 16 条
[1]   Quality assurance for image-guided radiation therapy utilizing CT-based technologies: A report of the AAPM TG-179 [J].
Bissonnette, Jean-Pierre ;
Balter, Peter A. ;
Dong, Lei ;
Langen, Katja M. ;
Lovelock, D. Michael ;
Miften, Moyed ;
Moseley, Douglas J. ;
Pouliot, Jean ;
Sonke, Jan-Jakob ;
Yoo, Sua .
MEDICAL PHYSICS, 2012, 39 (04) :1946-1963
[2]   Advances in image-guided radiation therapy [J].
Dawson, Laura A. ;
Jaffray, David A. .
JOURNAL OF CLINICAL ONCOLOGY, 2007, 25 (08) :938-946
[3]   A study on adaptive IMRT treatment planning using kV cone-beam CT [J].
Ding, George X. ;
Duggan, Dennis M. ;
Coffey, Charles W. ;
Deeley, Matthew ;
Hallahan, Dennis E. ;
Cmelak, Anthony ;
Malcolm, Arnold .
RADIOTHERAPY AND ONCOLOGY, 2007, 85 (01) :116-125
[4]   Dosimetric effects of patient rotational setup errors on prostate IMRT treatments [J].
Fu, Weihua ;
Yang, Yong ;
Li, Xiang ;
Heron, Dwight E. ;
Huq, M. Saiful ;
Yue, Ning J. .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (20) :5321-5331
[5]   Magnitude and clinical relevance of translational and rotational patient setup errors: A cone-beam CT study [J].
Guckenberger, M ;
Meyer, J ;
Vordermark, D ;
Baier, K ;
Wilbert, J ;
Flentje, M .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 65 (03) :934-942
[6]   INFLUENCE OF ROTATIONS ON DOSE DISTRIBUTIONS IN SPINAL STEREOTACTIC BODY RADIOTHERAPY (SBRT) [J].
Gutfeld, Orit ;
Kretzler, Annette E. ;
Kashani, Rojano ;
Tatro, Daniel ;
Balter, James M. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 73 (05) :1596-1601
[7]   Flat-panel cone-beam computed tomography for image-guided radiation therapy [J].
Jaffray, DA ;
Siewerdsen, JH ;
Wong, JW ;
Martinez, AA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 53 (05) :1337-1349
[8]   An image quality comparison study between XVI and OBI CBCT systems [J].
Kamath, Srijit ;
Song, William ;
Chvetsov, Alexei ;
Ozawa, Shuichi ;
Lu, Haibin ;
Samant, Sanjiv ;
Liu, Chihray ;
Li, Jonathan G. ;
Palta, Jatinder R. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2011, 12 (02) :376-390
[9]   Task Group 142 report: Quality assurance of medical accelerators [J].
Klein, Eric E. ;
Hanley, Joseph ;
Bayouth, John ;
Yin, Fang-Fang ;
Simon, William ;
Dresser, Sean ;
Serago, Christopher ;
Aguirre, Francisco ;
Ma, Lijun ;
Arjomandy, Bijan ;
Liu, Chihray ;
Sandin, Carlos ;
Holmes, Todd .
MEDICAL PHYSICS, 2009, 36 (09) :4197-4212
[10]   Dosimetric consequences of rotational setup errors with direct simulation in a treatment planning system for fractionated stereotactic radiotherapy [J].
Peng, Jean L. ;
Liu, Chihray ;
Chen, Yu ;
Amdur, Robert J. ;
Vanek, Kenneth ;
Li, Jonathan G. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2011, 12 (03) :61-70