A closed-loop reduced oxygen breathing device for inducing hypoxia in humans

被引:0
|
作者
Sausen, KP [1 ]
Bower, EA [1 ]
Stiney, ME [1 ]
Feigl, C [1 ]
Wartman, R [1 ]
Clark, JB [1 ]
机构
[1] USN, Aerosp Med Res Lab, Pensacola, FL 32508 USA
来源
AVIATION SPACE AND ENVIRONMENTAL MEDICINE | 2003年 / 74卷 / 11期
关键词
hypoxia; aerospace physiology; reduced oxygen breathing device; impedance cardiography; apnea; asthma; diving; flying;
D O I
暂无
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Background: Hypoxia poses a documented threat to aerospace and diving operations in healthy people and is a component of many clinical conditions. Practical training for aircrew, and research on clinically relevant hypoxic conditions frequently rely exclusively on large, expensive hypobaric chambers. Purpose: Here we describe and report the efficacy of a compact, economical, closed-loop rebreather-type reduced-oxygen breathing device (ROBD) for hypoxia induction in humans. Methods: Subjects were four healthy student Naval flight surgeons. During baseline, subjects breathed normoxic air (21% O-2, equivalent to sea-level). Baseline was followed by an altitude period, during which participants were exposed to hypoxic air (about 10% O-2, approximating 5,486 m [18,000 ft mean sea-level]) for 30 min followed by a normoxic recovery period. Subjects' peripheral arterial oxygen saturation (SpO(2)), BP, impedance cardiography, cardiac output, and systemic vascular resistance served as dependent measures along with the fraction of inspired oxygen (FIO2). Results: Circuit FIO2 and subjects' SpO(2) were significantly lower during the altitude period than the baseline and recovery periods. HR and cardiac output were significantly higher, and systemic vascular resistance was significantly lower, during hypoxic air exposure than during baseline or recovery. Conclusions: These data support the closed-loop ROBD as a potentially useful device for training and research involving acute hypoxia in healthy and clinical populations.
引用
收藏
页码:1190 / 1197
页数:8
相关论文
共 50 条
  • [1] Closed-loop control of analgesia in humans
    Gentilini, A
    Morari, M
    Bieniok, C
    Wymann, R
    Schnider, TW
    PROCEEDINGS OF THE 40TH IEEE CONFERENCE ON DECISION AND CONTROL, VOLS 1-5, 2001, : 861 - 866
  • [2] CLOSED-LOOP CONTROL FOR ANESTHESIA BREATHING SYSTEMS
    WESTENSKOW, DR
    WALLROTH, CF
    JOURNAL OF CLINICAL MONITORING, 1990, 6 (03): : 249 - 256
  • [3] A Closed-Loop Optogenetic Stimulation Device
    Edward, Epsy S.
    Kouzani, Abbas Z.
    ELECTRONICS, 2020, 9 (01)
  • [4] Identification of open-loop transfer functions in closed-loop baroreflex system using random breathing in humans
    Yingthawornsuk, T
    Kawada, T
    Sato, T
    Inagaki, M
    Sunagawa, K
    Cox, J
    Shiavi, RG
    Diedrich, A
    COMPUTERS IN CARDIOLOGY 2002, VOL 29, 2002, 29 : 461 - 464
  • [5] A closed-loop stepper motor waist-pull system for inducing protective stepping in humans
    Pidcoe, PE
    Rogers, MW
    JOURNAL OF BIOMECHANICS, 1998, 31 (04) : 377 - 381
  • [6] Perioperative closed-loop control of analgesia in humans
    Schaniel, C
    Gentilini, A
    Morari, M
    Bieniok, C
    Wymann, R
    Schnider, TW
    PROCEEDINGS OF THE 23RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4: BUILDING NEW BRIDGES AT THE FRONTIERS OF ENGINEERING AND MEDICINE, 2001, 23 : 449 - 449
  • [7] Portable Closed-loop Optogenetic Stimulation Device
    Edward, Epsy S.
    Kouzani, Abbas Z.
    Kale, Rajas P.
    Tye, Susannah J.
    2016 38TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2016, : 5250 - 5253
  • [8] CLoSES: A platform for closed-loop intracranial stimulation in humans
    Zelmann, Rina
    Paulk, Angelique C.
    Basu, Ishita
    Sarma, Anish
    Yousefi, Ali
    Crocker, Britni
    Eskandar, Emad
    Williams, Ziv
    Cosgrove, G. Rees
    Weisholtz, Daniel S.
    Dougherty, Darin D.
    Truccolo, Wilson
    Widge, Alik S.
    Cash, Sydney S.
    NEUROIMAGE, 2020, 223
  • [9] A Miniature Closed-loop Deep Brain Stimulation Device
    Parastarfeizabadi, Mahboubeh
    Kouzani, Abbas Z.
    Gibson, Ian
    Tye, Susannah J.
    2016 38TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2016, : 1786 - 1789
  • [10] A new closed-loop cryosurgical device for endometrial ablation
    Dobak, JD
    Ryba, E
    Kovalcheck, S
    JOURNAL OF THE AMERICAN ASSOCIATION OF GYNECOLOGIC LAPAROSCOPISTS, 2000, 7 (02): : 245 - 249