Development of a circulatory mock loop for biventricular device testing with various heart conditions

被引:11
作者
Kado, Yuichiro [1 ]
Miyamoto, Takuma [1 ]
Horvath, David J. [1 ,2 ]
Gao, Shengqiang [1 ,3 ]
Fukamachi, Kiyotaka [1 ]
Karimov, Jamshid H. [1 ]
机构
[1] Cleveland Clin, Dept Biomed Engn, Lerner Res Inst, ND20,9500 Euclid Ave, Cleveland, OH 44195 USA
[2] R1 Engn LLC, Euclid, OH USA
[3] Cleveland Clin, Med Device Solut, Lerner Res Inst, Cleveland, OH 44106 USA
关键词
Assisted circulation; heart failure; modeling; mechanical circulatory support; heart-assist devices; in vitro; ventricular assist device; VENTRICULAR ASSIST DEVICE; FAILURE; BLOOD;
D O I
10.1177/0391398820903316
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study aimed to evaluate a newly designed circulatory mock loop intended to model cardiac and circulatory hemodynamics for mechanical circulatory support device testing. The mock loop was built with dedicated ports suitable for attaching assist devices in various configurations. This biventricular mock loop uses two pneumatic pumps (Abiomed AB5000((TM)), Danvers, MA, USA) driven by a dual-output driver (Thoratec Model 2600, Pleasanton, CA, USA). The drive pressures can be individually modified to simulate a healthy heart and left and/or right heart failure conditions, and variable compliance and fluid volume allow for additional customization. The loop output for a healthy heart was tested at 4.2 L/min with left and right atrial pressures of 1 and 5 mm Hg, respectively; a mean aortic pressure of 93 mm Hg; and pulmonary artery pressure of 17 mm Hg. Under conditions of left heart failure, these values were reduced to 2.1 L/min output, left atrial pressure = 28 mm Hg, right atrial pressure = 3 mm Hg, aortic pressure = 58 mm Hg, and pulmonary artery pressure = 35 mm Hg. Right heart failure resulted in the reverse balance: left atrial pressure = 0 mm Hg, right atrial pressure = 30 mm Hg, aortic pressure = 100 mm Hg, and pulmonary artery pressure = 13 mm Hg with a flow of 3.9 L/min. For biventricular heart failure, flow was decreased to 1.6 L/min, left atrial pressure = 13 mm Hg, right atrial pressure = 13 mm Hg, aortic pressure = 52 mm Hg, and pulmonary artery pressure = 18 mm Hg. This mock loop could become a reliable bench tool to simulate a range of heart failure conditions.
引用
收藏
页码:600 / 605
页数:6
相关论文
共 20 条
[1]  
B Libby PO., 2008, BRAUNWALDS HEART DIS
[2]   Pulsatile operation of a centrifugal ventricular assist device with magnetic bearings [J].
Bearnson, GB ;
Olsen, DB ;
Khanwilkar, PS ;
Long, JW ;
Allaire, PE ;
Maslen, EH .
ASAIO JOURNAL, 1996, 42 (05) :M620-M624
[3]   LEFT ATRIAL AND LEFT VENTRICULAR PRESSURES IN SUBJECTS WITHOUT CARDIOVASCULAR DISEASE - OBSERVATIONS IN 18 PATIENTS STUDIED BY TRANSSEPTAL LEFT HEART CATHETERIZATION [J].
BRAUNWALD, E ;
FRAHM, CJ ;
BROCKENBROUGH, EC ;
ROSS, J .
CIRCULATION, 1961, 24 (02) :267-&
[4]   A Physical Heart Failure Simulation System Utilizing the Total Artificial Heart and Modified Donovan Mock Circulation [J].
Crosby, Jessica R. ;
DeCook, Katrina J. ;
Tran, Phat L. ;
Betterton, Edward ;
Smith, Richard G. ;
Larson, Douglas F. ;
Khalpey, Zain I. ;
Burkhoff, Daniel ;
Slepian, Marvin J. .
ARTIFICIAL ORGANS, 2017, 41 (07) :E52-E65
[5]   Initial in vitro testing of a paediatric continuous-flow total artificial heart [J].
Fukamachi, Kiyotaka ;
Karimov, Jamshid H. ;
Horvath, David J. ;
Sunagawa, Gengo ;
Byram, Nicole A. ;
Kuban, Barry D. ;
Moazami, Nader .
INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2018, 26 (06) :897-901
[6]   An innovative, sensorless, pulsatile, continuous-flow total artificial heart: Device design and initial in vitro study [J].
Fukamachi, Kiyotaka ;
Horvath, David J. ;
Massietto, Alex L. ;
Fumoto, Hideyuki ;
Horai, Tetsuya ;
Rao, Santosh ;
Golding, Leonard A. R. .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2010, 29 (01) :13-20
[7]   Development and Validation of a Life-Sized Mock Circulatory Loop of the Human Circulation for Fluid-Mechanical Studies [J].
Gehron, Johannes ;
Zirbes, Julian ;
Bongert, Markus ;
Schaefer, Stefan ;
Fiebich, Martin ;
Krombach, Gabriele ;
Boening, Andreas ;
Grieshaber, Philippe .
ASAIO JOURNAL, 2019, 65 (08) :788-797
[8]  
GUYTON A.C., 1976, TXB MED PHYSL, V5th
[9]   Control of centrifugal blood pump based on the motor current [J].
Iijima, T ;
Inamoto, T ;
Nogawa, M ;
Takatani, S .
ARTIFICIAL ORGANS, 1997, 21 (07) :655-660
[10]   Early in vivo experience with the pediatric continuous-flow total artificial heart [J].
Karimov, Jamshid H. ;
Horvath, David J. ;
Byram, Nicole ;
Sunagawa, Gengo ;
Kuban, Barry D. ;
Gao, Shengqiang ;
Dessoffy, Raymond ;
Fukamachi, Kiyotaka .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2018, 37 (08) :1029-1034