An active approach of pressure waveform matching for stress-based testing of arteries

被引:6
作者
Agrafiotis, Emmanouil [1 ]
Geith, Markus A. [1 ]
Golkani, Mohammad A. [2 ]
Hergesell, Vera [3 ]
Sommer, Gerhard [1 ]
Spiliopoulos, Sotirios [3 ]
Holzapfel, Gerhard A. [1 ,4 ]
机构
[1] Graz Univ Technol, Inst Biomech, Graz, Austria
[2] Graz Univ Technol, Inst Automat & Control, Graz, Austria
[3] Med Univ Graz, Dept Cardiac Surg, Graz, Austria
[4] Norwegian Univ Sci & Technol NTNU, Dept Struct Engn, Trondheim, Norway
关键词
active compliance chamber; arterial physiology; cardiovascular system; feedback control; in vitro testing of arteries; mock circulation loop; variable compliance; MOCK CIRCULATION LOOP; SHEAR-STRESS; DEVICE; DESIGN; MODEL; PERFORMANCE;
D O I
10.1111/aor.14064
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background Arterial compliance assists the cardiovascular system with three key roles: (i) storing up to 50% of the stroke volume; (ii) ensuring blood flow during diastole; (iii) dampening pressure oscillations through arterial distension. In mock circulation loops (MCLs), arterial compliance was simulated either with membrane, spring, or Windkessel chambers. Although they have been shown to be suitable for cardiac device testing, their passive behavior can limit stress-based testing of arteries. Here we present an active compliance chamber with a feedback control of variable compliance as part of an MCL designed for biomechanical evaluation of arteries under physiological waveforms. Materials and Methods The chamber encloses a piston that changes the volume via a cascaded controller when there is a difference between the real-time pressure and the physiological reference pressure with the aim to equilibrate both pressures. Results The experimental results showed repeatable physiological waveforms of aortic pressure in health (80-120 mm Hg), systemic hypertension (90-153 mm Hg), and heart failure reduced ejection fraction (78-108 mm Hg). Statistical validation (n = 20) of the function of the chamber is presented against compared raw data. Conclusion We demonstrate that the active compliance chamber can track the actual pressure of the MCL and balance it in real time (every millisecond) with the reference values in order to shape the given pressure waveform. The active compliance chamber is an advanced tool for MCL applications for biomechanical examination of stented arteries and for preclinical evaluation of vascular implants.
引用
收藏
页码:1562 / 1575
页数:14
相关论文
共 43 条
[1]   Identification of viscoelastic properties of Dacron aortic grafts subjected to physiological pulsatile flow [J].
Amabili, Marco ;
Balasubramanian, Prabakaran ;
Ferrari, Giovanni ;
Franchini, Giulio ;
Giovanniello, Francesco ;
Tubaldi, Eleonora .
JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2020, 110
[2]   Nonlinear Dynamics of Human Aortas for Material Characterization [J].
Amabili, Marco ;
Balasubramanian, Prabakaran ;
Bozzo, Isabella ;
Breslavskyi, Ivan ;
Ferrari, Giovanni ;
Franchini, Giulio ;
Giovanniello, Francesco ;
Pogue, Chloe .
PHYSICAL REVIEW X, 2020, 10 (01)
[3]  
Babbs CF., 2018, WELDON SCH BIOMED EN
[4]  
Chen C.-T., 1995, Analog and Digital Control System Design: Transfer-function, Statespace, and Algebraic Methods
[5]   Energy loss, a novel biomechanical parameter, correlates with aortic aneurysm size and histopathologic findings [J].
Chung, Jennifer ;
Lachapelle, Kevin ;
Wener, Evan ;
Cartier, Raymond ;
De Varennes, Benoit ;
Fraser, Richard ;
Leask, Richard L. .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2014, 148 (03) :1082-1088
[6]   A Modified Elastance Model to Control Mock Ventricles in Real-Time: Numerical and Experimental Validation [J].
Colacino, Francesco Maria ;
Moscato, Francesco ;
Piedimonte, Fabio ;
Danieli, Guido ;
Nicosia, Salvatore ;
Arabia, Maurizio .
ASAIO JOURNAL, 2008, 54 (06) :563-573
[7]   Design and evaluation of a hybrid mock circulatory loop for total artificial heart testing [J].
Cuenca-Navalon, Elena ;
Finocchiaro, Thomas ;
Laumen, Marco ;
Fritschi, Andreas ;
Schmitz-Rode, Thomas ;
Steinseifer, Ulrich .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2014, 37 (01) :71-80
[8]   The role of shear stress in the pathogenesis of atherosclerosis [J].
Cunningham, KS ;
Gotlieb, AI .
LABORATORY INVESTIGATION, 2005, 85 (01) :9-23
[9]   Pulsatile In Vitro Simulation of the Pediatric Univentricular Circulation for Evaluation of Cardiopulmonary Assist Scenarios [J].
Dur, Onur ;
Lara, Mikhail ;
Arnold, Dorian ;
Vandenberghe, Stijn ;
Keller, Bradley B. ;
DeGroff, Curt ;
Pekkan, Kerem .
ARTIFICIAL ORGANS, 2009, 33 (11) :967-976
[10]   Systemic hypertension [J].
Elliott, William J. .