摘要
Objectives. We examined the feasibility of transesophageal echocardiography (TEE) using a 10F monoplane probe developed for intracardiac ultrasound (AcuNav, Acuson/Siemens, Mountain View, Calif). Background: Traditional TEE uses a 10- to 12-mm-diameter probe, and conscious sedation is customary to minimize patient discomfort. Because of its small size (3.2-mm diameter), the 10F monoplane probe can be inserted into the esophagus using only topical anesthesia. This provides the potential for a more easily tolerated examination. Methods. A total of 20 patients underwent a comprehensive TEE using an adult multiplane probe. Immediately afterward, the 10F monoplane probe was inserted into the esophagus and a targeted examination completed. The 1OF monoplane studies were blindly reviewed by 3 observers for the study indication and for 16 diagnostic elements. These were graded against an expert's review of standard TEE. Results. The 10F monoplane probe was well tolerated in all patients. Observers A, B, and C answered the clinical question in 80%, 85%, and 100%, respectively, with the 10F probe. The percentage of clinical elements deemed evaluable was 71%, 78%, and 80%, respectively. Limitations included incomplete visualization of the mitral valve and a systematic underestimation of the severity of valve regurgitation. Conclusions: The 10F monoplane probe is safe, well-tolerated, and capable of evaluating many clinical questions. Because of its small size, conscious sedation may not be necessary. It may be useful for targeted evaluations, for monitoring invasive procedures, or for intermediate or long-term monitoring in an intensive care department. (J Am Soc Echocardiogr 2003;16:682-7.).
引用
收藏
页码:682 / 687
页数:6
相关论文
共 8 条
[1]   Intracardiac Doppler hemodynamics and flow: New vector, phased-array ultrasound-tipped catheter [J].
Bruce, CJ ;
Packer, DL ;
Seward, JB .
AMERICAN JOURNAL OF CARDIOLOGY, 1999, 83 (10) :1509-+
[2]   Feasibility study: Transesophageal echocardiography with a 10F (3.2-mm), multifrequency (5.5-to 10-MHz) ultrasound catheter in a small rabbit model [J].
Bruce, CJ ;
Packer, DL ;
O'Leary, PW ;
Seward, JB .
JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 1999, 12 (07) :596-600
[3]   Miniaturized transesophageal echocardiography in newborn infants [J].
Bruce, CJ ;
O'Leary, P ;
Hagler, DJ ;
Seward, JB ;
Cabalka, AK .
JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2002, 15 (08) :791-797
[4]   INTRACARDIAC ECHOCARDIOGRAPHY DURING RADIOFREQUENCY CATHETER ABLATION OF CARDIAC-ARRHYTHMIAS IN HUMANS [J].
CHU, E ;
KALMAN, JM ;
KWASMAN, MA ;
JUE, JCY ;
FITZGERALD, PJ ;
EPSTEIN, LM ;
SCHILLER, NB ;
YOCK, PG ;
LESH, MD .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 24 (05) :1351-1357
[5]   Catheter-based interventions guided solely by a new phased-array intracardiac imaging catheter: In vivo experimental studies [J].
Dairywala, IT ;
Li, P ;
Liu, Z ;
Bowie, D ;
Stewart, SR ;
Bayoumy, AA ;
Murthy, TH ;
Vannan, MA .
JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2002, 15 (02) :150-158
[6]   Multimodal fetal transesophageal echocardiography for fetal cardiac intervention in sheep [J].
Kohl, T ;
Westphal, M ;
Strümper, D ;
Achenbach, S ;
Halimeh, S ;
Petry, P ;
Aryee, S ;
Buller, T ;
Aleksiene, R ;
Asfour, B ;
Witteler, R ;
Vogt, J ;
Van Aken, H ;
Scheld, HH .
CIRCULATION, 2001, 104 (15) :1757-1760
[7]   ROLE OF RIGHT ATRIAL ENDOCARDIAL STRUCTURES AS BARRIERS TO CONDUCTION DURING HUMAN TYPE-I ATRIAL-FLUTTER - ACTIVATION AND ENTRAINMENT MAPPING GUIDED BY INTRACARDIAC ECHOCARDIOGRAPHY [J].
OLGIN, JE ;
KALMAN, JM ;
FITZPATRICK, AP ;
LESH, MD .
CIRCULATION, 1995, 92 (07) :1839-1848
[8]   Intracardiac phased-array imaging: Methods and initial clinical experience with high resolution, under blood visualization - Initial experience with intracardiac phased-array ultrasound [J].
Packer, DL ;
Stevens, CL ;
Curley, MG ;
Bruce, CJ ;
Miller, FA ;
Khandheria, BK ;
Oh, JK ;
Sinak, LJ ;
Seward, JB .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2002, 39 (03) :509-516