Exposure Creep in Computed Radiography: A Longitudinal Study

被引:39
作者
Gibson, Dale J. [1 ,2 ]
Davidson, Robert A. [1 ]
机构
[1] Charles Sturt Univ, Sch Dent & Hlth Sci, Wagga Wagga, NSW 2678, Australia
[2] Flinders Med Ctr, Bedford Pk, SA, Australia
关键词
Exposure creep; computed radiography; digital radiography; exposure indices; DIGITAL RADIOGRAPHY; INDEX; SYSTEMS; QUALITY;
D O I
10.1016/j.acra.2011.12.003
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Exposure creep is the gradual increase in x-ray exposures over time that results in increased radiation dose to the patient. It has been theorized as being a phenomenon that results from the wide-exposure latitude of computed radiography (CR) and direct/indirect digital radiography (DR). This project evaluates radiographic exposures over 43 months to determine if exposure creep exists and if measures can be applied to halt or reverse exposure creep trends. Methods: Exposure indices were initially recorded over 29 months between August 2007 and December 2009 from the intensive and critical care unit (ICCU) and the emergency department (ED) departments where manual CR exposures were used. The data from this period were then assessed and the exposure indexes (El) values from the radiographic images were compared to the radiology department criteria of El values between 1400 to 1800 as being in the optimal exposure range. El values below this were considered underexposed and over this as overexposed. An intervention was required to be used in ICCU and implemented in January 2010 to halt a noted trend of overexposure. The El value for each chest x-ray (CXR) was recorded in the patients' ICCU records and was to be used by radiologic technologists/radiographers in determine exposure factors in subsequent CXR. After the intervention, El values were recorded and evaluated for an additional 15 months between February 2010 and March 2011. Results: Between August 2007 and December 2009, 17,678 ICCU CXR images and 69,327 ED x-ray examinations were evaluated for over-and underexposure. A trend was noted in ICCU that showed a significant increase (P=.023) in El values from the beginning to the end of the evaluation. No such trend was seen in the ED El values (P=.120). After the intervention in ICCU, the overexposure trend was halted. Conclusions: Exposure creep has been show to exist. It is surmised that this occurs where judgment to determine the correct radiographic exposure factors is needed when taking into account a large range of patient sizes. It has also been shown that providing radiologic technologists/radiographers with previous El values for the same x-ray examination can halt a trend of exposure creep.
引用
收藏
页码:458 / 462
页数:5
相关论文
共 8 条
[1]   Effect of x-ray tube potential on image quality and patient dose for lumbar spine computed radiography examinations [J].
Brindhaban A. ;
Al Khalifah K. ;
Al Wathiqi G. ;
Al Ostath H. .
Australasian Physics & Engineering Sciences in Medicine, 2005, 28 (4) :216-222
[2]   Are exposure index values consistent in clinical practice? A multi-manufacturer investigation [J].
Butler, M. L. ;
Rainford, L. ;
Last, J. ;
Brennan, P. C. .
RADIATION PROTECTION DOSIMETRY, 2010, 139 (1-3) :371-374
[3]   Advances in computed radiography systems and their physical imaging characteristics [J].
Cowen, A. R. ;
Davies, A. G. ;
Kengyelics, S. M. .
CLINICAL RADIOLOGY, 2007, 62 (12) :1132-1141
[4]   Computed Radiography and Dosimetry: Some Practical Tips for Dose Optimization Procedures [J].
Davidson, Rob ;
Sim, Jenny .
JOURNAL OF MEDICAL IMAGING AND RADIATION SCIENCES, 2008, 39 (03) :109-114
[5]   Digital radiography: are the manufacturers' settings too high? Optimisation of the Kodak digital radiography system with aid of the computed radiography dose index [J].
Peters, SE ;
Brennan, PC .
EUROPEAN RADIOLOGY, 2002, 12 (09) :2381-2387
[6]   Correlation of image quality with exposure index and processing protocol in a computed radiography system [J].
Tsalafoutas, I. A. ;
Blastaris, G. A. ;
Moutsatsos, A. S. ;
Chios, P. S. ;
Efstathopoulos, E. P. .
RADIATION PROTECTION DOSIMETRY, 2008, 130 (02) :162-171
[7]   Quality assurance of computed and digital radiography systems [J].
Walsh, C. ;
Gorman, D. ;
Byrne, P. ;
Larkin, A. ;
Dowling, A. ;
Malone, J. F. .
RADIATION PROTECTION DOSIMETRY, 2008, 129 (1-3) :271-275
[8]   An assessment of exposure indices in computed radiography for the posterior-anterior chest and the lateral lumbar spine [J].
Warren-Forward, H. ;
Arthur, L. ;
Hobson, L. ;
Skinner, R. ;
Watts, A. ;
Clapham, K. ;
Lou, D. ;
Cook, A. .
BRITISH JOURNAL OF RADIOLOGY, 2007, 80 (949) :26-31