The renewable bismuth bulk annular band working electrode: Fabrication and application in the adsorptive stripping voltammetric determination of nickel(II) and cobalt(II)

被引:48
作者
Bas, Boguslaw [1 ]
Wegiel, Krystian [1 ]
Jedlinska, Katarzyna [1 ]
机构
[1] AGH Univ Sci & Technol, Fac Mat Sci & Ceram, PL-30059 Krakow, Poland
关键词
Bismuth bulk electrode; Stripping voltammetry; Nickel; Cobalt; Water analysis; Mercury-free electroanalysis; FILM ELECTRODE; WATER SAMPLES; ENVIRONMENTAL WATER; RING ELECTRODE; HEAVY-METALS; TRACE-METALS; SILVER; LEAD; NI; CO;
D O I
10.1016/j.aca.2015.05.005
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The paper presents the first report on fabrication and application of a user friendly and mercury free electrochemical sensor, with the renewable bismuth bulk annular band working electrode (RBiABE), in stripping voltammetry (SV). The sensor body is partly filled with the internal electrolyte solution, in which the RBiABE is cleaned and activated before each measurement. Time of the RBiABE contact with the sample solution is precisely controlled. The usefulness of this sensor was tested by Ni(II) and Co(II) traces determination by means of differential pulse adsorptive stripping voltammetry (DP AdSV), after complexation with dimethylglyoxime (DMG) in ammonia buffer (pH 8.2). The experimental variables (composition of the supporting electrolyte, pre-concentration potential and time, potential of the RBiABE activation, and DP parameters), as well as possible interferences, were investigated. The linear calibration graphs for Ni(II) and Co(II), determined individually and together, in the range from 1 x 10(-8) to 70 x 10(-8) mol L-1 and from 1 x 10(-9) to 70 x 10(-9) mol L-1 respectively, were obtained. The calculated limit of detection (LOD), for 30 s of the accumulation time, was 3 x 10(-9) mol L-1 for Ni(II) in case of a single element's analysis, whereas the LOD was 5 x 10(-9) mol L-1 for Ni(II) and 3 x 10(-10) mol L-1 for Co(II), when both metal ions were measured together. The repeatability of the Ni(II) and Co(II) adsorptive stripping voltammetric signals obtained at the RBiABE were equal to 5.4% and 2.5%, respectively (n = 5). Finally, the proposed method was validated by determining Ni(II) and Co(II) in the certified reference waters (SPS-SW1 and SPS-SW2) with satisfactory results. (C) 2015 Elsevier B.V. All rights reserved.
引用
收藏
页码:44 / 53
页数:10
相关论文
共 56 条
[1]   ASSESSMENT OF DIFFERENTIAL-PULSE ADSORPTION VOLTAMMETRY FOR THE SIMULTANEOUS DETERMINATION OF NICKEL AND COBALT IN BIOLOGICAL-MATERIALS [J].
ADELOJU, SB ;
BOND, AM ;
BRIGGS, MH .
ANALYTICA CHIMICA ACTA, 1984, 164 (OCT) :181-194
[2]   Potential of porphyrins as chromogenic reagents for determining metals in capillary electrophoresis [J].
Almeda, S. ;
Gandolfi, H. E. ;
Arce, L. ;
Valcarcel, M. .
JOURNAL OF CHROMATOGRAPHY A, 2009, 1216 (34) :6256-6258
[3]   Simultaneous Determination of Nickel and Cobalt Using a Solid Bismuth Vibrating Electrode by Adsorptive Cathodic Stripping Voltammetry [J].
Alves, Georgina M. S. ;
Magalhaes, Julia M. C. S. ;
Soares, Helena M. V. M. .
ELECTROANALYSIS, 2013, 25 (05) :1247-1255
[4]  
[Anonymous], 2001, ELECTROCHEMICAL METH
[5]   Individual and simultaneous determination of lead, cadmium, and zinc by anodic stripping voltammetry at a bismuth bulk electrode [J].
Armstrong, Kristie C. ;
Tatum, Clarissa E. ;
Dansby-Sparks, Royce N. ;
Chambers, James Q. ;
Xue, Zi-Ling .
TALANTA, 2010, 82 (02) :675-680
[6]   TIC working electrode. Voltammetric characteristics and application for determination of lead traces by stripping voltammetry [J].
Bas, B. ;
Piech, R. ;
Niewiara, E. ;
Ziemnicka, A. ;
Stobierski, L. ;
Kubiak, W. W. .
ELECTROANALYSIS, 2008, 20 (15) :1655-1664
[7]   Renewable Ceramic (TiN) Ring Electrode in Stripping Voltammetry. Determination of Pb(II) Without Removal of Oxygen [J].
Bas, B. ;
Piech, R. ;
Ziemnicka, M. ;
Reczynski, W. ;
Robotka, M. .
ELECTROANALYSIS, 2009, 21 (16) :1773-1780
[8]   The renovated silver ring electrode [J].
Bas, Boguslaw .
ELECTROCHEMISTRY COMMUNICATIONS, 2008, 10 (01) :156-160
[9]   Rapid pretreatment of a solid silver electrode for routine analytical practice [J].
Bas, Boguslaw ;
Jakubowska, Malgorzata ;
Kowalski, Zygmunt .
ELECTROANALYSIS, 2006, 18 (17) :1710-1717
[10]   Rapidly renewable silver and gold annular band electrodes [J].
Bas, Boguslaw ;
Jakubowska, Malgorzata ;
Reczynski, Witold ;
Ciepiela, Filip ;
Kubiak, Wladyslaw W. .
ELECTROCHIMICA ACTA, 2012, 73 :98-104