SETUP REPRODUCIBILITY FOR THORACIC AND UPPER GASTROINTESTINAL RADIATION THERAPY: INFLUENCE OF IMMOBILIZATION METHOD AND ON-LINE CONE-BEAM CT GUIDANCE

被引:14
|
作者
Li, Winnie [1 ]
Moseley, Douglas J.
Bissonnette, Jean-Pierre
Purdie, Thomas G.
Bezjak, Andrea
Jaffray, David A.
机构
[1] Princess Margaret Hosp, Radiat Med Program, Toronto, ON M5G 2M9, Canada
关键词
Thoracic and upper gastrointestinal radiotherapy; Cone-beam CT; Immobilization; Margins; IMAGE-GUIDED RADIOTHERAPY; FLAT-PANEL IMAGER; CELL LUNG-CANCER; COMPUTED-TOMOGRAPHY; CONFORMAL RADIOTHERAPY; PROSTATE-CANCER; ERRORS; LOCALIZATION; VERIFICATION; MARGINS;
D O I
10.1016/j.meddos.2009.09.003
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
We report the setup reproducibility of thoracic and upper gastrointestinal (UGI) radiotherapy (RT) patients for 2 immobilization methods evaluated through cone-beam computed tomography (CBCT) image guidance, and present planning target volume (PTV) margin calculations made on the basis of these observations. Daily CBCT images from 65 patients immobilized in a chestboard (CB) or evacuated cushion (EC) were registered to the planning CT using automatic bony anatomy registration. The standardized region-of-interest for matching was focused around vertebral bodies adjacent to tumor location. Discrepancies >3 mm between the CBCT and CT datasets were corrected before initiation of RT and verified with a second CBCT to assess residual error (usually taken after 90 s of the initial CBCT). Positional data were analyzed to evaluate the magnitude and frequencies of setup errors before and after correction. The setup distributions were slightly different for the CB (797 scans) and EC (757 scans) methods, and the probability of adjustment at a 3-mm action threshold was not significantly different (p = 0.47). Setup displacements >10 mm in any direction were observed in 10% of CB fractions and 16% of EC fractions (p = 0.0008). Residual error distributions after CBCT guidance were equivalent regardless of immobilization method. Using a published formula, the PTV margins for the CB were L/R, 3.3 mm; S/I, 3.5 mm; and A/P, 4.6 mm), and for EC they were L/R, 3.7 mm; S/I, 3.3 mm; and A/P, 4.6 mm. In the absence of image guidance, the CB slightly outperformed the EC in precision. CBCT allows reduction to a single immobilization system that can be chosen for efficiency, logistics, and cost. Image guidance allows for increased geometric precision and accuracy and supports a corresponding reduction in PTV margin. 0 2010 American Association of Medical Dosimetrists.
引用
收藏
页码:287 / 296
页数:10
相关论文
共 50 条
  • [21] Impact of belly board immobilization devices and body mass factor on setup displacement using daily cone-beam CT in rectal cancer radiotherapy
    Ruan, Junjie
    Huang, Xiaotong
    Wang, Tong
    Mai, Xiuying
    Lin, Chuyan
    Li, Fanghua
    Li, Yunfeng
    Chi, Feng
    Li, Bin
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2025, 26 (03):
  • [22] Algorithm-enabled exploration of image-quality potential of cone-beam CT in image-guided radiation therapy
    Han, Xiao
    Pearson, Erik
    Pelizzari, Charles
    Al-Hallaq, Hania
    Sidky, Emil Y.
    Bian, Junguo
    Pan, Xiaochuan
    PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (12) : 4601 - 4633
  • [23] Investigation of cone-beam CT image quality trade-off for image-guided radiation therapy
    Bian, Junguo
    Sharp, Gregory C.
    Park, Yang-Kyun
    Ouyang, Jinsong
    Bortfeld, Thomas
    El Fakhri, Georges
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (09) : 3317 - 3346
  • [24] IMAGING DOSES AND SECONDARY CANCER RISK FROM KILOVOLTAGE CONE-BEAM CT IN RADIATION THERAPY
    Kim, Dong Wook
    Chung, Weon Kuu
    Yoon, Myonggeun
    HEALTH PHYSICS, 2013, 104 (05): : 499 - 503
  • [25] Initial results for patient setup verification using transperineal ultrasound and cone beam CT in external beam radiation therapy of prostate cancer
    Anne Richter
    Bülent Polat
    Ingulf Lawrenz
    Stefan Weick
    Otto Sauer
    Michael Flentje
    Frederick Mantel
    Radiation Oncology, 11
  • [26] Radiation exposure during TACE procedures using additional cone-beam CT (CBCT) for guidance: safety and precautions
    Jonczyk, M.
    Collettini, F.
    Geisel, D.
    Schnapauff, D.
    Boening, G.
    Wieners, G.
    Gebauer, G.
    ACTA RADIOLOGICA, 2018, 59 (11) : 1277 - 1284
  • [27] Cone-beam computed tomography for on-line image guidance of lung stereotactic radiotherapy: Localization, verification, and intrafraction tumor position
    Purdie, Thomas G.
    Bissonnette, Jean-Pierre
    Franks, Kevin
    Bezjak, Andrea
    Payne, David
    Sie, Fanny
    Sharpe, Michael B.
    Jaffray, David A.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 68 (01): : 243 - 252
  • [28] The influence of target and patient characteristics on the volume obtained from cone beam CT in lung stereotactic body radiation therapy
    Liu, Hong-Wei
    Khan, Rao
    D'Ambrosi, Rafael
    Krobutschek, Krista
    Nugent, Zoann
    Lau, Harold
    RADIOTHERAPY AND ONCOLOGY, 2013, 106 (03) : 312 - 316
  • [29] Post-prostatectomy image-guided radiation therapy: Evaluation of toxicity and inter-fraction variation using online cone-beam CT
    Eldredge, Harriet B.
    Studenski, Matthew
    Keith, Scott W.
    Trabulsi, Edouard
    Lallas, Costas D.
    Gomella, Leonard G.
    Dicker, Adam P.
    Showalter, Timothy N.
    JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY, 2011, 55 (05) : 507 - 515
  • [30] Cone-beam computed tomography for lung cancer - validation with CT and monitoring tumour response during chemo-radiation therapy
    Michienzi, Alissa
    Kron, Tomas
    Callahan, Jason
    Plumridge, Nikki
    Ball, David
    Everitt, Sarah