High-resolution wide-field synchrotron radiation micro-CT for large human lung specimen imaging

被引:5
作者
Umetani, Keiji [1 ]
Itoh, Harumi [2 ]
Kawata, Yoshiki [3 ]
Niki, Noboru [3 ]
机构
[1] Japan Synchrotron Radiat Res Inst, Sayo Cho, Hyogo, Japan
[2] Univ Fukui, Fac Med Sci, Fukui, Japan
[3] Univ Tokushima, Dept Opt Sci & Technol, Tokushima, Tokushima, Japan
来源
ADVANCED OPTICAL IMAGING TECHNOLOGIES | 2018年 / 10816卷
关键词
biomedical imaging; digital single-lens reflex camera; lung specimen; micro-CT; secondary pulmonary lobule; synchrotron radiation; x-ray detector;
D O I
10.1117/12.2501282
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We have been developing a large-field and high-resolution micro-CT system that is highly specialized for visualization of the entire secondary pulmonary lobule in a large human lung specimen using a consumer-grade 36M-pixel single lens reflex camera. The secondary pulmonary lobule, a fundamental unit of lung structure, reproduces the lung in miniature. The lung specimen was set in an acrylic cylindrical case with 36 mm diameter and 40 mm height. A field of view of the micro-CT is 40.6 mm wide x 15.1 mm high with pixel size of 3.07 mu m using offset CT scanning for enlargement of the field of view. In all, 7200 projections were acquired over an angular range of 360 deg with an angular step of 0.05 deg. Then, we constructed a 13,220 x 13,220 x 4912 voxel image with an isotropic voxel size of 3.07 mu m for three-dimensional visualization of the entire secondary lobule. The whole specimen is depicted in the image using large field of view imaging and micro-architectures of alveolar shapes visualized in the identical image.
引用
收藏
页数:16
相关论文
共 23 条
  • [1] Reduction of metal streak artifacts in x-ray computed tomography using a transmission maximum a posteriori algorithm
    De Man, B
    Nuyts, J
    Dupont, P
    Marchal, G
    Suetens, P
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2000, 47 (03) : 977 - 981
  • [2] DR with a DSLR: Digital Radiography with a Digital Single-Lens Reflex camera
    Fan, Helen
    Durko, Heather L.
    Moore, Stephen K.
    Moore, Jared
    Miller, Brian W.
    Furenlid, Lars R.
    Pradhan, Sunil
    Barrett, Harrison H.
    [J]. MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING, 2010, 7622
  • [3] Fu J, 2007, NUCL INSTRUM METH A, V575, P519, DOI [10.1016/.j.nima.2007.03.008, 10.1016/j.nima.2007.03.008]
  • [4] Visualization and stereological characterization of individual rat lung acini by high-resolution X-ray tomographic microscopy
    Haberthuer, David
    Barre, Sebastien F.
    Tschanz, Stefan A.
    Yao, Eveline
    Stampanoni, Marco
    Schittny, Johannes C.
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 2013, 115 (09) : 1379 - 1387
  • [5] Radiation dose optimized lateral expansion of the field of view in synchrotron radiation X-ray tomographic microscopy
    Haberthuer, David
    Hintermueller, Christoph
    Marone, Federica
    Schittny, Johannes C.
    Stampanoni, Marco
    [J]. JOURNAL OF SYNCHROTRON RADIATION, 2010, 17 : 590 - 599
  • [6] Heitzman ER, 1984, LUNG RADIOLOGICPATHO
  • [7] In vitro evaluation of normal and abnormal lungs with ultra-high-resolution CT
    Ikura, H
    Shimizu, K
    Ikezoe, J
    Nagareda, T
    Yagi, N
    [J]. JOURNAL OF THORACIC IMAGING, 2004, 19 (01) : 8 - 15
  • [8] Architecture of the lung - Morphology and function
    Itoh, H
    Nishino, M
    Hatabu, H
    [J]. JOURNAL OF THORACIC IMAGING, 2004, 19 (04) : 221 - 227
  • [9] Itoh H, 1993, J Thorac Imaging, V8, P176, DOI 10.1097/00005382-199322000-00003
  • [10] Kak A.C. Slaney M., 1999, PRINCIPLES COMPUTERI