Assessment of coronary arterial stents by multislice-CT angiography

被引:72
作者
Maintz, D
Grude, M
Fallenberg, EM
Heindel, W
Fischbach, R
机构
[1] Univ Munster, Dept Clin Radiol, DE-48129 Munster, Germany
[2] Univ Munster, Dept Cardiol & Angiol, D-4400 Munster, Germany
关键词
stents and protheses; artifacts; computed tomography; coronary angiography; coronary vessels; SPIRAL COMPUTED-TOMOGRAPHY; MAGNETIC-RESONANCE ANGIOGRAPHY; IN-VITRO EVALUATION; IMPLANTATION; STENOSES; DISEASE; LUMEN;
D O I
10.1080/02841850312331287719
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To assess patency and lumen visibility of coronary artery stents by multislice-CT angiography (MSCTA) in comparison with conventional coronary angiography as the standard of reference. Material and Methods: 47 stents of 13 different types were evaluated in 29 patients. MSCTA was performed on a 4-slice scanner with a standard coronary protocol (detector collimation 4 x 1 mm; table feed 1.5 mm/rotation, 400 mAs, 120 kV). Image evaluation was performed by two readers who were blinded to the reports from the catheter angiography. MIP reconstructions were evaluated for image quality on a 4-point scale (1 = poor, 4 = excellent) and stent patency (contrast distal to the stent as an indirect patency sign). Axial images and multiplanar reformations through the stents were used for assessment of stent lumen visibility (measurement of the visible stent lumen diameter) and detection of relevant in-stent stenosis (greater than or equal to50%). Results: Image quality was fair to good on average (score 2.64 +/- 1.0) and depended on the heart rate (heart rate 45-60: average score 3.2, heart rate 61-70: average score 2.8, heart rate >71: average score 1.4). Thirty-seven stents were correctly classified as patent, 1 was correctly classified as occluded and 9 stents were not assessible due to insufficient image quality because of triggering artifacts. Parts of the stent lumen could be visualized in 30 cases. On average, 20-40% of the stent lumen diameter was visible. Twenty-five stents were correctly classified as having no stenosis, 1 was falsely classified as stenosed, 1 was correctly classified as occluded. In 20 stents lumen visibility was not sufficient for stenosis evaluation. Conclusion: Although the stent lumen may be partly visualized in most stents, a reliable evaluation of in-stent stenoses does not seem practical by 4-slice MSCT. Nevertheless, for stent patency evaluation, MS-CTA might provide valuable clinical information. With submillimeter MSCT (e.g., 16-slice scanners) and more sophisticated reconstruction algorithms, further improvements may be expected.
引用
收藏
页码:597 / 603
页数:7
相关论文
共 17 条
[1]  
Achenbach S, 2001, CIRCULATION, V103, P2535
[2]   ECG-gated reconstructed multi-detector row CT coronary angiography:: Effect of varying trigger delay on image quality [J].
Hong, C ;
Becker, CR ;
Huber, A ;
Schoepf, UJ ;
Ohnesorge, B ;
Knez, A ;
Brüning, R ;
Reiser, MF .
RADIOLOGY, 2001, 220 (03) :712-717
[3]   Coronary magnetic resonance angiography for the detection of coronary stenoses [J].
Kim, WY ;
Danias, PG ;
Stuber, M ;
Flamm, SD ;
Plein, S ;
Nagel, E ;
Langerak, SE ;
Weber, OM ;
Pedersen, EM ;
Schmidt, M ;
Botnar, RM ;
Manning, WJ .
NEW ENGLAND JOURNAL OF MEDICINE, 2001, 345 (26) :1863-1869
[4]   Three-year follow-up after implantation of metallic coronary-artery stents [J].
Kimura, T ;
Yokoi, H ;
Nakagawa, Y ;
Tamura, T ;
Kaburagi, S ;
Sawada, Y ;
Sato, Y ;
Yokoi, H ;
Hamasaki, N ;
Nosaka, H ;
Nobuyoshi, M .
NEW ENGLAND JOURNAL OF MEDICINE, 1996, 334 (09) :561-566
[5]   Imaging of coronary artery stents using multislice computed tomography: in vitro evaluation [J].
Maintz, D ;
Juergens, KU ;
Wichter, T ;
Grude, M ;
Heindel, W ;
Fischbach, R .
EUROPEAN RADIOLOGY, 2003, 13 (04) :830-835
[6]   Multislice CT angiography of the iliac arteries in the presence of various stents - In vitro evaluation of artifacts and lumen visibility [J].
Maintz, D ;
Fischbach, R ;
Juergens, KU ;
Allkemper, T ;
Wessling, J ;
Heindel, W .
INVESTIGATIVE RADIOLOGY, 2001, 36 (12) :699-704
[7]   Coronary magnetic resonance angiography for assessment of the stent lumen: A phantom study [J].
Maintz, D ;
Botnar, RM ;
Fischbach, R ;
Heindel, W ;
Manning, WJ ;
Stuber, M .
JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2002, 4 (03) :359-367
[8]  
Möhlenkamp S, 1999, CATHETER CARDIO INTE, V48, P39
[9]   Reliable noninvasive coronary angiography with fast submillimeter multislice spiral computed tomography [J].
Nieman, K ;
Cademartiri, F ;
Lemos, PA ;
Raaijmakers, R ;
Pattynama, PMT ;
de Feyter, PJ .
CIRCULATION, 2002, 106 (16) :2051-2054
[10]   Non-invasive coronary angiography with multislice spiral computed tomography: impact of heart rate [J].
Nieman, K ;
Rensing, BJ ;
van Geuns, RJM ;
Vos, J ;
Pattynama, PMT ;
Krestin, GP ;
Serruys, PW ;
de Feyter, PJ .
HEART, 2002, 88 (05) :470-474