The Dosimetric Impact of Shifts in Patient Positioning during Boron Neutron Capture Therapy for Brain Tumors

被引:7
作者
Lee, Jia-Cheng [1 ,2 ]
Chen, Yi-Wei [1 ,3 ]
Chuang, Keh-Shih [2 ]
Hsu, Fang-Yuh [4 ]
Chou, Fong-In [4 ]
Hsu, Shih-Ming [5 ]
Yen, Sang-Hue [1 ]
Wu, Yuan-Hung [1 ,3 ,5 ,6 ]
机构
[1] Taipei Vet Gen Hosp, Dept Oncol, Taipei, Taiwan
[2] Natl Tsing Hua Univ, Dept Biomed Engn & Environm Sci, Hsinchu, Taiwan
[3] Natl Yang Ming Univ, Sch Med, Taipei, Taiwan
[4] Natl Tsing Hua Univ, Nucl Sci & Technol Dev Ctr, Hsinchu, Taiwan
[5] Natl Yang Ming Univ, Dept Biomed Imaging & Radiol Sci, Taipei, Taiwan
[6] Natl Yang Ming Univ, Inst Publ Hlth, Taipei, Taiwan
关键词
BEAM; THOR; BNCT; CODE;
D O I
10.1155/2018/5826174
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
Unlike conventional photon radiotherapy, sophisticated patient positioning tools are not available for boron neutron capture therapy (BNCT). Thus, BNCT remains vulnerable to setup errors and intra-fractional patient motion. The aim of this study was to estimate the impact of deviations in positioning on the dose administered by BNCT for brain tumors at the Tsing Hua open-pool reactor (THOR). For these studies, a simulated head model was generated based on computed tomography (CT) images of a patient with a brain tumor. A cylindrical brain tumor 3 cm in diameter and 5 cm in length was modeled at distances of 6.5 cm and 2.5 cm from the posterior scalp of this head model (T-6.5cm and T-2.5cm, respectively). Radiation doses associated with positioning errors were evaluated for each distance, including left and right shifts, superior and inferior shifts, shifts from the central axis of the beam aperture, and outward shifts from the surface of the beam aperture. Rotational and tilting effects were also evaluated. The dose prescription was 20 Gray-equivalent (Gy-Eq) to 80 % of the tumor. The treatment planning system, NCTPlan, was used to perform dose calculations. The average decreases in mean tumor dose for T6.5cm for the 1 cm, 2 cm, and 3 cm lateral shifts composed by left, right, superior, and inferior sides, were approximately 1 %, 6 %, and 11 %, respectively, compared to the dose administered to the initial tumor position. The decreases in mean tumor dose for T-6.5cm were approximately 5 %, 11 %, and 15 % for the 1 cm, 2 cm, and 3 cm outward shifts, respectively. For a superficial tumor at T-2.5cm, no significant decrease in averagemean tumor dose was observed following lateral shifts of 1 cm. Rotational and tilting up to 15 degrees did not result in significant difference to the tumor dose. Dose differences to the normal tissues as a result of the shifts in positioning were also minimal. Taken together, these data demonstrate that the mean dose administered to tumors at greater depths is potentially more vulnerable to deviations in positioning, and greater shift distances resulted in reduced mean tumor doses at the THOR. Moreover, these data provide an estimation of dose differences that are caused by setup error or intra-fractionalmotion during BNCT, and these may facilitate more accurate predictions of actual patient dose in future treatments.
引用
收藏
页数:11
相关论文
共 21 条
  • [1] Abramoff M.D., 2004, Biophotonics Int., V11, P36
  • [2] [Anonymous], 2009, THESIS
  • [3] FORSTER RA, 1985, LECT NOTES PHYS, V240, P33
  • [4] Gonzalez S., NCT PLAN NEW PC VERS
  • [5] Hatanaka H., 1987, BORON NEUTRON CAPTUR
  • [6] Assessment of dose rate scaling factors used in NCTPlan treatment planning code for the BNCT beam of THOR
    Hsu, F. Y.
    Liu, M. T.
    Tung, C. J.
    Liu, Y. W. Hsueh
    Chang, C. C.
    Liu, H. M.
    Chou, F. I.
    [J]. APPLIED RADIATION AND ISOTOPES, 2009, 67 (7-8) : S130 - S133
  • [7] Optimization of the epithermal neutron beam for Boron Neutron Capture Therapy at the Brookhaven Medical Research Reactor
    Hu, JP
    Reciniello, RN
    Holden, NE
    [J]. HEALTH PHYSICS, 2004, 86 (05): : S103 - S109
  • [8] IAEA, 2001, Iaea, V2001, P75
  • [9] Kortesniemi M., 2002, SOLUTIONS CLIN IMPLE
  • [10] Preliminary dosimetric study on feasibility of multi-beam boron neutron capture therapy in patients with diffuse intrinsic pontine glioma without craniotomy
    Lee, Jia-Cheng
    Chuang, Keh-Shih
    Chen, Yi-Wei
    Hsu, Fang-Yuh
    Chou, Fong-In
    Yen, Sang-Hue
    Wu, Yuan-Hung
    [J]. PLOS ONE, 2017, 12 (06):