ADAPT system is a dramatic advance in computer-assisted surgery for femoral trochanteric fractures

被引:7
作者
Murakami, Tomotoshi [1 ]
Yamazaki, Kazuhiro [1 ]
Ogasa, Hiroyoshi [1 ]
机构
[1] Hagi Civil Hosp, Dept Orthoped Surg, 3460-3 Tubaki, Hagi City, Yamaguchi 7580061, Japan
关键词
Computer-assisted surgery; Femoral trochanteric fracture; Fluoroscopy based navigation; Operative time; Improving surgical efficiency; TIP-APEX DISTANCE; FIXATION; NAVIGATION; FAILURE;
D O I
10.1051/sicotj/2021056
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction: In recent years, computer-assisted surgery has made it possible to undergo surgery with a high degree of precision. This study aimed to investigate the usefulness of computer-assisted surgery for femoral trochanteric fractures using the ADAPT (ADAptive Positioning Technology) system. Methods: A total of forty patients with femoral trochanteric fracture underwent intramedullary nailing for fracture fixation: in twenty patients, the ADAPT system (ADAPT group), and in the other twenty, it was not used (control group). The operative time, intraoperative fluoroscopy time, tip apex distance (TAD), and tip to head surface distance (TSD) were measured and compared between the two groups to assess the efficiency and accuracy of the surgery. Results: The operative time was significantly shorter (P < 0.05), intraoperative fluoroscopy time was significantly reduced (P < 0.01), and implant placement was significantly better in the ADAPT group (P < 0.01). Conclusion: Navigation systems have been developed to improve the efficiency of surgery. The ADAPT system was considered a very useful device for intramedullary nailing of femoral trochanteric fractures, as it reduced the intraoperative fluoroscopy time and improved the accuracy of implant placement, also reducing the operative time.
引用
收藏
页数:6
相关论文
共 20 条
[1]   Computer assisted orthopaedic surgery - current state and impact [J].
Adhikari, Ajeya .
SICOT-J, 2018, 4
[2]   Tip apex distance, hip screw placement, and neck shaft angle as potential risk factors for cut-out failure of hip screws after surgical treatment of intertrochanteric fractures [J].
Andruszkow, Hagen ;
Frink, Michael ;
Froemke, Cornelia ;
Matityahu, Amir ;
Zeckey, Christian ;
Mommsen, Philipp ;
Suntardjo, Stefanie ;
Krettek, Christian ;
Hildebrand, Frank .
INTERNATIONAL ORTHOPAEDICS, 2012, 36 (11) :2347-2354
[3]   Computer-assisted Trauma Surgery [J].
Atesok, Kivanc ;
Schemitsch, Emil H. .
JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2010, 18 (05) :247-258
[4]   THE VALUE OF THE TIP-APEX DISTANCE IN PREDICTING FAILURE OF FIXATION OF PERITROCHANTERIC FRACTURES OF THE HIP [J].
BAUMGAERTNER, MR ;
CURTIN, SL ;
LINDSKOG, DM ;
KEGGI, JM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1995, 77A (07) :1058-1064
[5]   Critical factors in cut-out complication after gamma nail treatment of proximal femoral fractures [J].
Bojan, Alicja J. ;
Beimel, Claudia ;
Taglang, Gilbert ;
Collin, David ;
Ekholm, Carl ;
Jonsson, Anders .
BMC MUSCULOSKELETAL DISORDERS, 2013, 14
[6]   Gap difference in navigated TKA: a measure of the imbalanced flexion-extension gap [J].
Chia, Zi-Yang ;
Pang, Hee-Nee ;
Tan, Mann-Hong ;
Yeo, Seng-Jin .
SICOT-J, 2018, 4
[7]   The use of computer navigation in performing minimally invasive surgery for intertrochanteric hip fractures - The experience in Singapore [J].
Chong, K. W. ;
Wong, M. K. ;
Rikhraj, I. S. ;
Howe, T. S. .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2006, 37 (08) :755-762
[8]  
Deep K, 2017, SICOT-J, V3, DOI 10.1051/sicotj/2017034
[9]   Enhanced cephalomedullary nail lag screw placement and intraoperative tip-apex distance measurement with a novel computer assisted surgery system [J].
Do, Mitchell Kuhl ;
Beimel, Claudia .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2016, 47 (10) :2155-2160
[10]   Tip-apex distance of intramedullary devices as a predictor of cut-out failure in the treatment of peritrochanteric elderly hip fractures [J].
Geller, Jeffrey A. ;
Saifi, Comron ;
Morrison, Todd A. ;
Macaulay, William .
INTERNATIONAL ORTHOPAEDICS, 2010, 34 (05) :719-722