An advanced physiological controller design for a left ventricular assist device to prevent left ventricular collapse

被引:47
作者
Wu, Y
Allaire, P
Tao, G
Wood, H
Olsen, D
Tribble, C
机构
[1] Univ Virginia, Artificial Heart Ctr, Charlottesville, VA USA
[2] Utah Artificial Heart Inst, Salt Lake City, UT USA
关键词
left ventricular assist device; continuous flow blood pump; controller; performance criteria; congestive heart failure; left ventricle collapse;
D O I
10.1046/j.1525-1594.2003.00032.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A continuous flow left ventricle assist device (LVAD), which is mainly composed of a continuous flow blood pump and a physiological controller, has only one control input, the rotational speed of the pump, but at least three performance criteria to meet. The challenge for the physiological controller of a long-term continuous flow LVAD is the adaptability to different cardiovascular loading situations and the ability to handle systemic and parametric uncertainties with only one control input. The physiological LVAD controller presented in this article exhibits good performance in terms of the three performance criteria in different physiological loading conditions, such as disturbance, resting, and moderate exercise, for a patient with congestive heart failure. The collapse of the left ventricle, which is an inherent problem for a continuous flow LVAD, has been prevented because of the control algorithm design.
引用
收藏
页码:926 / 930
页数:5
相关论文
共 3 条
[1]   Intelligent control design for heart assist devices [J].
Boston, JR ;
Simaan, MA ;
Antaki, JF ;
Yu, YC ;
Choi, SJ .
JOINT CONFERENCE ON THE SCIENCE AND TECHNOLOGY OF INTELLIGENT SYSTEMS, 1998, :497-502
[2]  
WU Y, 2001, ANN BIOMED ENG S1, V29, pS2
[3]  
WU Y, 2002, ASME INT MECH ENG C