Qualification and validation of software and computer systems in laboratories - Part 4. Evaluation and validation of existing systems

被引:2
作者
Huber, L [1 ]
机构
[1] Hewlett Packard Waldbronn, D-76337 Waldbronn, Germany
关键词
validation; qualification; computers; analytical laboratories; existing systems;
D O I
10.1007/s007690050253
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Existing software and computer systems in laboratories require retrospective evaluation and validation if their initial validation was not formally documented. The key steps in this process are similar to those for the validation of new software and systems: user requirements and system specification, formal qualification, and procedures to ensure ongoing performance during routine operation. The main difference is that frequently qualification of an existing system is based primarily on reliable operation and proof of performance in the past rather than on qualification during development and installation.
引用
收藏
页码:317 / 321
页数:5
相关论文
共 8 条
[1]  
AGALLOCO J, 1987, FARM TECH, P38
[2]  
HAMBLOCH H, 1994, COMPUTER VALIDATION, P93
[3]   Qualification and validation of software and computer systems in laboratories - Part 3. Installation and operational qualification [J].
Huber, L .
ACCREDITATION AND QUALITY ASSURANCE, 1998, 3 (04) :140-144
[4]   Qualification and validation of software and computer systems in laboratories - Part 1: Validation during development [J].
Huber, L ;
Wiederoder, H .
ACCREDITATION AND QUALITY ASSURANCE, 1997, 2 (08) :360-366
[5]   Qualification and validation of software and computer systems in laboratories - Part 2: Qualification of vendors [J].
Huber, L .
ACCREDITATION AND QUALITY ASSURANCE, 1998, 3 (01) :2-5
[6]  
HUBER L, 1995, VALIDATION COMPUTERI
[7]  
*OECD, 1995, SER PRINC GOOD LAB P, V10
[8]  
1996, QUALITY CONTROL REPO, V30, P3