The p48 MW flow modulation device for treatment of unruptured, saccular intracranial aneurysms: a single center experience from 77 consecutive aneurysms

被引:17
作者
AlMatter, Muhammad [1 ]
Henkes, Elina [1 ]
Sirakov, Alexander [1 ,2 ]
Perez, Marta Aguilar [1 ]
Hellstern, Victoria [1 ]
Candel, Carmen Serna [1 ]
Ganslandt, Oliver [3 ]
Henkes, Hans [1 ,4 ]
机构
[1] Klinikum Stuttgart, Neuroradiol Klin, Kriegsberg Str 60, D-70174 Stuttgart, Germany
[2] UH St Ivan Rilski, Radiol Dept, Sofia, Bulgaria
[3] Klinikum Stuttgart, Neurochirurg Klin, Stuttgart, Germany
[4] Univ Duisburg Essen, Med Fak, Essen, Germany
关键词
Aneurysm; Flow diversion; New device; p48; MW; Endovascular treatment; BIFURCATION ANEURYSMS; EMBOLIZATION DEVICE; DIVERTING STENTS;
D O I
10.1186/s42155-020-00131-4
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BackgroundThe p48 MW Flow Modulation Device (phenox, Bochum Germany) is a low profile flow diverter stent (FDS), designed for implantation into intracranial arteries with a diameter of less than 3.5mm.ObjectiveTo evaluate the safety and efficacy of the p48 MW FDS in the treatment of unruptured aneurysms located at intracranial arteries with less than 3.5mm diameter based on a retrospective analysis from a single tertiary neurovascular center.MethodsA prospectively maintained database was retrospectively reviewed to identify all cases of intracranial saccular aneurysms treated electively with the p48 MW device. Records were made of basic demographics, aneurysmal characteristics, interventional procedures, adverse events, clinical outcomes and occlusion rates on angiographic follow-ups.ResultsA total of 77 aneurysms and 74 patients were included. The mean size of the treated aneurysms was 3.52.4mm and the mean aspect ratio was 1.3 +/- 0.4. A total of 80 endovascular procedures were performed with a total of 12 (15%) adverse events leading to two (2.5%) permanent morbidities/mortalities. Technical issues were encountered in 3 (3.9%) cases. Adequate occlusion of the treated aneurysm was recorded in 55.6% and 63.9% on the first and latest available DSA follow-ups, respectively. There were no cases of side-branch occlusion.ConclusionsThe p48 MW is an easy-to-use implant with very good safety margins. Side branch occlusion and significant in-stent stenosis are infrequently encountered. The time from implantation to sufficient aneurysm occlusion takes longer than with FDS with lower porosity.
引用
收藏
页数:8
相关论文
共 24 条
[1]   The p48 Flow Modulation Device with Hydrophilic Polymer Coating (HPC) for the Treatment of Acutely Ruptured Aneurysms: Early Clinical Experience Using Single Antiplatelet Therapy [J].
Aguilar-Perez, Marta ;
Hellstern, Victoria ;
AlMatter, Muhammad ;
Wendl, Christina ;
Baezner, Hansjoerg ;
Ganslandt, Oliver ;
Henkes, Hans .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2020, 43 (05) :740-748
[2]   The p48MW Flow Diverter-Initial Human Experience [J].
Bhogal, P. ;
Bleise, C. ;
Chudyk, J. ;
Lylyk, I ;
Viso, R. ;
Perez, N. ;
Henkes, H. ;
Lylyk, P. .
CLINICAL NEURORADIOLOGY, 2021, 31 (01) :135-145
[3]   Treatment of posterior circulation non-saccular aneurysms with flow diverters: a single-center experience and review of 56 patients [J].
Bhogal, P. ;
Perez, M. Aguilar ;
Ganslandt, O. ;
Baezner, H. ;
Henkes, H. ;
Fischer, S. .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2017, 9 (05) :471-+
[4]   The p48_HPC antithrombogenic flow diverter: initial human experience using single antiplatelet therapy [J].
Bhogal, Pervinder ;
Bleise, Carlos ;
Chudyk, Jorge ;
Lylyk, Ivan ;
Perez, Nicolas ;
Henkes, Hans ;
Lylyk, Pedro .
JOURNAL OF INTERNATIONAL MEDICAL RESEARCH, 2020, 48 (01)
[5]   The Use of Flow Diversion in Vessels ≤2.5 mm in Diametere-A Single-Center Experience [J].
Bhogal, Pervinder ;
Chudyk, Jorge ;
Bleise, Carlos ;
Lylyk, Ivan ;
Perez, Nicolas ;
Henkes, Hans ;
Lylyk, Pedro .
WORLD NEUROSURGERY, 2018, 118 :E575-E583
[6]   Endovascular Treatment of Intracranial Aneurysms With Flow Diverters A Meta-Analysis [J].
Brinjikji, Waleed ;
Murad, Mohammad H. ;
Lanzino, Giuseppe ;
Cloft, Harry J. ;
Kallmes, David F. .
STROKE, 2013, 44 (02) :442-447
[7]   Treatment of Unruptured Distal Anterior Circulation Aneurysms with Flow-Diverter Stents: A Meta-Analysis [J].
Cagnazzo, F. ;
Perrini, P. ;
Dargazanli, C. ;
Lefevre, P-H ;
Gascou, G. ;
Morganti, R. ;
di Carlo, D. ;
Derraz, I ;
Riquelme, C. ;
Bonafe, A. ;
Costalat, V .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2019, 40 (04) :687-693
[8]   Treatment of Distal Anterior Cerebral Artery Aneurysms with Flow-Diverter Stents: A Single-Center Experience [J].
Cagnazzo, F. ;
Cappucci, M. ;
Dargazanli, C. ;
Lefevre, P. -H. ;
Gascou, G. ;
Riquelme, C. ;
Bonafe, A. ;
Costalat, V. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2018, 39 (06) :1100-1106
[9]   Flow diversion treatment of complex bifurcation aneurysms beyond the circle of Willis: a single-center series with special emphasis on covered cortical branches and perforating arteries [J].
Gawlitza, Matthias ;
Januel, Anne-Christine ;
Tall, Philippe ;
Bonneville, Fabrice ;
Cognard, Christophe .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2016, 8 (05) :481-487
[10]   Safety and efficacy of the Pipeline embolization device for treatment of intracranial aneurysms: a pooled analysis of 3 large studies [J].
Kallmes, David F. ;
Brinjikji, Waleed ;
Cekirge, Saruhan ;
Fiorella, David ;
Hanel, Ricardo A. ;
Jabbour, Pascal ;
Lopes, Demetrius ;
Lylyk, Pedro ;
McDougall, Cameron G. ;
Siddiqui, Adnan .
JOURNAL OF NEUROSURGERY, 2017, 127 (04) :775-780