Flow-through optosensor combined with photochemically induced fluorescence for simultaneous determination of binary mixtures of sulfonamides in pharmaceuticals, milk and urine

被引:25
作者
Lopez Flores, Javier [1 ]
Fernandez de Cordova, Maria L. [1 ]
Molina Diaz, Antonio [1 ]
机构
[1] Univ Jaen, Fac Expt Sci, Dept Phys & Analyt Chem, E-23071 Jaen, Spain
关键词
multi-optosensor; sulfamethoxazole; sulfathiazole; sulfanilamide; photochemically induced; fluorescence; milk; urine; pharmaceuticals;
D O I
10.1016/j.aca.2007.02.025
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A sensitive and selective flow-through optosensor implemented with photochemically induc ed fluorescence (PIF) is proposed for the simultaneous determination of mixtures sulfamethoxazole/sulfanilamide and sulfathiazole/sulfanilamide. The resolution was accomplished by placing in the flow system a minicolumn filled with an appropriate solid support. Whereas one of the sulfonamides is not retained in the minicolumn and is determined by measuring its native fluorescence on the solid surface of the sensing microbeads in the detection area, the other one is retained and, after its elution, it is photochemically converted into a strongly fluorescent photoproduct which is transitorily retained on the sensing support in the flow cell and monitored. Linear calibration graphs were obtained over a concentration range of 2-3 orders of magnitude. The detection limits for the determination of sulfamethoxazole, sulfanilamide and sulfathiazole are 8.1, 2.9 and 5.7 ng mL(-1), respectively. The method was applied to pharmaceuticals, milk and human urine. The recovery of sulfamethoxazole from pharmaceuticals was 102.5% indicating no interference from trimethoprim which is not photochemically active. The recoveries for urine and milk samples fortified with sulfonamides at levels between 0.1 and 0.7 mu g mL(-1) agreed within 9S.0-107.5% of spiked levels. (C) 2007 Elsevier B.V All rights reserved.
引用
收藏
页码:164 / 171
页数:8
相关论文
共 27 条