Calibration of RGBD camera and cone-beam CT for 3D intra-operative mixed reality visualization

被引:27
作者
Lee, Sing Chun [1 ]
Fuerst, Bernhard [1 ,2 ]
Fotouhi, Javad [1 ]
Fischer, Marius [1 ,3 ]
Osgood, Greg [4 ]
Navab, Nassir [1 ,2 ]
机构
[1] Johns Hopkins Univ, Comp Aided Med Procedures, Baltimore, MD USA
[2] Tech Univ Munich, Comp Aided Med Procedures, D-80290 Munich, Germany
[3] Ludwig Maximilians Univ Munchen, Klinikum Innenstadt, Dept Trauma Surg, Munich, Germany
[4] Johns Hopkins Univ Hosp, Dept Orthopaed Surg, Orthopaed Trauma, Baltimore, MD 21287 USA
关键词
Augmented reality; Cone-beam CT; C-arm; Intra-operative imaging; 3D-3D calibration; NAVIGATION; RADIATION; FLUOROSCOPY; ACCURACY; SYSTEM;
D O I
10.1007/s11548-016-1396-1
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This work proposes a novel algorithm to register cone-beam computed tomography (CBCT) volumes and 3D optical (RGBD) camera views. The co-registered real-time RGBD camera and CBCT imaging enable a novel augmented reality solution for orthopedic surgeries, which allows arbitrary views using digitally reconstructed radiographs overlaid on the reconstructed patient's surface without the need to move the C-arm. An RGBD camera is rigidly mounted on the C-arm near the detector. We introduce a calibration method based on the simultaneous reconstruction of the surface and the CBCT scan of an object. The transformation between the two coordinate spaces is recovered using Fast Point Feature Histogram descriptors and the Iterative Closest Point algorithm. Several experiments are performed to assess the repeatability and the accuracy of this method. Target registration error is measured on multiple visual and radio-opaque landmarks to evaluate the accuracy of the registration. Mixed reality visualizations from arbitrary angles are also presented for simulated orthopedic surgeries. To the best of our knowledge, this is the first calibration method which uses only tomographic and RGBD reconstructions. This means that the method does not impose a particular shape of the phantom. We demonstrate a marker-less calibration of CBCT volumes and 3D depth cameras, achieving reasonable registration accuracy. This design requires a one-time factory calibration, is self-contained, and could be integrated into existing mobile C-arms to provide real-time augmented reality views from arbitrary angles.
引用
收藏
页码:967 / 975
页数:9
相关论文
共 50 条
  • [21] 3D morphometric analysis of the human incudomallear complex using clinical cone-beam CT
    Soons, Joris A. M.
    Danckaers, Femke
    Keustermans, William
    Huysmans, Toon
    Sijbers, Jan
    Casselman, Jan W.
    Dirckx, Joris J. J.
    HEARING RESEARCH, 2016, 340 : 79 - 88
  • [22] Interest of intra-operative 3D imaging in spine surgery: a prospective randomized study
    Ruatti, Sebastien
    Dubois, C.
    Chipon, E.
    Kerschbaumer, G.
    Milaire, M.
    Moreau-Gaudry, A.
    Tonetti, J.
    Merloz, Ph.
    EUROPEAN SPINE JOURNAL, 2016, 25 (06) : 1738 - 1744
  • [23] Intraoperative 3D imaging with cone-beam computed tomography leads to revision of pedicle screws in dorsal instrumentation: a retrospective analysis
    Zimmermann, Felix
    Kohl, Katharina
    Privalov, Maxim
    Franke, Jochen
    Vetter, Sven Y.
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2021, 16 (01)
  • [24] Correction of patient motion in cone-beam CT using 3D-2D registration
    Ouadah, S.
    Jacobson, M.
    Stayman, J. W.
    Ehtiati, T.
    Weiss, C.
    Siewerdsen, J. H.
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (23) : 8813 - 8831
  • [25] Intra-operative visualisation of 3D temperature maps and 3D navigation during tissue cryoablation
    Samset, E
    Mala, T
    Aurdal, L
    Balasingham, I
    COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2005, 29 (06) : 499 - 505
  • [26] Frequency-Combined Extended 3D Reconstruction for Multiple Circular Cone-Beam CT Scans
    Grimmer, Rainer
    Baek, Jongduk
    Pelc, Norbert
    Kachelriess, Marc
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 4089 - 4092
  • [27] Local and global 3D noise power spectrum in cone-beam CT system with FDK reconstruction
    Baek, Jongduk
    Pelc, Norbert J.
    MEDICAL PHYSICS, 2011, 38 (04) : 2122 - 2131
  • [28] Dose comparative study between two different 3D mobile intra-operative systems
    Sans Merce, M.
    Pitteloud, N.
    Savoye, F.
    RADIOPROTECTION, 2020, 55 (03) : 179 - 185
  • [29] Intra-Operative 3D Registration of MIS Reconstructed Surfaces to Pre-Operative Models
    Peleka, Georgia
    Zampokas, Georgios
    Mariolis, Ioannis
    Malasiotis, Sotiris
    Tzovaras, Dimitrios
    2018 IEEE INTERNATIONAL CONFERENCE ON IMAGING SYSTEMS AND TECHNIQUES (IST), 2018, : 402 - 407
  • [30] Repositioning accuracy of two different mask systems - 3D revisited: Comparison using true 3D/3D matching with cone-beam CT
    Boda-Heggemann, Judit
    Walter, Cornelia
    Rahn, Angelika
    Wertz, Hansjoerg
    Loeb, Iris
    Lohr, Frank
    Wenz, Frederik
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 66 (05): : 1568 - 1575