Preliminary Investigation to Improve Point Spread Function Modeling for a Multi-Pinhole SPECT Camera

被引:0
|
作者
Banerjee, Soumyanil [1 ]
Konik, Arda [1 ]
Mukherjee, Joyeeta M. [2 ]
Kalluri, Kesava S. [1 ]
Goding, Justin C. [1 ]
Caucci, Luca [4 ,5 ]
Zubal, George I. [3 ]
Furenlid, Lars R. [4 ,5 ]
King, Michael A. [1 ]
机构
[1] UMass Med Sch, Dept Radiol, Worcester, MA 01655 USA
[2] MathWorks, Natick, MA USA
[3] Z Concepts LLC, East Haven, CT USA
[4] Univ Arizona, Dept Radiol, Tucson, AZ USA
[5] Coll Opt Sci, Tucson, AZ USA
来源
2017 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC) | 2017年
基金
美国国家卫生研究院;
关键词
SPECT; DaTscan; Multi-pinhole (MPH); Point Spread Function (PSF); Gaussian; GaussianC; GaussianCC; Gaussian CC4;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Herein we report on the mathematical modeling of the simulated point spread functions (PSFs) of pinhole apertures for clinical I-123 DaTscan imaging on a dual-head SPECT system consisting of fan and multi-pinhole (MPH) collimators on separate heads. The PSFs can be measured sparsely by translating a point source within the volume of interest (VoI). These PSFs were generated using GATE Monte Carlo simulation software and were then modeled using standard 2D Gaussian having 6 parameters, and three other models using higher order polynomial terms as well as cross terms in the exponential. The goal is to efficiently store the parameters of the modeled PSF, measured across the VOI and then interpolate them on the fly during reconstruction. It has been shown that MPH reconstruction can be improved with accurate modeling of the PSF. However, for our application it has been determined that improved accuracy in PSF modeling (reduced NRMSE) can be obtained by incorporating more polynomial terms in the exponential than employed by the standard 2D Gaussian, especially with increased pinhole angulations. In this paper we introduce higher order polynomial terms (degree 3 and 4) as an extension to the Gaussian model and observe that these added terms could significantly reduce the NRMSE.
引用
收藏
页数:2
相关论文
共 42 条
  • [41] The effect of time-of-flight and point spread function modeling on 82Rb myocardial perfusion imaging of obese patients
    Dasari, Paul K. R.
    Jones, Judson P.
    Casey, Michael E.
    Liang, Yuanyuan
    Dilsizian, Vasken
    Smith, Mark F.
    JOURNAL OF NUCLEAR CARDIOLOGY, 2018, 25 (05) : 1521 - 1545
  • [42] Impact of point spread function modeling and time-of-flight on myocardial blood flow and myocardial flow reserve measurements for rubidium-82 cardiac PET
    Armstrong, Ian S.
    Tonge, Christine M.
    Arumugam, Parthiban
    JOURNAL OF NUCLEAR CARDIOLOGY, 2014, 21 (03) : 467 - 474