Disparities in Hypoxemia Detection by Pulse Oximetry Across Self-Identified Racial Groups and Associations With Clinical Outcomes*

被引:88
作者
Henry, Nicole R. [1 ]
Hanson, Andrew C. [2 ]
Schulte, Phillip J. [2 ]
Warner, Nafisseh S. [3 ]
Manento, Megan N. [4 ]
Weister, Timothy J. [5 ]
Warner, Matthew A. [4 ]
机构
[1] Mayo Clin, Dept Anesthesiol & Perioperat Med, Rochester, MN USA
[2] Mayo Clin, Dept Biomed Stat & Informat, Rochester, MN USA
[3] Mayo Clin, Div Pain Med, Dept Anesthesiol & Perioperat Med, Rochester, MN USA
[4] Mayo Clin, Div Crit Care Med, Dept Anesthesiol & Perioperat Med, Rochester, MN 55902 USA
[5] Mayo Clin, Anesthesia Clin Res Unit, Dept Anesthesiol & Perioperat Med, Rochester, MN USA
基金
美国国家卫生研究院;
关键词
disparities; hypoxemia; oxygen; pulse oximetry; race; MORTALITY; ACCURACY;
D O I
10.1097/CCM.0000000000005394
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
OBJECTIVES: To assess disparities in hypoxemia detection by pulse oximetry across self-identified racial groups and associations with clinical outcomes. DESIGN: Observational cohort study from May 5, 2018, to December 31, 2020. SETTING: Three academic medical centers in the United States. PATIENTS: Adults greater than or equal to 18 years who self-identified as White, Black, Asian, or American Indian admitted to the ICU or undergoing surgery during inpatient hospitalization with simultaneous measurements of pulse oximetry-estimated oxygen saturation and arterial blood gas-derived oxygen saturation. INTERVENTIONS: None. MEASUREMENTS AND MAIN RESULTS: Multivariable models were employed to assess the relationships between race, occult hypoxemia (i.e., arterial blood gas-derived oxygen saturation < 88% despite pulse oximetry-estimated oxygen saturation >= 92%), and clinical outcomes of hospital mortality and hospital-free days. One-hundred twenty-eight-thousand two-hundred eighty-five paired pulse oximetry-estimated oxygen saturation-arterial blood gas-derived oxygen saturation measurements were included from 26,603 patients. Pulse oximetry-estimated oxygen saturation on average overestimated arterial blood gas-derived oxygen saturation by 1.57% (1.54-1.61%). Black, Asian, and American Indian patients were more likely to experience occult hypoxemia during hospitalization (estimated probability 6.2% [5.1-7.6%], 6.6% [4.9-8.8%], and 6.6% [4.4-10.0%], respectively) compared with White patients (3.6% [3.4-3.8%]). Black patients had increased odds of occult hypoxemia compared with White patients after adjustment (odds ratio, 1.65; 1.28-2.14; p < 0.001). Differences in occult hypoxemia between Asian and American Indian patients compared with White patients were not significant after adjustment (odds ratio, 1.53; 0.95-2.47; p = 0.077 and odds ratio, 1.31; 0.80-2.16; p = 0.288, respectively). Occult hypoxemia was associated with increased odds of mortality in surgical (odds ratio, 2.96; 1.20-7.28; p = 0.019) and ICU patients (1.36; 1.03-1.80; p = 0.033). Occult hypoxemia was associated with fewer hospital-free days in surgical (-2.5 d [-3.9 to -1.2 d]; p < 0.001) but not ICU patients (0.4 d [-0.7 to 1.4 d]; p = 0.500). CONCLUSIONS: Occult hypoxemia is more common in Black patients compared with White patients and is associated with increased mortality, suggesting potentially important outcome implications for undetected hypoxemia. It is imperative to validate pulse oximetry with expanded racial inclusion.
引用
收藏
页码:204 / 211
页数:8
相关论文
共 10 条
[1]   Effects of skin pigmentation on pulse oximeter accuracy at low saturation [J].
Bickler, PE ;
Feiner, JR ;
Severinghaus, JW .
ANESTHESIOLOGY, 2005, 102 (04) :715-719
[2]   THE INCIDENCE AND EFFECT ON OUTCOME OF HYPOXEMIA IN HOSPITALIZED MEDICAL PATIENTS [J].
BOWTON, DL ;
SCUDERI, PE ;
HAPONIK, EF .
AMERICAN JOURNAL OF MEDICINE, 1994, 97 (01) :38-46
[3]   Dark Skin Decreases the Accuracy of Pulse Oximeters at Low Oxygen Saturation: The Effects of Oximeter Probe Type and Gender [J].
Feiner, John R. ;
Severinghaus, John W. ;
Bickler, Philip E. .
ANESTHESIA AND ANALGESIA, 2007, 105 :S18-S23
[4]   THE APACHE-III PROGNOSTIC SYSTEM - RISK PREDICTION OF HOSPITAL MORTALITY FOR CRITICALLY ILL HOSPITALIZED ADULTS [J].
KNAUS, WA ;
WAGNER, DP ;
DRAPER, EA ;
ZIMMERMAN, JE ;
BERGNER, M ;
BASTOS, PG ;
SIRIO, CA ;
MURPHY, DJ ;
LOTRING, T ;
DAMIANO, A ;
HARRELL, FE .
CHEST, 1991, 100 (06) :1619-1636
[5]   A review of ASA physical status - historical perspectives and modern developments [J].
Mayhew, D. ;
Mendonca, V. ;
Murthy, B. V. S. .
ANAESTHESIA, 2019, 74 (03) :373-379
[6]   Racial Bias in Pulse Oximetry Measurement [J].
Sjoding, Michael W. ;
Dickson, Robert P. ;
Iwashyna, Theodore J. ;
Gay, Steven E. ;
Valley, Thomas S. .
NEW ENGLAND JOURNAL OF MEDICINE, 2020, 383 (25) :2477-2478
[7]   Racial Bias in Pulse Oximetry Measurement Among Patients About to Undergo Extracorporeal Membrane Oxygenation in 2019-2020 A Retrospective Cohort Study [J].
Valbuena, Valeria S. M. ;
Barbaro, Ryan P. ;
Claar, Dru ;
Valley, Thomas S. ;
Dickson, Robert P. ;
Gay, Steven E. ;
Sjoding, Michael W. ;
Iwashyna, Theodore J. .
CHEST, 2022, 161 (04) :971-978
[8]   The strengthening the reporting of observational studies in epidemiology (STROBE) statement: guidelines for reporting observational studies [J].
von Elm, Erik ;
Altman, Douglas G. ;
Egger, Matthias ;
Pocock, Stuart J. ;
Gotzsche, Peter C. ;
Vandenbroucke, Jan P. .
BULLETIN OF THE WORLD HEALTH ORGANIZATION, 2007, 85 (11) :867-872
[9]   The effect of patient ethnicity on the accuracy of peripheral pulse oximetry in patients with COVID-19 pneumonitis: a single-centre, retrospective analysis [J].
Wiles, M. D. ;
El-Nayal, A. ;
Elton, G. ;
Malaj, M. ;
Winterbottom, J. ;
Gillies, C. ;
Moppett, I. K. ;
Bauchmuller, K. .
ANAESTHESIA, 2022, 77 (02) :143-152
[10]   Association Between Hypoxemia and Mortality in Patients With COVID-19 [J].
Xie, Jiang ;
Covassin, Naima ;
Fan, Zhengyang ;
Singh, Prachi ;
Gao, Wei ;
Li, Guangxi ;
Kara, Tomas ;
Somers, Virend K. .
MAYO CLINIC PROCEEDINGS, 2020, 95 (06) :1138-1147