The Effect of Central Corneal Thickness on Tonometry

被引:1
作者
Boehm, A. G. [1 ]
机构
[1] Elblandklinikum Radebeul, Augenklin, D-01445 Radebeul, Germany
关键词
DYNAMIC CONTOUR TONOMETRY; INTRAOCULAR-PRESSURE MEASUREMENT; BIOMECHANICAL PROPERTIES; APPLANATION TONOMETRY; AGE; CURVATURE; ICARE; XL;
D O I
10.1055/s-0029-1246039
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Central corneal thickness (CCT) affects IOP measurements and is an independent risk factor for the development of glaucoma. IOP measurements of all common tonometers, such as the Goldmann applanation tonometer, non-contact tonometer and rebound tonometer, are affected by CCT. Nomograms to correct IOP measurements according to CCT have been established. These nomograms lead to a reduction of the measurement error caused by CCT in groups of patients. However, one has to be aware of the fact that, in individuals, the correction of IOP measurements can even increase the deviation of the IOP measurement from the actual IOP. The effect of CCT on dynamic contour tonometry and IOP measured by ORA is negligible. CCT is an important parameter in glaucoma management and needs to be considered when interpreting IOP measurements. This can be done by using nomograms or by implementing CCT in the calculation of the individual target pressure.
引用
收藏
页码:114 / 117
页数:4
相关论文
共 26 条
[11]   The relative effects of corneal thickness and age on Goldmann applanation tonometry and dynamic contour tonometry [J].
Kotecha, A ;
White, ET ;
Shewry, JM ;
Garway-Heath, DF .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2005, 89 (12) :1572-1575
[12]   Corneal thickness- and age-related biomechanical properties of the cornea measured with the ocular response analyzer [J].
Kotecha, Aachal ;
Elsheikh, Ahmed ;
Roberts, Cynthia R. ;
Zhu, Haogang ;
Garway-Heath, David F. .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2006, 47 (12) :5337-5347
[13]   Intraocular Pressure Measurement Precision with the Goldmann Applanation, Dynamic Contour, and Ocular Response Analyzer Tonometers [J].
Kotecha, Aachal ;
White, Edward ;
Schlottmann, Patricio G. ;
Garway-Heath, David F. .
OPHTHALMOLOGY, 2010, 117 (04) :730-737
[14]   Influence of corneal biomechanical properties on intraocular pressure measurement - Quantitative analysis [J].
Liu, J ;
Roberts, CJ .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (01) :146-155
[15]   Determining in vivo biomechanical properties of the cornea with an ocular response analyzer [J].
Luce, DA .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (01) :156-162
[16]   Reproducibility and clinical evaluation of rebound tonometry [J].
Martinez-de-la-Casa, JM ;
Garcia-Feijoo, J ;
Castillo, A ;
Garcia-Sanchez, J .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46 (12) :4578-4580
[17]   Effect of corneal thickness on dynamic contour, rebound, and Goldmann tonometry [J].
Martinez-de-la-Casa, Jose M. ;
Garcia-Feijoo, Julian ;
Vico, Eva ;
Fernandez-Vidal, Ana ;
del Castillo, Jose M. Benitez ;
Wasfi, Mohamed ;
Garcia-Sanchez, J. .
OPHTHALMOLOGY, 2006, 113 (12) :2156-2162
[18]   Correlations between corneal biomechanical properties measured with the Ocular Response Analyzer and ICare rebound tonometry [J].
Martins Jorge, Jorge Manuel ;
Gonzalez-Meijome, Jose M. ;
Queiros, Antonio ;
Fernandes, Paulo ;
Parafita, Manuel A. .
JOURNAL OF GLAUCOMA, 2008, 17 (06) :442-448
[19]   Evaluation of the influence of corneal biomechanical properties on Intraocular pressure measurements using the Ocular Response Analyzer [J].
Medeiros, Felipe A. ;
Weinreb, Robert N. .
JOURNAL OF GLAUCOMA, 2006, 15 (05) :364-370
[20]   Determination of the true intraocular pressure and modulus of elasticity of the human cornea in vivo [J].
Orssengo, GJ ;
Pye, DC .
BULLETIN OF MATHEMATICAL BIOLOGY, 1999, 61 (03) :551-572