Simultaneous determination of ascorbic acid, epinephrine, and uric acid by differential pulse voltammetry using poly(3,3′-bis[N,N-bis(carboxymethyl)aminomethyl]-o-cresolsulfonephthalein) modified glassy carbon electrode

被引:50
作者
Ensafi, Ali A. [1 ]
Rezaei, B. [1 ]
Zare, S. Z. Mirahmadi [1 ]
Taei, M. [1 ]
机构
[1] Isfahan Univ Technol, Dept Chem, Esfahan 8415683111, Iran
关键词
Simultaneous determination; Ascorbic acid; Epinephrine; Uric acid; Voltammetry; Poly(3,3'-bis[N,N-bis(carboxymethyl)aminomethyl]-o-cresolsulfonephthalein); ELECTROCHEMICAL-BEHAVIOR; LIQUID-CHROMATOGRAPHY; OXIDATION; FILM; SELECTIVITY; NANOTUBES; DOPAMINE; SENSOR; NAFION;
D O I
10.1016/j.snb.2010.06.066
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A selective and stable poly(3,3'-bis[N,N-bis(carboxymethyl)aminomethyl]-o-cresolsulfonephthalein) modified glassy carbon electrode was prepared using the electrochemical polymerization technique for the simultaneous determination of ascorbic acid (AA), epinephrine (EP), and uric acid (UA). The modified electrode showed an excellent electrocatalytic activity for AA, EP, and UA oxidations and for accelerated electron transfer between the electrode and the substances. The separation potentials of the oxidation peak potentials for EP-AA and EP-UA were about 180 mV and 130 [DV. respectively. Calibration curves in the ranges 1 0-2000 mu mol L-1, 020-175 mu mol L-1, and 0.020-2000 mu mol L-1 were obtained for AA, EP, and UA, respectively. The lowest detection limits (S/N = 3) were 0.4 mu mol L-1, 003 mu mol L-1, and 0.009 mu mol L-1 for AA, EP, and UA, respectively. The proposed procedure was also successfully applied for the simultaneous detection of AA, EP, and UA in injectable medicine, blood plasma, and urine samples (C) 2010 Elsevier B.V. All rights reserved
引用
收藏
页码:321 / 329
页数:9
相关论文
共 34 条
[21]   Determination of catecholamines in human plasma by high-performance liquid chromatography with electrochemical detection [J].
Raggi, MA ;
Sabbioni, C ;
Casamenti, G ;
Gerra, G ;
Calonghi, N ;
Masotti, L .
JOURNAL OF CHROMATOGRAPHY B, 1999, 730 (02) :201-211
[22]   Simultaneous voltammetric measurement of ascorbic acid, epinephrine and uric acid at a glassy carbon electrode modified with caffeic acid [J].
Ren, W ;
Luo, HQ ;
Li, NB .
BIOSENSORS & BIOELECTRONICS, 2006, 21 (07) :1086-1092
[23]   Application of carbon-paste electrode modified with iron phthalocyanine for voltammetric determination of epinephrine in the presence of ascorbic acid and uric acid [J].
Shahrokhian, Saeed ;
Ghalkhani, Masoumeh ;
Amini, Mohammad Kazem .
SENSORS AND ACTUATORS B-CHEMICAL, 2009, 137 (02) :669-675
[24]   Optical fiber biosensor coupled to chromatographic separation for screening of dopamine, norepinephrine and epinephrine in human urine and plasma [J].
Silva, Lurdes I. B. ;
Ferreira, Filipe D. P. ;
Freitas, Ana C. ;
Rocha-Santos, Teresa A. P. ;
Duarte, A. C. .
TALANTA, 2009, 80 (02) :853-857
[25]   Amperometric biosensors based on multiwalled carbon nanotube-Nafion-tyrosinase nanobiocomposites for the determination of phenolic compounds [J].
Tsai, Yu-Chen ;
Chiu, Chian-Cheng .
SENSORS AND ACTUATORS B-CHEMICAL, 2007, 125 (01) :10-16
[26]   Carbon-coated nickel magnetic nanoparticles modified electrodes as a sensor for determination of acetaminophen [J].
Wang, Sheng-Fu ;
Xie, Fen ;
Hu, Ruo-Fei .
SENSORS AND ACTUATORS B-CHEMICAL, 2007, 123 (01) :495-500
[27]   A novel poly(taurine) modified glassy carbon electrode for the simultaneous determination of epinephrine and dopamine [J].
Wang, Yan ;
Chen, Zhen-Zhen .
COLLOIDS AND SURFACES B-BIOINTERFACES, 2009, 74 (01) :322-327
[28]   A simple high-performance liquid chromatography assay for simultaneous determination of plasma norepinephrine, epinephrine, dopamine and 3,4-dihydroxyphenyl acetic acid [J].
Wang, YS ;
Fice, DS ;
Yeung, PKF .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1999, 21 (03) :519-525
[29]   Simultaneous determination of epinephrine and dopamineat with poly(L-arginine) modified electrode [J].
Wei, Ma ;
Sun Deng-Ming .
CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2007, 35 (01) :66-70
[30]   HPLC determination of catecholamines based on precolumn derivatization with 6-aminomethylphthalhydrazide followed by chemiluminescence detection [J].
Yakabe, T ;
Yoshida, H ;
Nohta, H ;
Yamaguchi, M .
ANALYTICAL SCIENCES, 2002, 18 (12) :1375-1378