ID-MAM: A Validated Identity and Multi-Attribute Monitoring Method for Commercial Release and Stability Testing of a Bispecific Antibody

被引:27
作者
Evans, Adam R. [1 ]
Hebert, Alexander S. [1 ]
Mulholland, Joseph [1 ]
Lewis, Michael J. [1 ]
Hu, Ping [1 ]
机构
[1] Janssen Res & Dev LLC, BioTherapeut Dev & Supply Analyt Dev, Malvern, PA 19355 USA
关键词
MASS-SPECTROMETRY; MONOCLONAL-ANTIBODIES; QUALITY-CONTROL; HETEROGENEITY; GLYCOSYLATION; ATTRIBUTES; PROTEIN; DESIGN;
D O I
10.1021/acs.analchem.1c01029
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Post-translational modifications (PTMs) that impact the safety or efficacy of protein therapeutics are critical quality attributes (CQAs) that need to be controlled to ensure product quality. Peptide mapping with online mass spectrometry (MS) is a powerful tool that has been used for many years to monitor PTM CQAs during product development. However, operating peptide mapping methods with high-resolution mass spectrometers in GMP compliant, commercial quality control (QC) labs can be difficult. Peptide mapping is also required as an identity test in several countries. To address these two different needs, we utilized high-resolution peptide mapping for comprehensive characterization during development and then developed and validated a targeted multi-attribute monitoring (MAM) method using the low- resolution Waters QDa MS system with a fully automated data processing workflow that is suitable for identity (ID) testing, sequence variant control, and CQA quantitation in commercial QC labs. The ID-MAM method was validated for the quantitation of three selected PTM CQAs (CDR isomerization, Fc Met oxidation, and CDR Met oxidation) to ensure control of the oxidation and isomerization degradation pathways of a bispecific antibody (BsAb). This ID-MAM method was successfully validated in six labs (three analytical development and three QC labs) across four countries for commercial release and stability testing of a BsAb. CQA results obtained with the ID-MAM method were similar to results obtained using high-resolution peptide mapping, and the method was robust and reproducible. To our knowledge, this ID-MAM method is the first MS-based peptide mapping method implemented in GMP compliant QC labs for commercial release and stability testing of a biotherapeutic.
引用
收藏
页码:9166 / 9173
页数:8
相关论文
共 31 条
[1]   Determination of critical quality attributes for monoclonal antibodies using quality by design principles [J].
Alt, Nadja ;
Zhang, Taylor Y. ;
Motchnik, Paul ;
Taticek, Ron ;
Quarmby, Valerie ;
Schlothauer, Tilman ;
Beck, Hermann ;
Emrich, Thomas ;
Harris, Reed J. .
BIOLOGICALS, 2016, 44 (05) :291-305
[2]  
[Anonymous], 1996, ICH Harmonized Tripartite Guideline: Validation of Analytical Procedures: Methodology, P1
[3]  
[Anonymous], 2005, Validation of analytical procedures: Text and methodology Q2 (R1)
[4]  
[Anonymous], 2009, INT C HARMONIZATION, VQ8
[5]   Characterization of Therapeutic Antibodies and Related Products [J].
Beck, Alain ;
Wagner-Rousset, Elsa ;
Ayoub, Daniel ;
Van Dorsselaer, Alain ;
Sanglier-Cianferani, Sarah .
ANALYTICAL CHEMISTRY, 2013, 85 (02) :715-736
[6]   Biosimilar, Biobetter, and Next Generation Antibody Characterization by Mass Spectrometry [J].
Beck, Alain ;
Sanglier-Cianferani, Sarah ;
Van Dorsselaer, Alain .
ANALYTICAL CHEMISTRY, 2012, 84 (11) :4637-4646
[7]   Multi-Attribute Monitoring of Complex Erythropoetin Beta Glycosylation by GluC Liquid Chromatography-Mass Spectrometry Peptide Mapping [J].
Buettner, Alexander ;
Maier, Maria ;
Bonnington, Lea ;
Bulau, Patrick ;
Reusch, Dietmar .
ANALYTICAL CHEMISTRY, 2020, 92 (11) :7574-7580
[8]   COVALENT STRUCTURE OF AN ENTIRE GAMMAG IMMUNOGLOBULIN MOLECULE [J].
EDELMAN, GM ;
CUNNINGHAM, BA ;
GALL, WE ;
GOTTLIEB, PD ;
RUTISHAUSER, U ;
WAXDAL, MJ .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1969, 63 (01) :78-+
[9]   Chromatographic, Electrophoretic, and Mass Spectrometric Methods for the Analytical Characterization of Protein Biopharmaceuticals [J].
Fekete, Szabolcs ;
Guillarme, Davy ;
Sandra, Pat ;
Sandra, Koen .
ANALYTICAL CHEMISTRY, 2016, 88 (01) :480-507
[10]  
Jefferis R., 2018, J BIOCH BIOTECHNOL, V7, P63