Simulation study of PET detector limitations using continuous crystals

被引:13
作者
Cabello, Jorge [1 ]
Etxebeste, Ane [2 ]
Llosa, Gabriela [2 ]
Ziegler, Sibylle I. [1 ]
机构
[1] Tech Univ Munich, Klinikum Rechts Isar, Nukl Med Klin & Poliklin, D-80290 Munich, Germany
[2] CSIC Univ Valencia, Inst Fis Corpuscular, Valencia, Spain
关键词
continuous crystals; parallax effects; depth of interaction; high resolution; small animal PET; SMALL-ANIMAL PET; 3D INTERACTION POSITION; DEPTH; PERFORMANCE; HEAD; RESOLUTION; READOUT; DESIGN;
D O I
10.1088/0031-9155/60/9/3673
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Continuous crystals can potentially obtain better intrinsic detector spatial resolution compared to pixelated crystals, additionally providing depth of interaction (DoI) information from the light distribution. To achieve high performance sophisticated interaction position estimation algorithms are required. There are a number of algorithms in the literature applied to different crystal dimensions and different photodetectors. However, the different crystal properties and photodetector array geometries have an impact on the algorithm performance. In this work we analysed, through Monte Carlo simulations, different combinations of realistic crystals and photodetector parameters to better understand their influence on the interaction position estimation accuracy, with special emphasis on the DoI. We used an interaction position estimation based on an analytical model for the present work. Different photodetector granulation schemes were investigated. The impact of the number of crystal faces readout by photodetectors was studied by simulating scenarios with one and two photodetectors. In addition, crystals with different levels of reflection and aspect ratios (AR) were analysed. Results showed that the impact of photodetector granularity is mainly shown near the edges and specially in the corners of the crystal. The resulting intrinsic spatial resolution near the centre with a 12 x 12 x 10 mm(3) LYSO crystal was 0.7-0.9 mm, while the average spatial resolution calculated on the entire crystal was 0.77 +/- 0.18 mm for all the simulated geometries with one and two photodetectors. Having front and back photodetectors reduced the DoI bias (Euclidean distance between estimated DoI and real DoI) and improved the transversal resolution near the corners. In scenarios with one photodetector, small AR resulted in DoI inaccuracies for absorbed events at the entrance of the crystal. These inaccuracies were slightly reduced either by increasing the AR or reducing the amount of reflected light, and highly mitigated using two photodetectors. Using one photodetector, we obtained a piecewise DoI error model with a DoI resolution of 0.4-0.9 mm for a 1.2 AR crystal, and we observed that including a second photodetector or reducing the amount of reflections reduced the DoI bias but did not significantly improve the DoI resolution. Translating the piecewise DoI error model obtained in this study to image reconstruction we obtained a spatial resolution variability of 0.39 mm using 85% of the FoV, compared to 2.59 mm and 1.87 mm without DoI correction or with a dual layer system, respectively.
引用
收藏
页码:3673 / 3694
页数:22
相关论文
共 32 条
[1]   Maximum-Likelihood Methods for Processing Signals From Gamma-Ray Detectors [J].
Barrett, Harrison H. ;
Hunter, William C. J. ;
Miller, Brian William ;
Moore, Stephen K. ;
Chen, Yichun ;
Furenlid, Lars R. .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (03) :725-735
[2]   Neural network-based position estimators for PET detectors using monolithic LSO blocks [J].
Bruyndonckx, P ;
Léonard, S ;
Tavernier, S ;
Lemaître, C ;
Devroede, O ;
Wu, YB ;
Krieguer, M .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2004, 51 (05) :2520-2525
[3]   High resolution detectors based on continuous crystals and SiPMs for small animal PET [J].
Cabello, J. ;
Barrillon, P. ;
Barrio, J. ;
Bisogni, M. G. ;
Del Guerra, A. ;
Lacasta, C. ;
Rafecas, M. ;
Saikouk, H. ;
Solaz, C. ;
Solevi, P. ;
de La Taille, C. ;
Llosa, G. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2013, 718 :148-150
[4]  
Cherry S R., 2012, Physics in Nuclear Medicine, V4th, P43
[5]   An interior trust region approach for nonlinear minimization subject to bounds [J].
Coleman, TF ;
Li, YY .
SIAM JOURNAL ON OPTIMIZATION, 1996, 6 (02) :418-445
[6]   Continuous depth-of-interaction encoding using phosphor-coated scintillators [J].
Du, Huini ;
Yang, Yongfeng ;
Glodo, Jarek ;
Wu, Yibao ;
Shah, Kanai ;
Cherry, Simon R. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (06) :1757-1771
[7]   Study and optimization of positioning algorithms for monolithic PET detectors blocks [J].
Garcia de Acilu, P. ;
Sarasola, I. ;
Canadas, M. ;
Cuerdo, R. ;
Rato Mendes, P. ;
Romero, L. ;
Willmott, C. .
JOURNAL OF INSTRUMENTATION, 2012, 7
[8]   Multiple-Hit Parameter Estimation in Monolithic Detectors [J].
Hunter, William C. J. ;
Barrett, Harrison H. ;
Lewellen, Tom K. ;
Miyaoka, Robert S. .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2013, 32 (02) :329-337
[9]   Calibration Method for ML Estimation of 3D Interaction Position in a Thick Gamma-Ray Detector [J].
Hunter, William C. J. ;
Barrett, Harrison H. ;
Furenlid, Lars R. .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (01) :189-196
[10]   Scintillation properties of Ce doped Gd2Lu1(Ga,Al)5O12 single crystal grown by the micro-pulling-down method [J].
Kamada, Kei ;
Yanagida, Takayuki ;
Pejchal, Jan ;
Nikl, Martin ;
Endo, Takanori ;
Tsutsumi, Kousuke ;
Usuki, Yoshiyuki ;
Fujimoto, Yutaka ;
Fukabori, Akihiro ;
Yoshikawa, Akira .
JOURNAL OF CRYSTAL GROWTH, 2012, 352 (01) :35-38