Nebulizer choice for inhaled colistin treatment in cystic fibrosis

被引:33
作者
Katz, SL [1 ]
Ho, SL [1 ]
Coates, AL [1 ]
机构
[1] Univ Toronto, Hosp Sick Children, Div Resp Med, Toronto, ON M5G 1X8, Canada
关键词
cystic fibrosis; inhaled antibiotics; nebulizing colistin; Pseudomonas aeruginosa;
D O I
10.1378/chest.119.1.250
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Study objectives: To develop practical ways of nebulizing colistin by determining the rate of drug output, total drug output, and particle-size distribution of two commercially available jet nebulizers, the disposable Hudson 1730 Updraft II (Hudson Respiratory Care; Temecula, CA) and the reusable Pari LC Star breath-enhanced nebulizer (Pari Respiratory Equipment; Midlothian, VA). Methods: The nebulizers contained colistin, 75 mg, in 4 mL of isotonic solution. Particle-size distribution was measured by helium-neon laser diffraction, allowing calculation of the respirable fraction (RF), the mass of aerosol comprised of droplets < 5 <mu>m. Results: The mean (95% confidence interval [CI]) total rate of output of the Updraft II was 2.6 mg/min (2.0, 3.1; n = 4) with 1.3 mg/min (1.0, 1.5) mg/min within the RF. The rate of output of the LC Star increased in a quadratic relationship to the inspiratory flow, delivering 1.8 mg/min (0.7, 2.0; n = 4) with 1.4 mg/min (1.3, 1.6) within the RF, and 6.2 mg/min (5.6, 6.8) with 5.3 mg/min (4.8, 5.7) within the RF, at 0 L/min and 20 L/min inspiratory flows, respectively. Efficiency, as the rate of expected pulmonary deposition divided by rate of total output, was then calculated. The LC Star estimated 56% (51, 61) efficiency, with pulmonary delivery of 29% (26, 32) of the charge of the nebulizer, compared to the Updraft II at 22% (22, 23) efficiency and expected pulmonary deposition of 10% (10, 10) of the dose. Conclusions: Colistin can be successfully nebulized with both nebulizers tested. This study provides an estimate of in vivo efficiency and expected pulmonary deposition that may be used in future trials.
引用
收藏
页码:250 / 255
页数:6
相关论文
共 35 条
[21]  
HODSON ME, 1981, LANCET, V2, P1137
[22]   COLISTIN INHALATION-THERAPY IN CYSTIC-FIBROSIS PATIENTS WITH CHRONIC PSEUDOMONAS-AERUGINOSA LUNG INFECTION [J].
JENSEN, T ;
PEDERSEN, SS ;
GARNE, S ;
HEILMANN, C ;
HOIBY, N ;
KOCH, C .
JOURNAL OF ANTIMICROBIAL CHEMOTHERAPY, 1987, 19 (06) :831-838
[23]   The use of osmometry as a means of determining changes in drug concentration during jet nebulization [J].
Kwong, E ;
Macneish, CF ;
Meisner, D ;
Kelemen, S ;
Vadas, EB ;
Coates, AL .
JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG, 1998, 11 (02) :89-100
[24]  
KWONG WTJ, 2000, IN PRESS J AEROSOL M
[25]   SALBUTAMOL OUTPUT FROM 2 JET NEBULIZERS [J].
LANGFORD, SA ;
ALLEN, MB .
RESPIRATORY MEDICINE, 1993, 87 (02) :99-103
[26]   EFFECTS OF PH AND OSMOLARITY ON AEROSOL-INDUCED COUGH IN NORMAL VOLUNTEERS [J].
LOWRY, RH ;
WOOD, AM ;
HIGENBOTTAM, TW .
CLINICAL SCIENCE, 1988, 74 (04) :373-376
[27]   LONG-TERM EFFECTS OF INHALED TOBRAMYCIN IN PATIENTS WITH CYSTIC-FIBROSIS COLONIZED WITH PSEUDOMONAS-AERUGINOSA [J].
MACLUSKY, IB ;
GOLD, R ;
COREY, M ;
LEVISON, H .
PEDIATRIC PULMONOLOGY, 1989, 7 (01) :42-48
[28]  
NEWMAN SP, 1994, EUR RESPIR J, V7, P1177
[29]   EVALUATION OF JET NEBULIZERS FOR USE WITH GENTAMICIN SOLUTION [J].
NEWMAN, SP ;
PELLOW, PGD ;
CLAY, MM ;
CLARKE, SW .
THORAX, 1985, 40 (09) :671-676
[30]   INACCURATE CALCULATION OF DRUG OUTPUT FROM NEBULIZERS [J].
OCALLAGHAN, C ;
CLARKE, AR ;
MILNER, AD .
EUROPEAN JOURNAL OF PEDIATRICS, 1989, 148 (05) :473-474